Accuracy of an Interstitial Continuous Glucose Management System in Critically Ill Patients: A Prospective Observational Study.

Sep 2026 · Journal of Diabetes Science and Technology · pp. 19322968261486451 · 0 citations · 36 references
Medicine

Abstract

INTRODUCTION Glycemic management in the intensive care unit (ICU) remains challenging because glucose levels fluctuate rapidly and monitoring is intermittent. Interstitial continuous glucose monitoring (CGM) could improve surveillance, but its performance during rapid glucose excursions remains uncertain. We assessed the accuracy of the Dexcom G7 CGM in critically ill patients under stable conditions and during intravenous glucose tolerance tests (IVGTT).

Methods

We conducted a prospective ancillary study of the Determination of Insulin Resistance in ICU (DIRICU) study. Twenty critically ill adults with stress hyperglycemia, no prior diabetes, and expected ICU stay ≥5 days were included. A Dexcom G7 sensor was compared with arterial blood glucose measurements. Accuracy was assessed over the first 10 ICU days using paired measurements, mean bias, mean absolute relative difference (MARD), linear regression (R2), and Clarke error grid analysis, both outside and during IVGTT. A post hoc 20-minute temporal shift analysis was performed to account for delayed CGM peaks.

Results

Outside IVGTT, 716 paired values showed moderate agreement with reference glucose (R2 = .58, mean bias +11 mg/dL, MARD 17%), with 98.9% of values in Clarke zones A/B. During IVGTT, accuracy deteriorated markedly (1018 paired values): R2 = .028, mean bias -34 mg/dL, MARD 24%, and only 80.3% of values in zones A/B, with 18.6% in zone D. After temporal shifting, performance improved (R2 = .34, MARD 19%, 94.7% in zones A/B). No sensor-related adverse events occurred.

Conclusions

Dexcom G7 showed acceptable accuracy during stable ICU conditions, but performance deteriorated during rapid glucose excursions. Hybrid monitoring strategies combining CGM with intermittent blood glucose confirmation may therefore be preferable in critically ill patients.

Read PDF

Similar papers

Open access Oct 2026

Continuous glucose monitoring for diabetes mellitus in the perioperative and intensive care setting: a prospective observational validation study✰.

BACKGROUND In-hospital diabetes mellitus management currently relies on intermittent glucose monitoring with point-of-care (POC) testing, rather than continuous glucose monitoring (CGM). Most studies on CGM have focused on its benefits and accuracy in the outpatient setting, in contrast to the hospital setting, where t...

C. Pedersen, Aslak Johansen, A. Philipsen et al. · 0 citations
Open access Oct 2026

Glycemic Variability and Its Outcome in Intensive Care Unit Patients with Sepsis in a Tertiary Care Centre, Puducherry

Background : Sepsis-induced metabolic dysregulation produces pronounced fluctuations in blood glucose that persist even in the absence of chronic diabetes. While absolute hyperglycaemia has historically guided glycaemic management in intensive care, emerging evidence suggests that glycemic variability (GV)—the degree o...

V. Priya · 0 citations
Open access Oct 2026

Comparison of the Safety and Efficacy of Continuous Insulin Infusion Versus Intermittent Insulin Bolus in the Management of Hyperglycemia in Critically Ill Patients

Objective: To compare the efficacy and safety of intravenous continuous Insulin infusion (CII) versus Intermittent Insulin Bolus (IIB) in managing blood glucose levels in critically ill patients. Study Design: Quasi-experimental study. Place and Duration of Study: Intensive care units (ICUs) of a tertiary care hospital...

Muhammad Rafi Ismail, Wahaj Hassan, Fuad Ahmad Siddiqi et al. · 0 citations
Open access Sep 2026

Pediatric Inpatient Continuous Glucose Monitor Accuracy by Sensor Wear Day, Hospital Day, and Glucose Rate of Change

In this pediatric hospital CGM accuracy study, MARD was within a favorable range of <14% overall and throughout sensor wear and hospitalization, highlighting the need to incorporate trend arrows into hospital CGM protocols.

Jessica L. Ruiz, A. Stein, Alexander Wheelock et al. · 0 citations
Open access Sep 2026

Clinical performance and practical implications of continuous glucose monitoring in very preterm infants: a cohort study.

OBJECTIVE To explore the clinical performance of continuous glucose monitoring (CGM) in very preterm infants during the first week of life, using data from the multicentre Real time continuous glucose monitoring (REACT) randomised controlled trial (RCT) and to consider how CGM can best support neonatal glucose manageme...

G. Liu, L. Marcovecchio, K. Beardsall · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.