Oct 2026· Biomedical engineering and physics express· Vol 12, pp. 055063· 0 citations· 10 references
MedicinePhysics
TL;DR
Optimize CT acquisition protocols reduced radiation dose while preserving the evaluated objective image quality metrics in the cranial, thoracic, and upper abdominal regions, and was identified as a candidate protocol for subsequent clinical and reader performance validation.
Abstract
Computed tomography (CT) plays an important role in oncologic surveillance after treatment of gynecological malignancies, including patients treated with vaginal electronic brachytherapy. This phantom study evaluated CT acquisition strategies for oncologic surveillance, focusing on radiation dose reduction and objective image quality in the cranial, thoracic, and upper abdominal regions assessed quantitatively. CT acquisition protocols were investigated using an anthropomorphic Kyoto Kagaku phantom. Two acquisition strategies were analyzed: a multi-scan approach with separate acquisitions for the head, neck, thorax, and abdomen-pelvis, and a single-scan head-torso acquisition with region-specific reconstructions. Different exposure parameter settings and reconstruction kernels were evaluated. Objective image quality was assessed using semi-automated region-of-interest measurements with a custom FIJI/ImageJ macro. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and Artifact Index (AI) were calculated for cranial, thoracic, and abdominal regions. For the single-scan protocols, CTDIvol decreased from 4.21 mGy in the standard protocol to 2.97–3.99 mGy in the optimized protocols, while DLP decreased from 410 mGy.cm to 287–388 mGy.cm. Across all anatomical regions, SNR and CNR remained comparable with the standard protocol. Thoracic CNR ranged from 2.68 to 3.01, abdominal CNR remained stable in both liver and aortic regions, and AI demonstrated only moderate variability across the optimized protocols. Optimized CT acquisition protocols reduced radiation dose while preserving the evaluated objective image quality metrics in the cranial, thoracic, and upper abdominal regions. Among the evaluated configurations, a reduced dose single scan head-torso acquisition using reduced tube voltage and tube current modulation showed a favorable balance between radiation dose reduction and objective image quality under the investigated phantom conditions and was identified as a candidate protocol for subsequent clinical and reader performance validation.
BACKGROUND
Head and neck computed tomography angiography (CTA) requires high spatial resolution while maintaining acceptable image noise. This study aimed to compare image quality between photon-counting CT (PCCT) and energy-integrating detector CT (EID-CT) and to evaluate PCCT reconstruction strategies for specific cl...
Wei Qiao, Xiao Lu, Wei Zhou et al.· European Journal of Radiolog...· 0 citations
This study aimed to optimize pediatric computed tomography (CT) protocols for abdomen and chest examinations, focusing on reducing radiation dose without compromising diagnostic image quality. Information was collected from factory protocols, current-modulated protocols, and those used in the routine of a university ho...
Maria Rosane Neczypor, Carla Roberta de Barros Dias, J. V. Real et al.· Brazilian Journal of Radiati...· 0 citations
The motorized attachment for SR3D acquisition reduced image variability and improved image quality, and may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.
David Contella, Chandima Edirisinghe, Kevin Barker et al.· Brachytherapy· 0 citations
Background Computed tomography (CT) plays an important role in diagnosing diseases affecting the head, thorax, abdomen, and other body regions. However, the increasing use of CT has also raiseds concerns regarding radiation exposure to patients. Dual-energy CT (DECT) is a newer imaging technique developed to improve im...
V. K. Maurya, M. Mital· Indian Journal of Radiology...· 0 citations
In PET/CT imaging, CT frequently accounts for more than 50% of the total radiation exposure, even when used solely for attenuation correction and anatomical localisation. This poses a major limitation for dose-sensitive populations, including pregnant women and paediatric patients. This study systematically evalu...
Georgios I. Angelis, Y. Gholami, K. Willowson et al.· EJNMMI Physics· 0 citations
BACKGROUND
Wide-volume acquisition may reduce radiation exposure in pediatric head CT by limiting unnecessary z-axis irradiation, but evidence on 80-row detector systems remains limited. This study compared wide-volume and conventional helical acquisition using a newly fabricated patient-derived anthropomorphic pediatr...
H. Sekkat, Y. Madkouri, O. El Mouden et al.· Journal of Medical Imaging a...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.