Optimization of PET acquisition time for thoracic lesions in 18F-FDG PET/MRI: a retrospective analysis

Oct 2026 · European Radiology Experimental · Vol 10 · 0 citations · 33 references
Medicine

Abstract

This study aimed to define the minimum positron emission tomography (PET) acquisition time for chest PET/magnetic resonance imaging (MRI), providing image quality and lesion detectability comparable to a 20-min reference. In this retrospective study, list-mode data from 101 patients undergoing chest 18F-fluorodeoxyglucose (18F-FDG) PET/MRI on a 3-T PET/MRI system were reconstructed to simulate 10 acquisition times (0.5–20 min). Two readers assessed subjective image quality using a 5-point Likert scale. Quantitative parameters, including maximum, mean of the lesion standardized uptake value (L_SUVmax, L_SUVmean), image noise ratio (INR), and lesion-to-background ratio (LBR), were measured. Lesion detectability was defined as the miss rate relative to lesions visible on the 20-min reference acquisition. In 101 patients, 283 lesions were identified. Subjective scores increased from a median of 2 (interquartile range 2–2) at 0.5 min to 5 (5–5) at 7 min (p < 0.001), then plateaued (7–20 min; all p = 1.000). L_SUVmax, L_SUVmean, and LBR inversely correlated with scan duration (all p < 0.001). L_SUVmean stabilized at 3 min (p = 1.000), while L_SUVmax/LBR stabilized at 5 min (p = 1.000). INR required 10 min (p ≥ 0.054). Lesion miss rates increased time dependently, from 0% at 15 min to 7.77% at 0.5 min. A 7-min acquisition time provided subjective quality and detectability comparable to the 20-min protocol, supporting routine use. For quantitative assessments, 3–5 min sufficed, and patients with shorter 18F-FDG uptake or larger lesions may be adequately assessed with 2 min. Question What is the minimum PET acquisition time for chest 18F-FDG PET/MRI that maintains image quality and lesion detectability equivalent to a 20‑min reference? Findings A 7‑min PET acquisition matches 20‑min image quality and detectability; quantitative parameters stabilize at 3‑5 min; 2 min suffices for patients with shorter uptake or larger lesions. Relevance statement Reducing PET acquisition time to 7 min enhances patient comfort, reduces motion artifacts, and increases scanner throughput without compromising chest PET/MRI diagnostic performance, thereby improving patient experience and workflow efficiency. Further clinical validation is needed. Fluorodeoxyglucose F18, Magnetic resonance imaging, Patient comfort, Positron emission tomography, Thorax

Read PDF

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.