Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment.

Aug 2026 · Journal of sport rehabilitation · pp. 1-7 · 0 citations · 29 references
Medicine

TL;DR

3D-printed PhysioCode showed adequate concurrent validity and excellent test-retest reliability for knee extensor/flexor strength, with performance comparable to the MicroFET II in this sample of young, healthy adults, suggesting that PhysioCode may represent an option for strength assessment, warranting further evaluation in target clinical populations.

Abstract

Context

Handheld dynamometry is common for isometric strength testing, but the high cost and import barriers of commercial devices (eg, MicroFET II) limit access. Three-dimensional (3D) printing enables low-cost, customizable instruments. We developed the 3D-printed PhysioCode and examined its measurement properties for knee strength.

Design

Methods and Technical Report evaluating concurrent validity and test-retest reliability versus a reference handheld dynamometer.

Methods

Forty-nine healthy adults (23.2 [3.9] y; body mass index 21.9 [2.8] kg/m2) performed 3 isometric trials of dominant-limb knee extension and flexion with both devices in randomized order, assessed by 2 trained novice raters under blinded analysis. Concurrent validity was evaluated using Pearson correlation coefficients (r) and root mean square error, and agreement between devices was assessed using Bland-Altman analyses. Reliability was assessed using intraclass correlation coefficients (ICC[2,1] and ICC[2,k]), along with the standard error of measurement and minimal detectable change at 95% confidence. We compared single and repeated test strategies.

Results

All mean values differed by ≤21 N between devices across knee extensors and flexors. Concurrent validity was strong-very strong (r = .88-.95; P < .001), with lower root mean square error for averaged repetitions. Bland-Altman showed small, direction-specific bias: PhysioCode slightly overestimated extensor force values (-17 to 21 N) and underestimated flexor force values (-12 N). Reliability was excellent: PhysioCode ICC(2,1) = .93 to .95 and ICC(2,k) = .98 to .98; MicroFET II ICC(2,1) = .89 to .94 and ICC(2,k) = .96 to .98. Averaging 2 to 3 trials reduced standard error of measurement and minimal detectable change at 95% confidence relative to single trials or maxima.

Conclusions

The 3D-printed PhysioCode showed adequate concurrent validity and excellent test-retest reliability for knee extensor/flexor strength, with performance comparable to the MicroFET II in this sample of young, healthy adults. Averaging 2 or 3 repetitions improved precision. These findings suggest that PhysioCode may represent an option for strength assessment, warranting further evaluation in target clinical populations.

View source

Similar papers

Open access Sep 2026

Reliability and measurement error of hip abductor muscle strength using a hand-held dynamometer in healthy adults

[Purpose] Patients with hip osteoarthritis commonly exhibit weakened hip abductor (HA) muscles. Accurate assessment of muscle strength is critical; however, the reliability of HA strength measurements remains underreported. This study aimed to evaluate reliability and measurement error for HA muscle strength values mea...

Yuko Akatsuka, Kenta Kuwahara, Keiji Asada · 0 citations
Open access Sep 2026

Test–Retest Reliability of Knee Proprioceptive Force Sense Assessed Using a Pressure Biofeedback Unit in Male Soccer Players

The aim of this study was to determine the test–retest reliability, measurement error and agreement of a pressure biofeedback-based knee proprioceptive force sense (KPFS) assessment in healthy competitive male soccer players. The study employed a methodological test–retest reliability design. A total of 64 male soccer...

Muhammet Özalp · 0 citations
Open access Sep 2026

The Reliability and Validity of Active Range of Motion Measurement for Fingers Using Three-Dimensional Motion Analysis

The Fugl–Meyer Assessment (FMA) and Range of Motion (ROM) are well-known measures for evaluating finger motor function. However, the finger FMA has a narrow scoring range, and goniometer ROM measurements lack reliability. Therefore, we investigated ROM measurement using a 3D motion capture system (Ultraleap Stereo IR 1...

Saki Moriguchi, Takashi Takebayashi, J. Arata et al. · 0 citations
Open access 2026

Diagnostic accuracy of four handheld dynamometer tests for identifying individuals with hamstring interlimb strength asymmetry

Abstract Introduction: Handheld dynamometer (HHD) strength tests are widely used in clinical settings. The limb symmetry index (LSI) is a key parameter for guiding rehabilitation progression and return-to-sport decisions following hamstring muscle injuries. However, it remains unclear whether HHD-derived results can ac...

Vinícius de Borba Capaverde, F. X. de Lima e Silva, Gabriel dos Santos Oliveira et al. · 0 citations
Open access Oct 2026

The Reliability and Validity of a Novel Isometric Trunk Rotation Strength Test

Background Trunk rotation (TR) strength and power play an important role in overhead sports for both sport-specific performance and injury risk-reduction. However, clinically applicable, reliable and valid methods to measure isometric TR strength have not been previously described. Purpose To assess a) the test-retest...

Kári Árnason, Hildur H. Sigmarsdottir, Rosey Bjorgvinsdottir et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.