Transcatheter Mitral Valve Replacement in an Incomplete Ring With LAMPOON Facilitated By Aortic Prosthesis.
Abstract
Background
Transcatheter mitral valve replacement in large incomplete flexible mitral bands is high risk because limited anchoring and anterior leaflet displacement may cause embolization or left ventricular outflow tract (LVOT) obstruction. CASE SUMMARY A 54-year-old man with ischemic cardiomyopathy, prior coronary artery bypass grafting, aortic valve-in-valve transcatheter aortic valve replacement, mitral repair with a 31-mm Duran AnCore band, and tricuspid annuloplasty presented with NYHA functional class IV heart failure from severe recurrent mitral regurgitation. Computed tomography excluded SAPIEN M3 (predicted neo-LVOT 0 mm2) and showed a long anterior mitral leaflet, raising concern for dynamic obstruction despite a predicted neo-LVOT of 347.6 mm2 for a 29-mm balloon-expandable valve. He underwent antegrade base-to-tip laceration of the anterior mitral leaflet to prevent outflow obstruction, followed by valve-in-ring transcatheter mitral valve replacement, with no paravalvular leak or significant LVOT gradient.
Discussion
Anterior aortic prosthetic anatomy facilitated valve anchoring despite the incomplete flexible band, which would have been prohibitive in the absence of anterior bioprosthesis. TAKE-HOME MESSAGE Adjacent prosthetic anatomy may be leveraged in selected complex valve-in-ring procedures after meticulous multimodality planning and procedural expertise.