The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.
Abstract
INTRODUCTION
Quantitative SPECT/CT provides objective assessment of radiotracer distribution but may be influenced by CT acquisition parameters used for attenuation correction and by object related factors such as the partial volume effect. Optimization of CT protocols is therefore important to reduce radiation exposure while maintaining quantitative reliability.
Methods
A study evaluated the effect of CT acquisition parameters on quantitative SPECT/CT measurements. A NEMA IEC Body phantom containing six spheres (10-37 mm) with a target to background activity ratio of 10:1 was scanned using six CT protocols combining 80, 110, and 130 kV with 35 and 70 mAs. Quantitative measurements were analysed using Shapiro-Wilk test, multiple regression, multiple linear regression, and Levene's test.
Results
Measurements showed normal distribution for CNR on CT to the Shapiro-Wilk test (p > 0.05), whereas SUVmean, SUVmax, and CNR on SPECT showed non-normal distribution (p < 0.05). Sphere diameter demonstrated a strong positive correlation with quantitative values (r ≈ 0.99, p < 0.001), whereas CT acquisition parameters showed no significant effect. Regression analysis identified sphere diameter as the main determinant of measurement variability. Levene's test confirmed homogeneous variance between acquisition protocols (p > 0.05).
Conclusion
Quantitative SPECT/CT measurements remained stable across the investigated CT protocols. Object size was the main factor influencing quantitative accuracy, likely due to partial volume effects.
IMPLICATION FOR PRACTICE
The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.
BACKGROUND
Quantitative post-therapy 177Lu SPECT depends on reliable attenuation correction. For centers without SPECT/CT, CT-free Chang attenuation correction is feasible, but its dependence on emission-derived body-contour thresholding requires validation.
METHODS
Twenty patients treated with 177Lu-DOTATATE underwe...
Abrar Qasım Mustafa Al-Ramadhani, Ahmet Murat Şenışık, Bilal Kovan et al.· Applied Radiation and Isotop...· 0 citations
In PET/CT imaging, CT frequently accounts for more than 50% of the total radiation exposure, even when used solely for attenuation correction and anatomical localisation. This poses a major limitation for dose-sensitive populations, including pregnant women and paediatric patients. This study systematically evalu...
Georgios I. Angelis, Y. Gholami, K. Willowson et al.· EJNMMI Physics· 0 citations
Background Use of virtual noncontrast (VNC) images from photon-counting CT (PCCT) may reduce radiation dose and support quantitative lung analysis by obviating the need for separate true noncontrast (TNC) acquisitions. Their accuracy for lung densitometry, however, has not been systematically assessed. Purpose To evalu...
Max L. Bodlien, S. Scharm, A. Hunkemöller et al.· Radiology· 0 citations
Background Quantitative bone SPECT/CT enables objective assessment of skeletal radiotracer uptake using standardised uptake values (SUVs) and related parameters. Advances in reconstruction, correction methods, calibration and detector technology have improved measurement reliability and expanded interest in oncological...
Jovana Živanović, Jules Zhang-Yin· Frontiers in Nuclear Medicin...· 1 citation
PURPOSE
The increased sensitivity of long-axial field-of-view (FOV) digital PET/CT systems enables potential reductions in administered activity and/or acquisition time with respect to standard-field-of-view systems and established clinical protocols, while preserving quantitative performance. This study aimed to evalu...
A. Piai, B. Bordigoni, L. Antunovic et al.· Physica medica (Testo stampa...· 0 citations
This study investigated the relationship between body size, administered activity, and image quality (IQ) using two patient-like phantoms representing normal and obese body profiles, and demonstrated the count rate capabilities up to and beyond the saturation limit.
Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al.· Physics in Medicine and Biol...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.