A-210 Analytical Performance of 25-OH Total Vitamin D Chemiluminescent Immunoassay Using DiaSorin LIASON XL Analyzer
TL;DR
DiaSorin LIASON XL 25-OH Total vitamin D immunoassay demonstrated acceptable and good analytical performance during assay verification and after implementation as part of the monitoring of patient testing.
Abstract
Vitamin D plays an essential role as a hormone in the control of calcium and phosphorus metabolism. Total 25-OH vitamin D concentration is considered the best indicator of vitamin D nutritional status, and it’s also recognized as a prohormone which has multiple roles in maintaining optimal health. It is important that an assay does not measure the inactive 3-epi compound usually detected in infants and only measures 25-OH vitamin D2 and D3 equally. Analytical performance evaluation and verification of new assays is required before placing them in use for patient testing. The objective of this study was to verify the analytical performance of precision, bias estimation, analytical measurement of range (AMR), reference ranges and monitor assay performance using peer group approach. 25-OH Total vitamin D immunoassay was performed using DiaSorin LIASON XL Analyzer (DiaSorin Inc. MN, USA). Assay precision and estimation of bias performance was conducted using Quality Control (QC) material from manufacturers (L1,L2), Bio-Rad QC (L1,L2,L3), including residual/spiked patient samples with low, normal and elevated levels. For precision, each material was performed in 10 replicates to verify within-run precision, and 1-2 replicates during 10-15 days for between-run precision. In addition, a CLSI EP15-A3 protocol was performed: 5 replicates of each QC material level during 5 days. AMR and reportable range verification were performed in duplicate using manufacturer and Chromosystem material. Instrument comparisons using patient residual samples (n=36) were performed between DiaSorin LIASON XL Analyzer and LC-MS/MS 4500MD Sciex Triple Quad (MA, USA) and with DiaSorin LIASON XL Analyzer from another pediatric hospital laboratory. With-run and between-run precision (SD, %CV), linear regression and bias plot analysis were calculated using Microsoft Excel. Assay performance was also monitored using Standard Deviation Index (SDI) and Coefficient of Variation Ratio (CVR) with monthly Bio-Rad instrument specific peer group reports. No potential outliers were identified using Grubb’s test. Overall, the precision observed for with-run and between-run precision were in general up to 5.3%, and up to 6.6%, and =1.0 and =1.5ng/mL for concentration <20ng/mL, respectively. Manufacturer‘s precision performance was also verified using CLSI EP15 approach using DiaSorin and Bio-Rad QC material. Verified AMR and reportable range were 4-150ng/mL and 4-600ng/mL, respectively. Instrument comparison between DiaSorin LIASON XL Analyzer and LC-MS/MS 4500MD AB Sciex Triple Quad and with other DiaSorin LIASON XL Analyzer had a slope of 0.9699 and 0.9839, with intercepts of 3.56 and 3.75, and coefficient correlation of 0.9891 and 0.9919 with assays mean absolute differences of 2.0 and 3.2 ng/mL, respectively. Overall relative bias estimation using CLSI EP15 approach showed up to 0.4ng/mL, and 9.3ng/mL of relative bias for concentrations =20 and =99ng/mL, respectively. Bio-Rad SDI and CVR during the next 4 months after implementation were =0.95 and =0.90, respectively. Pediatric Endocrine Society (PES) vitamin D status decision limits were verified and placed in use. DiaSorin LIASON XL 25-OH Total vitamin D immunoassay demonstrated acceptable and good analytical performance during assay verification and after implementation as part of the monitoring of patient testing.