Transitioning Beyond Mercury Sphygmomanometers: Diagnostic Accuracy and Agreement of Digital Blood Pressure Monitors in a South Indian Adult Population

Aug 2026 · Cureus · Vol 18 · 0 citations · 18 references
Medicine

TL;DR

The automated digital monitor demonstrates good population-level agreement and diagnostic validity, making it a reliable tool for routine screening and epidemiological surveillance, however, given the wide limits of agreement, critical clinical decisions, especially in borderline cases, should be made with caution.

Abstract

Introduction: Hypertension is a leading cause of morbidity and mortality, and its prevalence is increasing worldwide. Accurate blood pressure (BP) measurement is fundamental to clinical diagnosis and epidemiological surveillance, and any measurement errors can lead to substantial misclassification of hypertension, resulting in inappropriate clinical management and inaccurate estimates of disease burden. While automated digital monitors offer a practical alternative to manual devices, their real-world accuracy requires robust local validation during the phase-out of traditional instruments. The objective of the current study was to assess the validity and accuracy of a digital BP monitor against the gold-standard mercury device in an adult population in South India. Materials and methods: A cross-sectional diagnostic evaluation study was conducted among 262 adults attending the outpatient department of a teaching hospital in South India. Baseline BP was measured twice using the digital BP monitor (Omron HEM-7120, OMRON Healthcare Co., Ltd., Kyoto, Japan) with a one- to two-minute interval between readings. After a five-minute interval, BP was measured using a mercury sphygmomanometer (Novaphon, NOVAFON GmbH, Weinstadt, Germany). Validity, reliability, agreement, and bias were analyzed using paired t-tests, Pearson correlation, and Bland-Altman analysis. Results: Digital and mercury measurements demonstrated exceptionally strong linear correlations for systolic BP (r = 0.945) and diastolic BP (r = 0.845). The mean difference (digital minus mercury) was non-significant for both parameters: -0.167 ± 7.46 mmHg for systolic and 0.296 ± 6.63 mmHg for diastolic. When comparing the two devices, absolute differences within 5 mmHg were observed in 135 (51.5%) participants for systolic BP and 164 (62.6%) participants for diastolic BP. Categorical assessment showed substantial diagnostic agreement (Cohen’s kappa = 0.77), with a sensitivity of 84.0% and specificity of 92.9% for hypertension detection. Bland-Altman analysis revealed a negligible mean bias of −0.167 mmHg for systolic BP (95% limits of agreement (LoA): −14.79 to 14.46 mmHg) and 0.296 mmHg for diastolic BP (95% LoA: −12.70 to 13.29 mmHg). Conclusions: The automated digital monitor demonstrates good population-level agreement and diagnostic validity, making it a reliable tool for routine screening and epidemiological surveillance. However, given the wide limits of agreement, critical clinical decisions, especially in borderline cases, should be made with caution.

Read PDF

Similar papers

2026

A Novel Instrument for Research and Calibration of Noninvasive Blood Pressure Measurement

Hypertension and cardiovascular diseases (CVDs) are the major causes of death, hospital admissions, and disability in the middle-aged, and regular blood pressure (BP) measurement is one of the simplest and most effective public health interventions available. Yet we know that noninvasive BP (NIBP) devices, ostensibly c...

Branko G. Celler, Raisa Rudino, A. Argha · 0 citations
Open access Aug 2026

Diagnostic Value of Ambulatory Blood Pressure Monitoring for Hypertension in Private Cardiology Practice in Senegal

ABPM improves the diagnostic classification of hypertension and reveals a high prevalence of discordant phenotypes and nocturnal blood pressure abnormalities, and wider use of ABPM may help optimize patient management in this setting.

N. Gaye, J. Mingou, Coumba Kaba et al. · 0 citations
#software testing Open access Aug 2026

Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.

This represents the first evaluation of a Device Type 3 cuffless blood pressure algorithm (automated, wearable, demographic-calibrated, not at heart level) meeting ESH awake/asleep test criteria, suggesting Alysis-001's potential as a clinically viable alternative for ambulatory blood pressure monitoring in hypertensiv...

K. Hongyo, Atsushi Hirayama, Kosuke Shimizu et al. · 0 citations
Open access Sep 2026

Routine versus standardized blood pressure measurements: Agreement and association with cardiovascular disease.

BACKGROUND Recommendations for blood pressure (BP) measurement are poorly followed in clinical practice, possibly leading to inaccurate BP estimates. OBJECTIVES To assess the agreement between routine and standardized BPs, and the association between BP level and cardiovascular disease (CVD) for each measurement type...

O. Gastgivars, J. Ögren, B. Carlberg et al. · 0 citations
Open access Sep 2026

Diagnostic accuracy of attended versus unattended automated blood pressure measurements compared to ambulatory blood pressure monitoring conducted by community health workers in rural sub-Saharan Africa

Community Health Workers (CHWs) are increasingly being involved in screening for hypertension in rural sub-Saharan Africa. However the diagnostic accuracy of CHW-led attended (CHW present during procedure) and unattended (CHW absent) automated blood pressure (BP) in home settings has not been compared. We...

R. Lucinde, Modu Jobe, Alexander D. Perkins et al. · 0 citations
Open access Sep 2026

Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography and electrocardiography.

PURPOSE Continuous, non-invasive blood pressure (BP) monitoring is essential for understanding cardiovascular dynamics in real-world settings. This study evaluated the reliability and validity of a cuffless method for ambulatory BP monitoring using pulse wave travel times derived from bio-impedance, impedance cardiogra...

Myrte Schoenmakers, Martin J. Gevonden, B. Sauce et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.