Real-world analytical and diagnostic evaluation of the Abbott i-STAT Alinity point-of-care high-sensitivity troponin I assay in a UK emergency department.

Oct 2026 · Annals of Clinical Biochemistry · pp. 45632261499915 · 0 citations
Medicine

TL;DR

The Abbott i-STAT Alinity POCT hs-cTnI assay demonstrated analytical performance and clinical concordance supportive of local implementation, with similar diagnostic discrimination to the Alinity laboratory hs-cTnI assay in a small cohort.

Abstract

Background

High-sensitivity cardiac troponin (hs-cTn) testing is central to emergency department assessment of suspected acute coronary syndrome. Point-of-care hs-cTn testing may reduce turnaround time, but local analytical and clinical evaluation is required before implementation.

Methods

In a single-centre real-world evaluation, paired whole-blood and plasma samples from patients with suspected acute coronary syndrome were analysed using the Abbott i-STAT Alinity point-of-care hs-cTnI assay and the Alinity laboratory hs-cTnI assay. Imprecision, method comparison, threshold concordance, diagnostic discrimination for adjudicated Type 1 myocardial infarction (MI), and potential effects of haemolysis, icterus, lipaemia and haematocrit were assessed.

Results

Low-level imprecision was below 10% CV. Concordance at the <5 ng/L rule-out threshold was high (84.1%); all discordant classifications reflected higher POCT values, with no false-negative rule-outs. Both assays showed similar discrimination for adjudicated Type 1 MI and 100% observed sensitivity and negative predictive value at the 5 ng/L threshold, although confidence intervals were wide. POCT specificity was lower than laboratory specificity. Method comparison showed a modest positive POCT bias, and regression analyses did not identify clinically meaningful effects of haemolysis, icterus, lipaemia or haematocrit.

Conclusion

The Abbott i-STAT Alinity POCT hs-cTnI assay demonstrated analytical performance and clinical concordance supportive of local implementation, with similar diagnostic discrimination to the Alinity laboratory hs-cTnI assay in a small cohort. Larger prospective studies are needed to confirm clinical safety and workflow impact.

View source

Similar papers

Open access Aug 2026

Single High-Sensitivity Cardiac Troponin Measurement to Rule-out/Rule-in Myocardial Infarction in the Emergency Department.

BACKGROUND Point-of-care (POC) testing and triage liaison physicians could help alleviate the impact of emergency department (ED) overcrowding. The aim of our study was to evaluate the diagnostic performance of the TriageTrue POC assay for the rapid rule-out of myocardial infarction based on a single measurement of hig...

A. Dupuy, Yanis Gueche, A. Bargnoux et al. · 0 citations
Oct 2026

A-041 Clinical Performance and Risk Classification Agreement of Point-of-Care i-STAT High-Sensitivity Troponin I Compared with Abbott Alinity hs-cTnI in ED Patients with Suspected Acute Coronary Syndrome

The i-STAT hs-cTnI assay demonstrated high agreement with the Abbott Alinity platform for troponin-based risk classification and correctly identified all myocardial infarction cases, suggesting that the i-STAT assay may be safely used for early risk stratification of ED patients, particularly for identifying low-risk p...

Merih T. Tesfazghi, Elif Yucebay, I. Espinosa et al. · 0 citations
Open access Sep 2026

Real-world impact of implementing a point-of-care high-sensitivity troponin pathway on emergency department flow.

BackgroundPoint-of-care (POC) high-sensitivity cardiac troponin I (hs-cTnI) testing can reduce diagnostic turnaround in the emergency department (ED), but real-world operational data remain mixed.MethodsObservational before-after service evaluation at a high-volume UK tertiary ED (Queen Elizabeth University Hospital, G...

Faheem Ahmad, U. Razzaq, Vaishnavi Muthukrishnan et al. · 0 citations
Oct 2026

A-028 Real-world comparative clinical performance of high-sensitivity cardiac troponin T Gen5 and Gen6 in routine acute care testing

Small analytical differences between generations of high-sensitivity cardiac troponin T (hs-cTnT) assays can shift whether and when results cross clinical decision thresholds, potentially influencing downstream care including diagnoses, serial testing needs, and time to disposition. As health systems consider tra...

Megan Palacio, Nolan R. Martin, Spencer Seely et al. · 0 citations
Review Oct 2026

A-375 Verifying analytical performance of i-STAT high sensitivity troponin cartridges in an urban academic medical center emergency department

Troponin I is a commonly used laboratory test for detection of myocardial infarction. A point-of-care (POC) option for Troponin I allows for quicker clinical decision making, which is particularly useful in the Emergency Department (ED). However, the Abbott contemporary cartilages currently used in this ED for PO...

Alice Ann Lever, Chneriko Thomas, Gwendolyn Griffin-Tyson et al. · 0 citations
Oct 2026

A-006 Diagnostic performance of an investigational high-sensitivity cardiac troponin T and long-T ITC troponin complex assays for acute myocardial infarction in females and males

Assessment of the diagnostic performance of an investigational Mindray CL-1200i hs-cTnT chemiluminescence immunoassay for type 1 MI and type 2 MI compared to standard of care Abbott Alinity hs-cTnI assay and investigational Mindray hs-cTnI assay investigated whether the long-T ITC complex assay can provide additional d...

Joshua R. Miller, Y. Sandoval, Stephen W. Smith et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.