Assessing Image Quality Differences Between Single-Energy and Dual-Energy Computed Tomography in Head, Thoracic, and Abdominal Scans: A Systematic Review
Jul 2026· Indian Journal of Radiology and Imaging· 0 citations· 30 references
TL;DR
This article aims to systematically compare image quality between SECT and DECT in head, thoracic, and abdominal CT, focusing on contrast resolution, image noise, artifact reduction, and diagnostic clarity.
Abstract
Background Computed tomography (CT) is an essential imaging modality for evaluating head, thoracic, and abdominal conditions. High-quality images are crucial for accurate diagnosis and clinical decision-making. Single-energy CT (SECT) is widely used but is limited by beam-hardening artifacts and suboptimal contrast resolution, particularly in complex anatomical regions. Dual-energy CT (DECT) acquires data at two energy levels, allowing spectral analysis that can improve image quality. Objective This article aims to systematically compare image quality between SECT and DECT in head, thoracic, and abdominal CT, focusing on contrast resolution, image noise, artifact reduction, and diagnostic clarity. Materials and Methods A structured literature search was conducted to identify studies directly comparing SECT and DECT. Included studies were analyzed for image quality outcomes across different anatomical regions and clinical tasks, considering the influence of acquisition protocols and reconstruction methods. Results DECT generally improved contrast-to-noise ratio and reduced beam-hardening artifacts, especially in complex or contrast-enhanced regions. However, benefits varied by anatomical area and technical implementation. Optimized SECT protocols often achieved comparable image quality, emphasizing the importance of protocol refinement. Conclusion DECT offers potential advantages over SECT in selected applications, but its performance depends on careful optimization of acquisition and reconstruction parameters. Tailored, evidence-based protocol selection is essential to maximize clinical benefits.
Background Computed tomography (CT) plays an important role in diagnosing diseases affecting the head, thorax, abdomen, and other body regions. However, the increasing use of CT has also raiseds concerns regarding radiation exposure to patients. Dual-energy CT (DECT) is a newer imaging technique developed to improve im...
V. K. Maurya, M. Mital· Indian Journal of Radiology...· 0 citations
BACKGROUND
Head and neck computed tomography angiography (CTA) requires high spatial resolution while maintaining acceptable image noise. This study aimed to compare image quality between photon-counting CT (PCCT) and energy-integrating detector CT (EID-CT) and to evaluate PCCT reconstruction strategies for specific cl...
Wei Qiao, Xiao Lu, Wei Zhou et al.· European Journal of Radiolog...· 0 citations
MPR image quality is typically inferior to axial images and deteriorates further when derived from thicker axial slices, therefore, appropriate selection of the reconstruction kernel, pixel size and slice thickness are essential to maintain diagnostic image quality.
P. Monnin, A. Viry, F. Becce et al.· Journal of Applied Clinical...· 0 citations
Background/Objectives: This study aimed to investigate the impact of low-energy virtual monoenergetic imaging (VMI) on quantitative and qualitative image characteristics, as well as its effect on the diagnostic assessability of thoracic arteries in photon-counting computed tomography angiography (CTA). Methods: We retr...
Andreea I. Nica, C. Booz, G. Bucolo et al.· Diagnostics· 0 citations
Purpose: To compare image quality between deep learning reconstruction and hybrid iterative reconstruction for conventional and virtual monoenergetic images derived from the same spectral CT examinations. Methods and Materials: This retrospective, post hoc secondary analysis of a multi-institutional, controlled, blinde...
M. Nazarian, A. Perkins, E. Wong et al.· medRxiv· 0 citations
Electron density provides a significantly better CNR for liver metastases with electron density derived from dual-energy computed tomography compared with both conventional and VMI at 40 keV reconstructions, in both the unenhanced and portal-dominant phases.
C. Chabrier, Maxime Leroy, Pr Olivier Ernst· 0 citations