TL;DR
This study provides an updated SoA database for precision that is of particular importance in circumstances where EFLM Biological Variation Database-derived APS do not exist, cannot be generated with current methods, or are inappropriate for pathophysiologic states.
Abstract
Background
The European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) has established a three-stage model of criteria to set analytical performance specifications (APS). The third model is based on state-of-the-art (SoA), which has been described as the highest level of analytical performance with currently available methods. However, an up-to-date, comprehensive, easily usable SoA database does not currently exist.
Methods
Data mining was performed using a data management package (Unity Real Time) for Internal quality control (IQC) and interlaboratory comparisons (ILC). For each analyte, the coefficient of variation (CV) values reported by the 50th, 90th, and 95th percentiles of laboratories were extracted and then the median intralaboratory CV50, CV90, and CV95 values were calculated.
Results
Using >775 million quality control (QC) measurements, an updated SoA database includes >100 analytes commonly performed in the fields of clinical biochemistry, hematology, hemostasis, therapeutic drug monitoring (TDM), and serology. The number of QC measurements of each analysis, for 3 specific levels of concentration, ranged from 1090 to 9 163 824 to evaluate the impact of concentration level on the method precision for each analysis.
Conclusions
This study provides an updated SoA database for precision that is of particular importance in circumstances where EFLM Biological Variation Database-derived APS do not exist, cannot be generated with current methods, or are inappropriate for pathophysiologic states. The impact of concentration level on precision data is also reported.
One of the foundations of medical decision-making is laboratory diagnostics, an interpretive field of science. The interpretation of results requires the definition of normal and pathological values based on reference intervals. The modern framework of reference intervals has gradually evolved, in parallel with the dev...
É. Papp, Barna Vásárhelyi, Orsolya Márta Láng· Orvosi Hetilap· 0 citations
Abstract Objectives To evaluate the analytical performance of a novel fully-automated molecular diagnostics system Molecision® R8, together with its multiplex testing capability and compatibility with different types of analytes and samples. Methods Precision, limit of detection (LoD) and linearity were evaluated based...
Ya-Ping Ren, Chao Chen, Wei-Zhong Wang et al.· Clinical Chemistry and Labor...· 0 citations
The College of American Pathologists (CAP) recommends estimating measurement uncertainty (MU) during validation (typically by analyzing replicates of quality control [QC] material over 5 days); and re-evaluating MU estimates in response to a clinical inquiry (typically by a top-down approach using extended length...
Vivek Kumar, A. Gallegos· Clinical Chemistry· 0 citations
Risk-based internal quality control (IQC) has been increasingly adopted in clinical laboratories to enhance analytical quality and patient safety. ISO 15189:2022 recommends designing IQC plans based on risk assessment rather than applying fixed QC frequencies. Sigma metrics combined with the QC Constellation tool...
Practical implementation of patient-based real-time quality control (PBRTQC) requires analyte- and institution-specific optimization, particularly in tertiary hospital settings characterized by substantial patient heterogeneity and case-mix variability. This study evaluated the operational feasibility of average-...
Se-eun Koo, H. Park· Clinical Chemistry· 0 citations
Biological variation data (BVD), which describes within-subject (CVI) and between-subject (CVG) variability of laboratory measurands, are essential reference data in laboratory medicine. BVD are used to define analytical performance specifications (APS) for imprecision, bias, maximum allowable measurement uncerta...
Kornelia D. Galior, William A. Bartlett, Sverre Sandberg et al.· Clinical Chemistry· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.