TL;DR
Color K-edge imaging is feasible with multiple high-Z elements, whilst outperforming the performance of conventional imaging, paving the way for lung SPCCT functional imaging.
Abstract
Purpose
To assess the feasibility and performance of color K-edge imaging using spectral photon-counting CT (SPCCT) combined with eight high-Z-elements candidates for lung functional imaging.
MATERIAL AND
Methods
Solutions of gadolinium, holmium, ytterbium, hafnium, tantalum, tungsten, gold, and bismuth, were prepared in 12 tubes from 0 to 2 mg/mL, and inserted into an anthropomorphic lung phantom (QRM GmbH). Scans were performed using lung protocols, on a clinical SPCCT (Philips, Israel), at 120 kVp and 3 radiation exposure levels (50, 75, 150 mAs). Noise and contrast-to-noise ratios (CNR) were measured in all images as well as relative percentage errors (RPE). Linear regressions (slope, intercept) of measured versus prepared concentrations and CNR versus concentration were plotted..
Results
On average, mean absolute RPE ranged from 15.9%±8.0% for ytterbium to 34.7%±19.0% for bismuth. Quantification accuracy was excellent for all elements but lower for bismuth. For all elements and exposure levels, mean noise in color K-edge and conventional images was 0.030 ± 0.007 mg/mL and 20.4 ± 2.7 HU, respectively. CNR in color K-edge images increased as the concentrations increased, with an average slope (a) of 43.2 ± 11.8 mL/mg, the highest being for ytterbium (60.1 ± 8.1 mL/mg) and the lowest for bismuth (23.8 ± 5.2 mL/mg). CNR in conventional images also increased as the concentrations increased, with an average slope of 1.7 ± 0.5 mL/mg; the highest was for gadolinium (2.3 ± 0.3 mL/mg) and the lowest for tungsten (1.2 ± 0.1 mL/mg)..
Conclusion
Color K-edge imaging is feasible with multiple high-Z elements, whilst outperforming the performance of conventional imaging, paving the way for lung SPCCT functional imaging.
To determine the best combination of reconstruction parameters on a spectral photon-counting CT (SPCCT) system to improve detection of peritoneal metastases (PMs) on virtual monoenergetic images (VMIs) at 40 keV. A spectral phantom was scanned using SPCCT and dual-energy CT (EID-DECT) at 8.5 mGy (CTDIvol). 40 keV VMIs...
J. Greffier, R. Grange, D. Dabli et al.· European Radiology Experimen...· 0 citations
This study demonstrates the feasibility of DE CBCT technology on the HyperSight imaging system through a novel statistical material specific imaging algorithm and VMI-based CNR improvements and metal artifact reduction.
Bryanna Smith, M. Sattarivand· Physics in Medicine and Biol...· 0 citations
OBJECTIVES
Photon-counting CT (PCCT) enables dose reduction beyond energy-integrating CT, but ultra-low-dose acquisitions may impair assessment of airways and lung parenchyma. We evaluated the impact of ultra-low-dose (ULD) PCCT on quantitative and qualitative lung parameters.
METHODS
In this prospective study, 101 p...
Jonas Kroschke, Lisa Jungblut, Oyunbileg von Stackelberg et al.· European Radiology· 0 citations
Objectives: To evaluate the feasibility of in vivo gadolinium K-edge imaging on a clinical dual-source photon-counting CT (PCCT) system in a porcine model and to assess its performance for dynamic contrast-enhanced imaging, functional biliary imaging, and dual-contrast (bi-phasic) imaging. Materials and Methods: A sing...
M. V. Rybertt, L.-P. Liu, P. Sahbaee et al.· medRxiv· 0 citations
This study investigated the kV dependence of VMI CT values in PCD-CT across a wide range of tissue types, keV levels, and kV settings for various phantom sizes, and to determine which kV settings yield the optimal contrast-to-noise ratio (CNR) for a given VMI energy and phantom size.
C. Jeukens, Floor Pinckaers, B. Martens et al.· Investigative Radiology· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.