Commissioning a single photon counting CT simulator for radiotherapy planning: Methods, challenges, and initial results

Sep 2026 · Journal of Applied Clinical Medical Physics · Vol 27 · 0 citations · 22 references
Medicine

TL;DR

The Siemens Naeotom Alpha PCCT scanner satisfies the requirements noted in task group reports for a radiotherapy simulator and has the potential to offer advantages for radiation oncology applications, including improved target delineation and treatment planning accuracy.

Abstract

Abstract Background Photon counting computed tomography (PCCT) offers distinct advantages over standard energy‐integrated detector (EID) CT by offering energy‐resolved measurements, enhanced spatial resolution, and improved soft‐tissue contrast. While PCCT shows great utility in diagnostic imaging, its application as a radiotherapy simulator has not been explored. Commissioning procedures for EID‐based CT simulators are well‐documented, but commissioning a PCCT scanner in a radiotherapy setting has not been investigated. Purpose We investigated whether PCCT can be used for radiotherapy simulation by evaluating the geometric accuracy, image quality, dosimetric performance, and relative electron density calibration of a Siemens Naeotom Alpha PCCT scanner with a commercially available carbon‐fiber couch top. Methods Commissioning followed AAPM Task Group 66 guidelines, involving electromechanical evaluation, image quality assessment, and radiation safety verification. The CATPHAN 604 phantom was used to evaluate CT number accuracy, uniformity, spatial resolution, low contrast detectability, and geometric accuracy across multiple clinical protocols. Dosimetric measurements were performed using CTDI phantoms with a 100‐mm pencil ion chamber. Relative electron density (RED) calibration was performed using an Advanced Electron Density phantom at selected tube voltages (120 and 140 kVp), slice thicknesses (0.4–10 mm), and reconstruction filters. Statistical analyses included repeated‐measures ANOVA with post‐hoc pairwise comparisons to assess protocol and position effects on Hounsfield unit values. Results Geometric accuracy measurements revealed in‐plane spatial linearity of 50.00 ± 0.06 mm (nominal 50 mm). CT number accuracy aligned with manufacturer specifications, except for bone, using specific protocols. The mean discrepancy between measured and nominal CTDIvol was 3.8 ± 4.6%, with excellent agreement on the dose‐length product (0.9 ± 2.3%). Relative electron density measurements suggested statistically significant, but clinically insignificant, variations across protocols (mean HU shifts: ‐23.0 to 7.9 HU). Position‐dependent effects suggested mean HU reductions up to 23.2 HU at peripheral locations. Iterative metal artifact reduction (iMAR) differentiated Titanium and Stainless Steel metallic implants without significantly affecting surrounding tissue CT numbers (P > 0.08). Conclusions The Siemens Naeotom Alpha PCCT scanner satisfies the requirements noted in task group reports for a radiotherapy simulator. The system's enhanced spatial resolution, spectral imaging capabilities, and flexible retrospective reconstruction have the potential to offer advantages for radiation oncology applications, including improved target delineation and treatment planning accuracy.

Read PDF

Similar papers

Comprehensive acceptance testing and quality control of Single Photon Emission Computed Tomography/Computed Tomography systems.

The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality and developed a structured quality control program based on acceptance testing results.

N. Abbasi, H. Younis, S. Ur Rehman et al. · 0 citations
Open access Aug 2026

A commissioning workflow for quantitative ¹⁷⁷Lu SPECT/CT across different Siemens scanner–reconstruction configurations

A practical commissioning workflow to optimize ¹⁷⁷Lu SPECT/CT imaging across different commercial scanner–reconstruction configurations is developed and a practical framework for local ¹⁷⁷Lu SPECT/CT optimization is provided.

L. Lampertico, G. Muti, C. Carbonini et al. · 0 citations
Open access Oct 2026

Comparison of proton, helium-ion, and carbon-ion CT imaging for hadron radiotherapy using Monte Carlo simulations.

OBJECTIVE To perform a comprehensive comparative analysis of proton, helium-ion, and carbon-ion computed tomography (CT) as direct imaging modalities for hadron therapy treatment planning, focusing on Relative Stopping Power (RSP) reconstruction accuracy and patient radiation dose. Approach: High-fidelity Monte Carlo s...

Z. Jólesz, G. Bíró, G. Papp et al. · 0 citations
Open access Sep 2026

Optimization of reconstruction strategies for head and neck computed tomography angiography using photon-counting computed tomography: A comparative study with energy-integrating detector computed tomography.

BACKGROUND Head and neck computed tomography angiography (CTA) requires high spatial resolution while maintaining acceptable image noise. This study aimed to compare image quality between photon-counting CT (PCCT) and energy-integrating detector CT (EID-CT) and to evaluate PCCT reconstruction strategies for specific cl...

Wei Qiao, Xiao Lu, Wei Zhou et al. · 0 citations
Open access Oct 2026

Correlation of size specific dose estimates with organ dose in abdominal spectral computed tomography

Abstract Background In abdominal spectral Computed Tomography (CT), organ dose assessment is important for radiation safety but remains challenging due to the complexity of Monte Carlo (MC) simulations. Purpose This study aimed to compare the correlations of three size‐specific dose estimate (SSDE) metrics (including t...

Tian Qin, Qing-Ting Sun, Yi-Han Fan et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.