Sep 2026· Journal of Neurosurgery : Spine· pp.
1-8
· 1 citation· 28 references
Medicine
TL;DR
The automated VBQ measurement closely aligns with manual methods and may serve as an adjunct screening tool for osteoporosis in clinical practice and as a scalable research application suitable for integration into clinical workflows and large-scale MRI dataset analysis.
Abstract
Objective
This study aimed to develop and validate a fully automated vertebral bone quality (VBQ) pipeline for lumbar MRI studies and to assess its diagnostic performance with dual-energy x-ray absorptiometry (DXA)-defined osteoporosis.
Methods
Lumbar MRI studies from patients being evaluated for low-back pain were collected, deidentified, and assigned to reviewers for manual annotations. Each MRI study was manually annotated by two independent reviewers and by a custom, automated artificial intelligence (AI)-based segmentation algorithm. The measured signal intensity from the regions of interest (ROIs) was then extracted. Agreement between manual and automated methods was assessed using intraclass correlation coefficients (ICCs), root mean square error, and the paired t-test. The automated algorithm was validated on a separate cohort of 100 MRI examinations with paired DXA studies. Diagnostic performance for identifying osteoporosis was evaluated using receiver operating characteristic analysis to calculate the area under the curve (AUC).
Results
A total of 1089 patients were included to compare manual and automated VBQ calculations. Correlation between reviewer and automation was strong (r = 0.87, p < 0.001), and agreement was good with an ICC of 0.86 (95% CI 0.81-0.89). Diagnostic capacity was validated in a separate cohort (n = 100) with an AUC of 0.79 (95% CI 0.63-0.92). The Youden-optimized VBQ score of 2.59 increased sensitivity from 0.20-0.50 to 0.66-0.80 while maintaining high specificity at 0.83-0.96.
Conclusions
Automated VBQ measurement closely aligns with manual methods. An optimized threshold of 2.59 improved sensitivity for DXA-defined osteoporosis while maintaining high specificity. The automated algorithm may serve as an adjunct screening tool for osteoporosis in clinical practice and as a scalable research application suitable for integration into clinical workflows and large-scale MRI dataset analysis.
Volume, thickness, and defect area provide complementary measures of cartilage structure and Controlled geometric results and longitudinal associations supported the feasibility of threshold-based defect-area estimation.
Bin-Bin Yang, Yong-Mei Jian, Cheng-Lei Liu et al.· 0 citations
INTRODUCTION
Dual-energy X-ray absorptiometry (DXA) is the reference standard for osteoporosis diagnosis, but access remains limited in many settings. MRI-based vertebral bone quality (VBQ) has emerged as an opportunistic screening tool derived from routine lumbar imaging; however, its diagnostic performance is uncerta...
M. D. Wulandari, P. Jorisal, C. B. H. Sulay et al.· Radiography· 0 citations
RATIONALE AND OBJECTIVES
To assess the agreement between automated cartilage thickness measurements derived from three-dimensional (3D) dual-echo steady-state (DESS) magnetic resonance imaging and manual radiologist measurements.
MATERIALS AND METHODS
In 111 patients, cartilage thickness measurements across 21 knee s...
Mirela Kostadinova, Iman Harrou, Scherwin Mahmoudi et al.· Academic Radiology· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.