Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise

Sep 2026 · Bioengineering · Vol 13, pp. 1099 · 0 citations · 26 references

TL;DR

It is demonstrated that the agreement of smartwatch-derived HR with electrocardiography measurements varies across time points during resistance exercise and indicates that smartwatch-derived workout-session EE should be interpreted cautiously when accurate metabolic assessment is required.

Abstract

Commercial smartwatches are increasingly used to monitor physiological responses during exercise; however, their validity for resistance exercise remains uncertain. This study evaluated the criterion validity of four commercial smartwatches (Apple Watch Series 8, Samsung Galaxy Watch 5, Fitbit Charge 5, and Garmin Vivosmart 5) for measuring heart rate (HR) and energy expenditure (EE) during resistance exercise. Seventy healthy men aged 20–39 years performed two sets each of the bench press, T-bar row, squat, and deadlift at individualized 10-repetition maximum loads. Smartwatch-derived HR was evaluated against electrocardiography, whereas smartwatch-derived workout-session EE was evaluated against indirect calorimetry. The Apple Watch showed the lowest measurement error for mean HR across the protocol (bias = −0.36 bpm; MAE = 1.81 bpm; ICC = 0.969) and for overall dynamic HR (bias = −0.72 bpm; MAE = 4.95 bpm; ICC = 0.904). HR agreement deteriorated markedly immediately after each set, particularly for the Samsung and Fitbit devices, which underestimated ECG-derived HR by 25.60 and 19.59 bpm, respectively. In contrast, EE estimates showed poor agreement with indirect calorimetry across all devices (ICC = 0.021–0.033). These findings demonstrate that the agreement of smartwatch-derived HR with electrocardiography measurements varies across time points during resistance exercise and indicate that smartwatch-derived workout-session EE should be interpreted cautiously when accurate metabolic assessment is required.

Read PDF

Similar papers

Conference Open access 2026

Accuracy of Wearable Heart Rate Monitors under Static and High-Intensity Activity Conditions

The widespread adoption of wearable devices has created new opportunities for continuous monitoring of physical activity, cardiovascular health, and sleep patterns. However, it remains an open question to what extent the physiological and kinematic data provided by commercial smartwatches can be considered accurate and...

János Dávid Balogh, Attila Adamkó · 0 citations
Open access Sep 2026

The accuracy of VO2 max, heart rate and energy expenditure measurements from Google Pixel Watch 3

Consumer wearables enable scalable physiological assessment and pre-symptomatic disease detection. However, validation of wearable data remains insufficient in scope and rigour, hindering integration into clinical and public health practice. Here, we evaluated the validity of VO2max, heart rate and energy expenditure f...

Rory Lambe, Moritz Schumann, Sherene Hamilton et al. · 0 citations
Open access 2025

Impact of physical exercise on the accuracy of blood pressure and heart rate data measured by smartwatch

INTRODUCTION: Heart rate (HR) and blood pressure (BP)—systolic (SBP) and diastolic (DBP)—are important for individuals who engage in physical activity. Smartwatches stand out because they are convenient and collect data noninvasively; however, the information they provide during physical exercise still requires validat...

Julia Basolli Gomiero, Gustavo D'Annibale Bech, Thiago Song Won Yoon et al. · 0 citations
Open access Sep 2026

Effect of exercise-induced short-term changes in blood pressure and heart rate on smartwatch measurement accuracy

Blood pressure (BP) is a key vital sign that many people around the world are increasingly monitoring continuously using smartwatches. It remains unclear whether smartwatches can fully replace traditional oscillometric cuff-based devices under real-world conditions. The aim of this study was to verify the accuracy of B...

Veronika Ráfl Huttová, Šimon Walzel, M. Rožánek et al. · 0 citations
Open access Oct 2026

Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report

Background/Objectives: Wearable devices offer a scalable approach to repeated cardiovascular assessment through ambulatory spot measurements and have demonstrated adequate performance in measuring heart rate (HR). However, their reliability for blood pressure (BP) assessment remains uncertain. This study aimed to evalu...

Octavio A. Lecona, V. Martínez-Garcés, B. Estañol et al. · 0 citations
Open access Sep 2026

VALIDATION OF HEART RATE MEASUREMENTS FROM THE POLAR VERITY SENSE DURING THE SUBMAXIMAL EKBLOM-BAK TEST IN ADULT INDIVIDUALS WITH OBESITY

Objective To assess the validity of Polar Verity Sense, an arm-worn heart rate (HR) monitor, for measuring HR in individuals with obesity during the submaximal ergometer cycle Ekblom-Bak test. Subjects/Patients Adults in an inpatient rehabilitation programme for individuals with obesity. Methods Participants completed...

Grete Rudi Onstad, Fredrik Klæboe Lohne, M. S. Fimland · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.