Oct 2026· Photodiagnosis and Photodynamic Therapy· pp.
105679
· 0 citations· 24 references
Medicine
TL;DR
Pentacam and Sirius measurements showed high correlation and good agreement and can provide complementary clinical references, however, systematic biases were observed in the absolute ACD, Kf, Ks, and AST values obtained with the two devices.
Abstract
Purpose
To compare the correlation and agreement between measurements of anterior segment biometric parameters obtained with Pentacam and Sirius in healthy myopic eyes.
Methods
A total of 679 patients with myopia aged 17-49 years (mean ± SD, 23.37 ± 5.93 years) were enrolled, including 431 males (63.5%) and 248 females (36.5%). The spherical equivalent (SE) of the right eye ranged from -0.63 D to -14.75 D (mean ± SD, -4.44 ± 1.84 D). Central corneal thickness (CCT), anterior chamber depth (ACD), flat keratometry (Kf), steep keratometry (Ks), and corneal astigmatism (AST) were measured with each device. Paired t-tests or Wilcoxon signed-rank tests, correlation analysis, intraclass correlation coefficient (ICC) analysis, and Bland-Altman analysis were used to assess correlation and agreement between the two systems.
Results
CCT did not differ significantly between the two devices (p > 0.05), whereas ACD, Kf, Ks, and AST differed significantly (all p < 0.05). Pentacam yielded lower ACD, Kf, and AST values but a higher Ks value than Sirius. All parameters were significantly and positively correlated (r = 0.929-0.984, all p < 0.001). All ICC values exceeded 0.92 (lower limits of the 95% CIs >0.91, all p < 0.001), indicating excellent agreement. Bland-Altman plots showed that the differences in all parameters clustered closely around zero, with narrow 95% limits of agreement.
Conclusion
Pentacam and Sirius measurements showed high correlation and good agreement and can provide complementary clinical references. However, systematic biases were observed in the absolute ACD, Kf, Ks, and AST values obtained with the two devices. Device-specific reference ranges are therefore advisable.
Aims: To compare biometric and topographic measurements obtained with the Aladdin optical biometer and the Pentacam Scheimpflug imaging system in eyes with low and moderate myopia, and to assess agreement between the two devices.Methods: This retrospective study included the right eyes of 58 patients with low or modera...
Elvin Halili Çelenk, Peykan Türkçüoğlu· Journal of Health Sciences a...· 0 citations
To evaluate anterior and posterior corneal surfaces and anterior segment parameters in the Turkish population using Pentacam, and to analyze age-related changes. This retrospective study included healthy participants aged 3–86 years who underwent Pentacam imaging at Dokuz Eylül University Faculty of Medicine. Participa...
Denizcan Özizmirliler, İ. Durak· BMC Ophthalmology· 0 citations
Background High myopia significantly increases the risk of serious ocular complications. The purpose of this study was to characterize ocular biometric variations across different degrees of myopia in young Saudi adults and to identify biometric parameters most strongly influenced by refractive severity. Methods In thi...
S. Alrasheed, Azzam A. Alharbi, N. Challa et al.· PeerJ· 0 citations
Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...
Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al.· Bioengineering· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.