Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.

Sep 2026 · American Journal of Hypertension · 0 citations
Medicine

TL;DR

Although the cuffless device shows promise in successfully tracking nocturnal BP, full validation is required before clinical use.

Abstract

Background

Cuffless blood pressure (BP) devices have demonstrated potential in estimating daytime BP but have not been able to predict nocturnal BP precisely, creating difficulties when the intended use is ambulatory BP monitoring (ABPM). We evaluated the ability of a photoplethysmography-based cuffless BP device, with the intended use 24-hour ABPM, to estimate nocturnal BP compared to a cuff-based oscillometric device.

Methods

Participants underwent 24-hour ABPM with the cuffless device placed on the non-dominant upper arm and a validated cuff-based oscillometric device on the dominant arm. The difference between mean awake and asleep BP (awake-asleep BP change) was calculated for both devices and used in the analyses. The statistical analyses were guided by the Awake/Asleep test (European Society of Hypertension 2023 Recommendations), however, there were methodological deviances from these recommendations. Most importantly, we used only cuffless BP values that were obtained simultaneously with the reference device.

Results

Data from 50 participants, aged 18-80 years, with mean (standard deviation; SD) inclusion office systolic/diastolic BP (SBP/DBP) 120.1 (11.5)/75.3 (7.9), were analyzed. The cuffless device achieved a mean difference [95% limits of agreement] of 1.6 [-8.0, 11.2] mmHg for SBP and 1.2 [-6.7, 9.1] mmHg for DBP in awake-asleep BP changes compared to the reference. Pearson correlation coefficients (95% confidence interval) for SBP/DBP were 0.75 (0.59, 0.85)/0.74 (0.58, 0.84). Our population had an overrepresentation of normal and low BP.

Conclusion

Although the cuffless device shows promise in successfully tracking nocturnal BP, full validation is required before clinical use.

Read PDF

Similar papers

Open access Sep 2026

Long-Term Stability and Blood Pressure Change Tracking with a Cuffless Wearable Device in Hypertensive Patients.

CSEM's wearable PPG-based cuffless BP monitor appears to provide stable BP estimation and concordant tracking of week-to-month BP changes during antihypertensive therapy when compared with an oscillometric device, and could support the clinical acceptance of PPG-based wearable BP monitors.

M. Proença, Pierre-Antonin Rigon, G. Bonnier et al. · 0 citations
Open access Sep 2026

Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring.

BACKGROUND Cuffless wearable technologies enable continual blood pressure (BP) monitoring and may improve tolerability compared with conventional cuff-based methods. Pediatric data comparing cuffless devices with ambulatory blood pressure monitoring (ABPM) remain limited. METHODS In this prospective observational stu...

M. Huml, A. Kotková, Tereza Fremuthová et al. · 0 citations
#software testing Open access Aug 2026

Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.

This represents the first evaluation of a Device Type 3 cuffless blood pressure algorithm (automated, wearable, demographic-calibrated, not at heart level) meeting ESH awake/asleep test criteria, suggesting Alysis-001's potential as a clinically viable alternative for ambulatory blood pressure monitoring in hypertensiv...

K. Hongyo, Atsushi Hirayama, Kosuke Shimizu et al. · 0 citations
Review Aug 2026

Wearable wrist-watch type cuff oscillometric blood pressure monitors: consensus statement by the European Society of Hypertension Working Group on blood pressure monitoring.

More research is needed to establish the accuracy and clinical utility of these novel devices and their role in improving the diagnosis and management of hypertension.

G. Stergiou, A. Menti, Kei Asayama et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.