Dual-Echo bSSFP for Rapid T2-Sensitive Quantitative Contrast Imaging at Ultra-Low Field.

Aug 2026 · IEEE transactions on bio-medical engineering · Vol PP · 0 citations
Medicine

TL;DR

The proposed dual-echo bSSFP/bDESS method provides a high-SNR and time-efficient strategy for sequence-specific T2-sensitive contrast characterization at ultra-low field.

Abstract

Objective

Rapid characterization of transverse-relaxation-sensitive MRI contrast is important for evaluating tissue-dependent signal behavior, but remains challenging in ultra-low-field (ULF) MRI because of limited signal-to-noise ratio (SNR) and acquisition-efficiency constraints. This study aims to develop a rapid, high-SNR, sequence-specific T2-sensitive quantitative contrast imaging method for ULF MRI.

Methods

A balanced dual-echo steady-state (bDESS) sequence was developed to acquire two echoes at predefined echo times within each repetition of a balanced steady-state free precession acquisition. A logarithmic-ratio operator, based on a mono-exponential attenuation approximation, was used to derive a sequence-specific T2-sensitive contrast index, termed T2SDI, from the two echo magnitudes. The proposed method was implemented on a custom-built 6.5 mT MRI system and evaluated using numerical simulations, CuSO₄ phantom experiments, and in vivo brain imaging.

Results

Numerical simulations showed that T2SDI exhibited a monotonic and approximately linear dependence on T2 under controlled field-inhomogeneity conditions. Phantom experiments confirmed that bDESS-derived T2SDI increased with CPMG-measured reference T2 and showed higher SNR than dual-echo SPGR. The method was further demonstrated in vivo by generating T2SDI maps of the human brain.

Conclusion

This study presents a sequence-specific, index-based method for rapid T2-sensitive quantitative contrast imaging in ULF MRI.

Significance

The proposed dual-echo bSSFP/bDESS method provides a high-SNR and time-efficient strategy for sequence-specific T2-sensitive contrast characterization at ultra-low field.

View source

Similar papers

Aug 2026

Evaluation of k‐Space Acquisition Strategies in Nonuniform B0 (NuBo) MRI

To evaluate k‐space acquisition strategies for magnetic resonance imaging (MRI) in a nonuniform B0 (NuBo) field‐cycling system, focusing on image quality, scan‐time efficiency, robustness to B0 inhomogeneity, and RF coil bandwidth constraints.

Y. Ha, Chenhao Sun, Anja Samardzija et al. · 0 citations
Open access Aug 2026

Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement

CS-MPRAGE provides high-quality 3D images and reliable volume data with significantly reduced acquisition time and comparable image quality by comparing its scan time and image quality with standard MPRAGE.

Jung-bin Lee, Hye Shin Ahn, Leehi Joo et al. · 0 citations
Open access Aug 2026

OPTIMIZING ECHO TIME AND REPETITION TIME FOR TISSUE CONTRAST USING DIGITAL PHANTOMS

Digital phantom simulations are leveraged to systematically evaluate the effect of varying TE and TR on tissue contrast across common anatomical structures, and validate known MRI contrast behaviors and provide scanner-agnostic TE/TR recommendations to accommodate practical clinical constraints.

Manvee Rai, Nisha Jhinkwan, Nidhi Goswami et al. · 0 citations
Aug 2026

Single-shot T2-Weighted Turbo Spin-echo Imaging of the Liver Improves Lesion Characterization Compared with Free-breathing Acquisition.

SS-TSE, independent of DLR, improved T2-weighted ancillary feature assessment and LI-RADS categorization of HCC over FB-TSE via a 91.8% reduction in acquisition time; DLR did not alter these outcomes but improved image quality and lesion conspicuity.

Kumi Ozaki, Shoya Haseyama, Eri Sugioka et al. · 0 citations
Open access Sep 2026

Fast and robust T1 mapping based on a 3D dual‐echo UTE sequence (PETALUTE) for superparamagnetic iron oxide nanoparticle (SPION) biodistribution assessment

A fast, B1 +‐corrected T1 mapping protocol based on PETALUTE (a 3D dual‐echo ultrashort echo time MRI sequence with a petal‐like rosette k‐space trajectory) using a variable flip‐angle (VFA) acquisition for T1 mapping to assess ferumoxytol distribution is developed.

Zhen Jiang, S. Sawiak, Alexandra Lipka et al. · 0 citations
Aug 2026

Evaluation of acquisition strategies in IVIM-DKI brain imaging: Effects on image quality, parameter stability, and comparison with DSC perfusion.

An optimized low b-value-weighted acquisition with targeted averaging (A2) provided the strongest overall balance between image quality, parameter stability, tissue contrast, and correspondence with DSC-derived perfusion measures.

A. S., S. Pendem, R. Kadavigere et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.