Numerical modelling of a vascular palpation device for assessing intracranial aneurysm wall mechanics under heterogeneous conditions.
BACKGROUND The mechanical properties of the intracranial aneurysm wall are not taken into account in current rupture risk scores. A deformation device prototype was previously developed to characterize aneurysm wall mechanics in vivo through controlled deformation and inverse analysis. Its ability to detect localized m...