Dual-energy cone-beam CT improves iodine contrast and Hounsfield unit value stability in the maxillofacial region: a phantom study

Aug 2026 · Biomedical engineering and physics express · Vol 12 · 0 citations · 44 references
Medicine Physics

TL;DR

Although DE-CBCT with VMI and frequency splitting improves HU agreement with CT, slightly improves soft tissue visualization, and reduces artifacts in head and neck phantom imaging, it did not reveal any additional targets compared with routine CBCT.

Abstract

This study was designed to assess the feasibility of dual-energy cone-beam computed tomography (DE-CBCT) and the associated image processing algorithms to improve delayed phase iodine enhancement. An anthropomorphic adult head phantom with iodine inserts in maxillofacial regions was imaged using a clinical CBCT and a helical CT scanner. DE-CBCT was acquired sequentially at 80 and 120 kV, and helical CT at 100 kV (routine clinical protocol), with radiation doses matched within 5%. Virtual monochromatic images (VMIs) at 50 and 200 keV were generated with (VMI+) and without (VMI) noise-optimized frequency splitting algorithm. Evaluation included Hounsfield Unit (HU) agreement with helical CT, iodine contrast, contrast-to-noise ratio (CNR), artifact reduction, and subjective radiologist assessment of target visibility. VMI50 and VMI50+ demonstrated the strongest HU agreement with CT. VMI50+ produced significantly better contrast than CBCT in the maxillary sinus region (mean percentage change 40.7% (15.3%–66.1%); p < 0.05), while improvements were minor in the mandibular region (9.7% (3.0%–16.5%); p < 0.05). VMI50+ significantly increased CNR in the maxillary sinus compared with CBCT (3.3 vs 2.8; p < 0.05), with no significant difference in the mandibular region. Although DE-CBCT increased contrast and reduced artifacts, it did not reveal any additional targets compared with routine CBCT. DE-CBCT with VMI and frequency splitting improves HU agreement with CT, slightly improves soft tissue visualization, and reduces artifacts in head and neck phantom imaging.

Read PDF

Similar papers

Open access Oct 2026

Evaluating Image Quality of Cone Beam Computed Tomography Using HyperSight® on the Varian TrueBeam.

PURPOSE The study evaluated the image quality of the cone beam computed tomography (CBCT) system on Varian TrueBeam linear accelerators with HyperSight (TB-HS) and compared it with the standard TrueBeam system (TB-S). Particular emphasis was placed on changes relevant to offline adaptive radiotherapy (ART). METHODS I...

J. Lohela, S. Rytky, S. Karhula et al. · 0 citations
Sep 2026

Photon-counting detector CT with virtual monoenergetic imaging and iterative metal artifact reduction for spinal fusion implants: optimal keV and combined benefits.

For spinal implants, PCD-CT offers effective approaches for metal artifact and 120 keV is preferred and the combination of iMAR and 120 keV introduce new artifacts and requires careful consideration.

Xin-Ying Huang, Zai-Zhu Zhang, Yong An et al. · 0 citations
Open access Sep 2026

Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter?

This study investigated the kV dependence of VMI CT values in PCD-CT across a wide range of tissue types, keV levels, and kV settings for various phantom sizes, and to determine which kV settings yield the optimal contrast-to-noise ratio (CNR) for a given VMI energy and phantom size.

C. Jeukens, F. M. E. Pinckaers, B. Martens et al. · 0 citations
Open access Sep 2026

CBCT Dose Is Reduced with Angular Tube Current Modulation While Maintaining Image Quality and Reducing Photon Starvation Artifacts.

PURPOSE This study evaluated whether scout-derived angular tube current modulation (ATCM) in cone beam computed tomography (CBCT) can reduce radiation dose while preserving image quality. ATCM is routine in helical CT but remains absent from clinical CBCT and is largely unexplored in the literature. METHODS A project...

Tero Hyvärinen, Samuel Onnela, T. Liimatainen et al. · 0 citations
Preprint Oct 2026

Anatomy-preserving unpaired cone-beam CT refinement for image-guided radiotherapy using pseudo-label guided diffusion

Cone-beam computed tomography (CBCT) is widely used in image-guided radiotherapy, but scatter, beam hardening, noise, truncation, and other artifacts limit image quality and CT number accuracy. Paired CBCT and CT data are difficult to obtain clinically because of motion, anatomical changes, and acquisition mismatch. We...

Qi Lai, Yu-Tong He · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.