Long-term analytical performance of apixaban measurements based on external quality assessment data.

Aug 2026 · Thrombosis Research · Vol 264, pp. 109795 · 0 citations · 17 references
Medicine

TL;DR

Apixaban assays demonstrate good long-term precision despite methodological heterogeneity, and uniform platforms showed significantly lower LCVa and ACD than partially or non-uniform platforms.

Abstract

Objectives

Although direct oral anticoagulants (DOACs) do not usually require monitoring and clinical cut-off values are not well established yet, measurement of apixaban is often requested in emergency settings and may be requested for patients with suspected variable pharmacokinetics. For safe interpretation, results must be comparable between laboratories and platforms. Most external quality assessment (EQA) schemes, however, only provide short-term performance data. This study aimed to assess the long-term analytical performance of apixaban assays using multicenter EQA data and to identify sources of analytical variation.

Methods

Anonymized data from the ECAT program were analyzed. Twice yearly, two lyophilized citrate plasma samples (apixaban range 24-413 ng/mL) were distributed. Consensus values were established, and linear regression of individual laboratory results versus consensus values was performed to calculate slope, intercept, and standard error. Total error, random error, bias (constant and proportional), long-term coefficient of variation (LCVa), and analytical critical difference (ACD) were determined. The effect of platform uniformity was also evaluated.

Results

Data from 233 laboratories representing 238 method-reagent-calibrator combinations met inclusion criteria. Consensus coefficients of variation were < 10% above 50 ng/mL but increased at lower concentrations, reaching 24.6%. Mean total error was 16.8% and was mainly driven by random error (10.7%). Mean total bias was 33.1 ng/mL, predominantly proportional. The median LCVa was 5% (range 1.7-19.9%), and the median ACD was 34.2 ng/mL. Uniform platforms (same-manufacturer equipment, method, and calibrator) showed significantly lower LCVa and ACD than partially or non-uniform platforms.

Conclusions

Apixaban assays demonstrate good long-term precision despite methodological heterogeneity. Proportional bias is the primary source of deviation. Platform harmonization improves reproducibility.

Read PDF

Similar papers

Open access Sep 2026

Long-Term Monitoring of Everolimus Concentrations in Routine Clinical Practice: Evaluation of Patient Percentiles as Quality Indicator and Impact of Analytical Method Change

Everolimus concentration distributions were stable over time and showed no meaningful sex-related or seasonal variation, and may serve as a practical supplementary quality indicator for TDM assays.

A. Larsson, Anna-Karin Hamberg, M. Eriksson · 0 citations
Open access Aug 2026

Same sample, different result: Variation in clozapine assay measurement in four UK laboratories

Background Laboratory measurement of clozapine blood concentrations is central to safe and effective treatment and is commonly assumed to be accurate and reproducible across providers. Chromatographic assays require batch-specific setup and are performed using different equipment and techniques, which may introduce err...

M. Atkins, J. Fritz, M. Hilaire et al. · 0 citations
Open access Aug 2026

Validation of point-of-care INR testing against standard laboratory methods in a Malaysian primary healthcare setting

The study concludes that the Wondfo POCT device demonstrates acceptable agreement with standard laboratory testing within therapeutic ranges, thereby offering a viable alternative for rapid, decentralized INR monitoring, particularly in outpatient care, and can potentially improve patient management and safety.

Rosline Hassan, S. Iberahim, S. S. Mohd Yusoff et al. · 0 citations
Open access Aug 2026

Analytical Performance Specifications Derived from External Quality Assessment Data: Comparison with Biological Variation-Based Models for Intermediate Precision Assessment.

APSs derived from pooled EQA peer group data provide realistic and technically achievable intermediate precision targets consistent with current AP and thus can complement BV-based specifications.

C. Ilardo, Emmanuel Reynaud, Nathalie Benaily · 0 citations
Open access Sep 2026

Development and validation of a pharmacokinetic tool to convert random topiramate concentrations to standardized trough levels in bipolar disorder

This preliminary PK conversion tool reliably estimates the standardized trough concentration of immediate-release topiramate from random TDM levels and could potentially aid safety monitoring by improving the interpretation of non-standard samples, though its clinical utility requires prospective external validation an...

Xuan Jv, Shu-Zhen Li, Ye Liu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.