Comparison of MALDI-TOF MS and VITEK 2 performance in the identification of brucella strains isolated from blood culture bottles; retrospective evaluation of antibiotic resistance profiles, clinical manifestations, and serological follow-up.

Nov 2026 · Diagnostic microbiology and infectious disease · Vol 117 2, pp. 117744 · 0 citations · 16 references
Medicine

Abstract

INTRODUCTION AND

Objectives

Brucella species cause severe zoonotic infections with an exceptionally high risk of laboratory-acquired transmission. We systematically evaluated and compared the identification performances of two contemporary MALDI-TOF MS platforms and the VITEK 2 automated system, analyzed serial standard tube agglutination (STA) titer kinetics, and determined in vitro antimicrobial susceptibility profiles of clinical blood culture isolates.

Materials And Methods

Twenty-five culture-confirmed Brucella isolates recovered from peripheral blood culture bottles between 2019 and 2025 were evaluated using the MALDI Biotyper Sirius (routine MBT Compass library), VITEK MS Prime (IVD database v3.3), and VITEK 2 GN automated identification cards. All sample manipulation and ethanol/formic acid extraction were performed inside a certified Class II Biosafety Cabinet ensuring complete inactivation prior to mass spectrometry. Antimicrobial susceptibility was determined by the disk diffusion method on Mueller-Hinton Fastidious (MH-F) agar according to EUCAST (v. 16.0) guidelines. Clinical data and serial STA titers were retrospectively analyzed.

Results

The MALDI Biotyper Sirius accurately identified all 25 isolates (100%) at the species level as Brucella melitensis, whereas the VITEK MS Prime identified all (100%) at the genus level (Brucella spp.). In contrast, the VITEK 2 GN automated system demonstrated markedly poor diagnostic utility, correctly identifying 92% (23/25) of isolates, failing to identify one isolate ('unidentified'), and misidentifying another blood culture isolate as Roseomonas gilardii. A statistically significant decrease in Wright tube agglutination titers was observed following treatment (median initial titer 1:320 [IQR, 1:160-1:640] vs. median final titer 1:80 [IQR, 1:40-1:160]; p = 0.007). Serological reductions did not differ significantly across therapeutic regimens (p = 0.3219). All isolates (100%) were fully susceptible to rifampin, doxycycline and gentamicin. Resistance rates to trimethoprim-sulfamethoxazole (SXT) and ciprofloxacin were 40% (10/25) and 24% (6/25), respectively. Favorable clinical cure without relapse was achieved in patients receiving SXT-containing regimens despite in vitro resistance.

Conclusions

The MALDI Biotyper Sirius provides superior diagnostic speed and biosafety via reliable species-level identification using routine spectral libraries. Automated phenotypic misidentifications (e.g., R. gilardii) represent critical biohazards that warrant urgent MALDI-TOF MS verification. High in vitro resistance to SXT highlights potential disk diffusion overestimation in fastidious organisms, emphasizing the necessity of broth microdilution MIC confirmation before precluding vital oral treatment options.