An Update on State-of-the-Art Performance for Precision in Medical Laboratories.

Aug 2026 · The Journal of Applied Laboratory Medicine · 0 citations
Medicine

TL;DR

This study provides an updated SoA database for precision that is of particular importance in circumstances where EFLM Biological Variation Database-derived APS do not exist, cannot be generated with current methods, or are inappropriate for pathophysiologic states.

Abstract

Background

The European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) has established a three-stage model of criteria to set analytical performance specifications (APS). The third model is based on state-of-the-art (SoA), which has been described as the highest level of analytical performance with currently available methods. However, an up-to-date, comprehensive, easily usable SoA database does not currently exist.

Methods

Data mining was performed using a data management package (Unity Real Time) for Internal quality control (IQC) and interlaboratory comparisons (ILC). For each analyte, the coefficient of variation (CV) values reported by the 50th, 90th, and 95th percentiles of laboratories were extracted and then the median intralaboratory CV50, CV90, and CV95 values were calculated.

Results

Using >775 million quality control (QC) measurements, an updated SoA database includes >100 analytes commonly performed in the fields of clinical biochemistry, hematology, hemostasis, therapeutic drug monitoring (TDM), and serology. The number of QC measurements of each analysis, for 3 specific levels of concentration, ranged from 1090 to 9 163 824 to evaluate the impact of concentration level on the method precision for each analysis.

Conclusions

This study provides an updated SoA database for precision that is of particular importance in circumstances where EFLM Biological Variation Database-derived APS do not exist, cannot be generated with current methods, or are inappropriate for pathophysiologic states. The impact of concentration level on precision data is also reported.

Read PDF

Similar papers

Review Open access Sep 2026

[What is (not) considered normal? - Contemporary challenges of establishing laboratory reference intervals].

One of the foundations of medical decision-making is laboratory diagnostics, an interpretive field of science. The interpretation of results requires the definition of normal and pathological values based on reference intervals. The modern framework of reference intervals has gradually evolved, in parallel with the dev...

É. Papp, Barna Vásárhelyi, Orsolya Márta Láng · 0 citations
Sep 2026

Multicenter evaluation of an innovative fully-automated molecular diagnostics system

Abstract Objectives To evaluate the analytical performance of a novel fully-automated molecular diagnostics system Molecision® R8, together with its multiplex testing capability and compatibility with different types of analytes and samples. Methods Precision, limit of detection (LoD) and linearity were evaluated based...

Ya-Ping Ren, Chao Chen, Wei-Zhong Wang et al. · 0 citations
Oct 2026

B-093 Establishing measurement uncertainty for tumor markers: impact of analytical variation in quality control material vs patient serum matrix

The College of American Pathologists (CAP) recommends estimating measurement uncertainty (MU) during validation (typically by analyzing replicates of quality control [QC] material over 5 days); and re-evaluating MU estimates in response to a clinical inquiry (typically by a top-down approach using extended length...

Vivek Kumar, A. Gallegos · 0 citations
Oct 2026

A-307 Determining risk-based internal quality control frequency for biochemical testing using Sigma metrics and the QC Constellation tool

Risk-based internal quality control (IQC) has been increasingly adopted in clinical laboratories to enhance analytical quality and patient safety. ISO 15189:2022 recommends designing IQC plans based on risk assessment rather than applying fixed QC frequencies. Sigma metrics combined with the QC Constellation tool...

S. Nguyễn · 0 citations
Oct 2026

B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laborator

Practical implementation of patient-based real-time quality control (PBRTQC) requires analyte- and institution-specific optimization, particularly in tertiary hospital settings characterized by substantial patient heterogeneity and case-mix variability. This study evaluated the operational feasibility of average-...

Se-eun Koo, H. Park · 0 citations
Review Oct 2026

A-203 Biological variation reference data; EFLM initiatives and developments to address data quality issues and deliver reliable data.

Biological variation data (BVD), which describes within-subject (CVI) and between-subject (CVG) variability of laboratory measurands, are essential reference data in laboratory medicine. BVD are used to define analytical performance specifications (APS) for imprecision, bias, maximum allowable measurement uncerta...

Kornelia D. Galior, William A. Bartlett, Sverre Sandberg et al. · 0 citations