MRI-based assessment of focused ultrasound parameters on boiling histotripsy lesions in agar-based phantoms.
Abstract
Purpose
Histotripsy is a noninvasive focused ultrasound (FUS) technique that fractionates tissue through mechanical effects. This study presents a systematic magnetic resonance imaging (MRI)-based dose-response characterization of the effects of key boiling histotripsy sonication parameters, including acoustic power, pulse count, duty cycle, and combined pulse repetition frequency (PRF)/pulse duration conditions, on mechanical lesion formation in agar-based phantoms.
Methods
Exposures were performed on agar phantoms using a 2.6 MHz single-element transducer (f-number 1.3), and T2-Weighted (T2-W) images were acquired post-hoc to visualize the formed lesions.
Results
Lesions appeared as hyperintense regions bordered by hypointense outlines on T2-W images, enabling quantification of lesion size. Acoustic power thresholds of 170 W (corresponding to simulated P+=90.8 MPa, P-=-15.7 MPa, and As = 94.8 MPa) were required for phantom erosion using 20 ms long pulses at 1 Hz PRF and a 2 % duty cycle, with increasing power levels between 170-215 W (simulated P+=90.8-96.8 MPa, P-=-15.7 to -18.3 MPa, As = 94.8-97.1 MPa) resulting in larger lesion dimensions. At the highest acoustic power, pulse counts of 300 or more were necessary for phantom fractionation. Although larger lesion dimensions were observed at higher pulse counts, with a non-significant trend toward a plateau above 500 pulses, these observations require confirmation in larger studies due to limited repetitions (n = 2). Duty cycles above 1 % applied at 1 Hz PRF using varied pulse lengths up to 50 ms, successfully induced lesions, with higher duty cycles between 2-5 % increasing lesion length. Combined variations in PRF (1-2 Hz) and pulse duration (10-20 ms) at a 2 % duty cycle required fewer pulses for lesion formation at lower PRFs and longer pulse lengths.
Conclusions
The study highlights the sensitivity of T2-W MRI for characterizing boiling histotripsy outcomes in phantom models, and provides valuable insights for optimizing sonication parameters under in-vitro conditions.