Impact of CT parameter reduction on quantitative SPECT/CT assessment in 99mTc imaging: A phantom study.

Sep 2026 · Radiography · Vol 32 7, pp. 103569 · 0 citations · 26 references
Medicine

TL;DR

The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.

Abstract

INTRODUCTION Quantitative SPECT/CT provides objective assessment of radiotracer distribution but may be influenced by CT acquisition parameters used for attenuation correction and by object related factors such as the partial volume effect. Optimization of CT protocols is therefore important to reduce radiation exposure while maintaining quantitative reliability.

Methods

A study evaluated the effect of CT acquisition parameters on quantitative SPECT/CT measurements. A NEMA IEC Body phantom containing six spheres (10-37 mm) with a target to background activity ratio of 10:1 was scanned using six CT protocols combining 80, 110, and 130 kV with 35 and 70 mAs. Quantitative measurements were analysed using Shapiro-Wilk test, multiple regression, multiple linear regression, and Levene's test.

Results

Measurements showed normal distribution for CNR on CT to the Shapiro-Wilk test (p > 0.05), whereas SUVmean, SUVmax, and CNR on SPECT showed non-normal distribution (p < 0.05). Sphere diameter demonstrated a strong positive correlation with quantitative values (r ≈ 0.99, p < 0.001), whereas CT acquisition parameters showed no significant effect. Regression analysis identified sphere diameter as the main determinant of measurement variability. Levene's test confirmed homogeneous variance between acquisition protocols (p > 0.05).

Conclusion

Quantitative SPECT/CT measurements remained stable across the investigated CT protocols. Object size was the main factor influencing quantitative accuracy, likely due to partial volume effects. IMPLICATION FOR PRACTICE The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.

View source

Similar papers

Aug 2026

Clinical evaluation of Chang attenuation-correction parameterization for CT-free 177Lu SPECT organ quantification: implications for safety dosimetry.

BACKGROUND Quantitative post-therapy 177Lu SPECT depends on reliable attenuation correction. For centers without SPECT/CT, CT-free Chang attenuation correction is feasible, but its dependence on emission-derived body-contour thresholding requires validation. METHODS Twenty patients treated with 177Lu-DOTATATE underwe...

Abrar Qasım Mustafa Al-Ramadhani, Ahmet Murat Şenışık, Bilal Kovan et al. · 0 citations
Open access Sep 2026

Quantitative accuracy and clinical utility of ultra-low dose CT protocols for attenuation correction in PET/CT

In PET/CT imaging, CT frequently accounts for more than 50% of the total radiation exposure, even when used solely for attenuation correction and anatomical localisation. This poses a major limitation for dose-sensitive populations, including pregnant women and paediatric patients. This study systematically evalu...

Georgios I. Angelis, Y. Gholami, K. Willowson et al. · 0 citations
Sep 2026

Quantitative Comparison of Virtual and True Noncontrast Lung Images in Photon-counting CT: Global and Regional Attenuation Assessment.

Background Use of virtual noncontrast (VNC) images from photon-counting CT (PCCT) may reduce radiation dose and support quantitative lung analysis by obviating the need for separate true noncontrast (TNC) acquisitions. Their accuracy for lung densitometry, however, has not been systematically assessed. Purpose To evalu...

Max L. Bodlien, S. Scharm, A. Hunkemöller et al. · 0 citations
Review Open access Sep 2026

Quantification in bone SPECT/CT

Background Quantitative bone SPECT/CT enables objective assessment of skeletal radiotracer uptake using standardised uptake values (SUVs) and related parameters. Advances in reconstruction, correction methods, calibration and detector technology have improved measurement reliability and expanded interest in oncological...

Jovana Živanović, Jules Zhang-Yin · 1 citation
Sep 2026

EARL-based quantitative evaluation of administered activity reduction for [68Ga]Ga-DOTATOC PET/CT on a 32-cm axial FOV digital BGO system: Phantom analysis and retrospective list-mode clinical assessment.

PURPOSE The increased sensitivity of long-axial field-of-view (FOV) digital PET/CT systems enables potential reductions in administered activity and/or acquisition time with respect to standard-field-of-view systems and established clinical protocols, while preserving quantitative performance. This study aimed to evalu...

A. Piai, B. Bordigoni, L. Antunovic et al. · 0 citations
Open access Sep 2026

Effects of count rate performance on image quality in an LAFOV PET/CT system

This study investigated the relationship between body size, administered activity, and image quality (IQ) using two patient-like phantoms representing normal and obese body profiles, and demonstrated the count rate capabilities up to and beyond the saturation limit.

Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.