The dynamics of averaged impedance drop percentage (AID%), a novel ablation parameter, and its association with conduction gaps and lesion characteristics suggest that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric.
Abstract
Background
Real-time monitoring of myocardial tissue response during radiofrequency applications using a mesh-shaped flexible irrigation tip ablation catheter (TactiFlex) remains challenging. This study assessed the dynamics of averaged impedance drop percentage (AID%), a novel ablation parameter, and its association with conduction gaps and lesion characteristics.
Methods
We retrospectively analyzed 1,423 radiofrequency applications in 30 patients undergoing initial pulmonary vein isolation using TactiFlex to examine the relationship between AID% and radiofrequency applications within conduction gap segments. Mechanistic validation was performed using an ex-vivo porcine myocardial model (152 lesions).
Results
Among 1,423 radiofrequency applications, 138 were located within conduction gap segments. AID% was lower in radiofrequency applications in conduction gap segments than in non-conduction gap segments, although the difference was modest (10.2 [8.3-12.2] % vs. 10.9 [9.0-13.1] %, p = 0.002). The overall discriminative performance was limited (area under the curve, 0.58; 95% confidence interval, 0.53-0.63). In mixed-effects logistic regression accounting for within-patient clustering, AID% <10% was not significantly associated with radiofrequency applications in conduction gap segments (odds ratio, 1.560; 95% confidence interval, 0.970-2.500; p = 0.065), whereas average radiofrequency power was significant. In the ex vivo study, lesion characteristics varied according to radiofrequency power, contact force, and duration, even under fixed target AID% conditions.
Conclusions
AID% reflects a composite impedance-based tissue response to radiofrequency delivery that is influenced by multiple ablation parameters. Its limited ability to discriminate conduction-gap segments, together with the variability in lesion characteristics across radiofrequency delivery conditions, suggests that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric. The averaged impedance drop percentage (AID%) reflects a composite impedance-based tissue response to radiofrequency delivery that is influenced by multiple ablation parameters. Its limited ability to discriminate conduction gap segments, together with the variability in lesion characteristics across radiofrequency delivery conditions, suggests that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric.
PURPOSE
Pulsed field ablation (PFA) has become an established treatment modality for atrial fibrillation (AF). Adequate catheter-to-tissue contact is essential for effective energy delivery. However, impedance characteristics associated with directly visualized catheter contact during PFA have not been well characteriz...
Kohki Kimura, Hirotoshi Nagae, Y. Yamamoto et al.· Indian Pacing and Electrophy...· 1 citation
BACKGROUND
The Sphere-9 lattice-tip catheter (Affera, Medtronic) enables delivery of both radiofrequency ablation (RFA) and pulsed field ablation (PFA) for ventricular arrhythmia (VA) ablation. Data on its impact on cardiac implantable electronic devices (CIEDs) are limited.
OBJECTIVES
This study sought to evaluate c...
Ehsan Mahmoodi, A. Younis, K. Higuchi et al.· JACC Clinical Electrophysiol...· 0 citations
BACKGROUND
High-intensity ultrasound (HIU) ablation can produce deep myocardial lesions but has been limited by the directional, side-facing transducer orientation, yielding inconsistent myocardial targeting with lesion efficiencies <75% and transmurality <60%.
OBJECTIVE
We designed HIU catheters and a real-time cont...
Craig C. Morris, Naveed Arang, Phoebe Dijour et al.· Heart Rhythm· 0 citations
INTRODUCTION
Achieving transmural lesions is critical for durable outcomes in catheter ablation. While electrogram amplitude reduction and pacing threshold changes have been reported as useful parameters, the optimal electrode configuration and cutoff values remain unclear. This preclinical study evaluated the efficacy...
I. Kawamura, Aiko Fujimaki, M. Takigawa et al.· Journal of interventional ca...· 0 citations
BACKGROUND
The Dual-Energy THERMOCOOL SMARTTOUCH® SurroundFlow (DE STSF) catheter combines radiofrequency (RF) and pulsed field ablation (PFA) within a single contact force-sensing, irrigated-tip focal platform, fully integrated with the CARTO® 3 three-dimensional mapping system.
METHODS
We reviewed available preclin...
M. Cherbi, J. le Polain de Waroux, M. D. De Smet et al.· Future Cardiology· 0 citations
INTRODUCTION
Catheter ablation is standard of care for cardiac arrhythmias, but conventional catheters are still limited by single-plane steering, variable tissue contact, and high irrigation demands, particularly in complex anatomy. Robotic magnetic navigation (RMN) addresses several of these constraints, but it has l...
Peter Karl Jacobsen, Aashi Mehrotra, C. Jøns et al.· Expert Review of Medical Dev...· 0 citations