Low baseline Tc-MEP amplitude is associated with false-positive alerts in spine surgery: a ROC-based analysis across spinal pathologies.

Sep 2026 · Journal of clinical monitoring and computing · 0 citations · 15 references
Medicine

TL;DR

Lower baseline Tc-MEP amplitudes were associated with FP alerts and may help identify low-amplitude signals that warrant optimization and cautious interpretation before percentage-based alarm criteria are applied.

Abstract

This study aimed to determine whether absolute baseline transcranial motor evoked potential (Tc-MEP) amplitude is associated with false-positive (FP) alerts during intraoperative neurophysiological monitoring and to estimate pathology-specific cutoff values. We retrospectively reviewed 307 high-risk spine surgery cases monitored with Tc-MEPs from 2018 to 2020. For waveform-level analysis, 1,748 muscle waveforms from 287 FP/true-negative cases were analyzed. Alerts were defined as final amplitude ≤ 30% of baseline, and FP alerts were defined as alerts without new postoperative motor deficit. Receiver operating characteristic analyses were performed to estimate baseline-amplitude cutoff values for FP alerts in spinal cord tumor surgery, spinal deformity surgery, ossification of the posterior longitudinal ligament surgery, and the pooled cohort. FP alerts occurred in 253 of 1,748 waveforms (14.5%). Baseline Tc-MEP amplitude was lower in FP than in true-negative waveforms (351 ± 533 µV vs. 604 ± 1007 µV). FP alerts were more frequent in lower-extremity recordings. Exploratory ROC-derived cutoff values were 122 µV for spinal cord tumor surgery, 180 µV for spinal deformity surgery, 160 µV for ossification of the posterior longitudinal ligament surgery, and 170 µV for the pooled cohort; the corresponding AUCs were 0.68, 0.56, 0.61, and 0.61, respectively. Lower baseline Tc-MEP amplitudes were associated with FP alerts. Baseline amplitude assessment may help identify low-amplitude signals that warrant optimization and cautious interpretation before percentage-based alarm criteria are applied.

Read PDF

Similar papers

Review Open access Sep 2026

Comparative diagnostic performance of motor-evoked potential alarm thresholds for predicting postoperative neurological decline in intradural extramedullary spinal tumor surgery: a systematic review and meta-analysis

Differences in sample size, tumor location, histology, histology, and outcome reporting between threshold groups should be interpreted cautiously because of differences in sample size, tumor location, histology, and outcome reporting.

Łukasz Antkowiak, Marta Antkowiak, A. Kasperczuk et al. · 0 citations
Sep 2026

Diagnostic Performance and Clinical Outcomes of Intraoperative Neuromonitoring in Adolescent Idiopathic Scoliosis: A Comprehensive Analysis of 5,206 Patients.

BACKGROUND Curve-specific risk factors for intraoperative neuromonitoring (IONM) alerts during surgery for adolescent idiopathic scoliosis (AIS) remain unclear. METHODS We identified 5,206 patients with AIS who underwent spinal fusion with multimodal IONM from a prospective database. Alerts were defined as ≥50% ampli...

Travis S. CreveCoeur, Amer F. Samdani, Alexander J. Schüpper et al. · 0 citations
Review Open access Sep 2026

Diagnostic Accuracy of Intraoperative Neurophysiologic Monitoring in Spinal Deformity Surgery: Systematic Review and Meta-Analysis

Study Design Systematic review and diagnostic accuracy meta-analysis. Objective To evaluate the diagnostic accuracy of intraoperative neurophysiologic monitoring (IONM) for detecting postoperative neurologic deficits in spinal deformity surgery and assess performance across monitoring modalities. Methods A systematic r...

B. O'Reilly, Alexandra M. Dunker, Drue M. Domagala et al. · 0 citations
Review Sep 2026

Diagnostic Performance of Pulse Pressure Variation for Predicting Fluid Responsiveness During Spine Surgery in the Prone Position: A Systematic Review and Meta-Analysis.

BACKGROUND Pulse pressure variation (PPV) is widely used to guide fluid responsiveness in mechanically ventilated patients, but its accuracy in the prone position remains uncertain. Prone positioning alters respiratory mechanics and venous return, potentially diminishing the respiratory variation in stroke volume that...

Anshit Goyal, Kiran S. Merchant, E. Deloso et al. · 0 citations
Open access Oct 2026

Non‐Invasive Intracranial Pressure Waveform Assessment as a Screening Test for Abnormal Lumbar Puncture Results

NIWA was feasible and detected IH and CSF changes with high sensitivity among real‐world neurological outpatients and pre‐LP screening with NIWA may enhance diagnostic yield and mitigate complications and healthcare‐related costs.

G. Kubota, Alessandra Santos da Fonseca, C. Hayashi et al. · 0 citations
Open access Aug 2026

Intraoperative neuromonitoring alerts during surgery for congenital spinal deformity: prevalence, risk factors, and outcomes in a prospective study of 100 patients.

BACKGROUND Surgical goals for congenital spinal deformity (CSD) include correction of deformity and prevention of progression without affecting neurological function. CSD has a higher incidence of intraoperative neuromonitoring (IONM) alerts than other scoliosis types and a potential for risk of postoperative neurologi...

Anuj Gupta, Brett A. Shannon, Munish C. Gupta et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.