Camera-based gait analysis system for a smart rollator
Abstract
Maintaining mobility in old age and in the event of illness is crucial to quality of life. The SmartRoll project is developing smart rollators with technological enhancements and assistive functions to reduce the barriers to independent activity for patients in in-patient care facilities. A key aspect of this is gait analysis, which is used to quantify mobility limitations and track their progression. SmartRoll utilises a camera-based, markerless system on the rollator that combines the advantages of mobile and precise stationary measurement technology. Depth cameras (ZED X mini) and a neural network (YOLOv8spose) are used to detect feet and key points. The AI was trained on a specific dataset which, despite its small size, delivers promising results. Using the speed of the human-rollator system, gait parameters such as stride length (sum of left and right step) and walking speed are calculated. The software architecture is modular and processes raw data through AIsupported feature extraction right through to visualisation. A comparison of the SmartRoll system with the RehaGait reference system showed good agreement in stride length (8% deviation) and walking speed (3% deviation). This underscores the reliability of the approach.