iCBCT with a dedicated HU-density calibration curve provides dose calculation accuracy comparable to planning CT, supporting reliable offline ART in prostate radiotherapy.
Abstract
Background
Adaptive Radiotherapy (ART) updates treatment plans based on newly acquired anatomical information to maintain dosimetric accuracy. Image-Guided Radiotherapy (IGRT) is essential for detecting anatomical changes and guiding replanning decisions.
Purpose
This study evaluates the feasibility of using iterative Cone Beam CT (iCBCT) acquired on a TrueBeam system as a reference image dataset for dose calculation and offline ART in prostate radiotherapy. The approach avoids additional simulation CTs and reduces uncertainties from deformable image registration (DIR).
Materials And Methods
A CIRS electron-density phantom was embedded in solid-water slabs and surrounded by bolus to simulate varying patient-scanned volumes. HU-density calibration curves were generated for a standard prostate patient volume (sHU-density) and volume-specific conditions (vHU-density) and imported into Eclipse planning system along with the planning (pCT) calibration curve (pHU-density). Twenty prostate VMAT plans were retrospectively recalculated on registered iCBCT images using these curves. Dosimetric parameters (Dmean, D98%, D2%) and 3D gamma analysis (1%/1 mm, 10% threshold) assessed dose differences. An Alderson Rando phantom study was used to compared pCT-based (pHU-density) and iCBCT-based (sHU-density) calculations.
Results
HU-density curves were stable across scanned volumes, particularly for soft tissues. Dosimetric differences between sHU-density and vHU-density curves were < 0.5%, with gamma pass rates > 99%. Using the pHU-density curve on iCBCT introduced 1% systematic dose deviations. Phantom measurements confirmed strong agreement between pCT- and iCBCT-based calculations using an iCBCT-specific calibration curve.
Conclusion
iCBCT with a dedicated HU-density calibration curve provides dose calculation accuracy comparable to planning CT, supporting reliable offline ART in prostate radiotherapy.
HyperSight has the potential to be an alternative image used for dose calculation directly on HS-CBCT in the ART workflow, particularly in prostate cancer, and demonstrates good image quality and CT number accuracy.
Daily CBCT image-guided couch corrections maintain dosimetric accuracy during hypofractionated prostate radiotherapy and support safe implementation of tight PTV margins in resource-constrained public-sector healthcare settings.
Leanore Legoale-Sebopa, Sheynaz Bassa, H. V. Maselesele· South African Journal of Onc...· 0 citations
Background and purpose This study presents a novel workflow for automated gross tumor volume (GTV) adaptation in online adaptive magnetic resonance imaging (MRI)-guided radiotherapy using promptable segmentation foundation models and evaluates its impact on the quality of adapted treatment plans. Materials and methods...
D. Dudás, T. Blöcker, Miguel Palacios et al.· Physics and Imaging in Radia...· 0 citations
Abstract Background Cone‐beam computed tomography (CBCT) is integral to image‐guided radiation therapy and increasingly used for adaptive radiotherapy and CBCT‐based dose calculation. Conventional kV‐CBCT systems on C‐arm TrueBeam linacs with Feldkamp–Davis–Kress (FDK) reconstruction have historically been limited by r...
Niamh Clarke, J. Hughes, D. Wanigaratne et al.· Journal of Applied Clinical...· 0 citations
Daily anatomical variations can compromise dosimetric plan quality in kidney stereotactic ablative radiotherapy (SABR). The purpose of this study is twofold: (1) To characterize the impact of daily anatomical variations on delivered dose in kidney SABR. (2) To compare linear accelerator (LINAC)-based adaptive radiother...
O. Moncrief, E. Warren, Laurel Z. Larramendi et al.· Medical Dosimetry· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.