Fundamental Performance Evaluation of Pulsed ToF and FMCW LiDAR for the Development of Measurement-Principle-Specific Standardized Test
Abstract
Reproducible LiDAR performance evaluation requires not only common test environments but also evaluation metrics and analysis conditions that reflect different measurement principles and sensor output characteristics. This study investigates pulsed time-of-flight (ToF) and frequency-modulated continuous-wave (FMCW) LiDAR to identify fundamental evaluation items for measurement-principle-specific standardized test procedures. For pulsed ToF LiDAR, axial displacement tracking, distance measurement, and spatial separation between adjacent targets were evaluated. Modification of the data-processing conditions reduced the mean relative displacement errors along the X, Y, and Z axes from 6.67%, 5.14%, and 6.90% to 0.46%, 0.60%, and 0.60%, respectively. An independent repeated-motion test also showed a difference between LiDAR-derived displacement and the physical reference, indicating the need to distinguish tracking consistency from reference-based accuracy. For FMCW LiDAR, range and velocity characteristics were analyzed according to analysis-region size, measurement mode, and target velocity. Increasing the analysis region reduced frame to frame range noise from 0.468mm to 0.114mm, while valid velocity output rate and velocity variation differed between the 0D and Scan modes. These results indicate that standardized LiDAR testing should define common test conditions while applying measurement-principle-specific performance metrics and analysis procedures.