Non-Contact Heart Rate Estimation From Carotid Pulse Using Ultra-Wideband mmWave Loaded-Slot Antenna

2026 · IEEE Access · Vol 14, pp. 169709-169723 · 0 citations · 57 references

Abstract

Non-contact carotid pulse monitoring is a promising alternative to conventional contact-based methods by enabling continuous remote monitoring while improving patient comfort, reducing infection risk, and supporting telehealth applications. On the other hand, reliable contact-free detection of the carotid pulse remains challenging due to the extremely small mechanical displacements of the neck surface and susceptibility to motion artifacts. Existing non-contact techniques, including camera-based optical systems, radar, and laser vibrometers, either require controlled illumination or lack sufficient robustness to patient motion. These limitations highlight the need for a compact, contact-free sensing approach suitable for both clinical and remote monitoring settings. This work introduces a novel non-contact carotid pulse detection and heart-rate estimation sensing platform and methodology using an ultra-wideband millimeter-wave loaded-slot antenna. The antenna operates at 22–29 GHz with a maximum gain of approximately 7 dBi, enabling detection of periodic variations in the complex reflection coefficient (S11) induced by arterial motion. During experimental validation on 22 volunteers (males, mean age 27), the antenna was placed 10 mm from the neck, and the temporal evolution of S11 was recorded using a vector network analyzer. The measured signal was processed with band-pass filtering and envelope extraction to isolate the pulse waveform, which was then compared against a commercial photoplethysmography (PPG) sensor. The results show agreement between the proposed method and the PPG reference, with only a small deviation in heart rate estimation across subjects. In comparison with camera- and radar-based non-contact methods, the proposed antenna-based sensor achieves similar accuracy while eliminating the need for optical illumination and bulky hardware. These findings highlight the feasibility of contact-free carotid pulse monitoring in wearable and point-of-care settings, enabling unobtrusive cardiovascular monitoring in clinical scenarios, such as emergency assessment, continuous bedside monitoring, and remote patient care, where rapid, contact-free pulse detection is advantageous.

Read PDF

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.