Abstract
Objective
To assess the feasibility of performing quantitative in vivo tissue sodium concentration measurements in the skin with MRI using a sodium-tuned single-loop transmit/receive surface coil at 3 T.
Materials And Methods
Proton and sodium imaging of skin at the lower back was performed in 17 healthy volunteers on a clinical 3 T system using an anisotropic (1 x 1 x 10 mm) 3D cones sequence. B1 mapping via the double-angle method was used to account for spatial variation in transmit and receive sensitivity. Apparent tissue sodium concentration (aTSC) was estimated relative to a concurrently imaged phantom of known concentration.
Results
The mean and standard deviation of B1-corrected aTSC across all subjects were 28 ± 7 mM, approximately 63% higher than previously published studies at 3 T. Although the study was not originally powered to detect gender or racial differences, a higher aTSC was observed in males relative to females (31 ± 5 mM vs. 23 ± 7 mM, distribution 11M/6F).
Discussion
The skin aTSC values recorded were slightly higher than in some previous studies, which possibly arose from the higher spatial resolution and coil sensitivity utilised here. The recorded values remain lower than those measured at 7 T or via direct tissue sampling. Sex-linked differences in 23Na-MRI-derived aTSC are consistent with biopsy results described in the literature. A scan duration of ~40 minutes was considered clinically acceptable.
OBJECTIVE
While most sodium (23Na) MRI studies focus on apparent tissue sodium concentration (aTSC), longitudinal (T₁) and transverse (T₂*) relaxation times may provide additional microstructural and physiological information. This study aimed to establish reference values for 23Na T₁ and bi-exponential T₂* in human lo...
Nour Zeitouni, Laurent Ruck, L. Gast et al.· MAGMA· 0 citations
Optimized and phantom calibrated MOLLI enabled rapid measurement of human ex vivo blood T1 at 5 T, and blood T1 at 5 T was higher than commonly reported adult values at 3 T and varied with both Hct and Y.
Shu-Wan Yu, Yuan Dong, Yu-Xiao Gu et al.· Magnetic Resonance in Medici...· 0 citations
OBJECTIVE
To quantify the reproducibility of knee articular cartilage T2 across a wide range of MR systems.
MATERIALS AND METHODS
The reproducibility of T2 relaxation time mapping was assessed by measuring the knee of a volunteer (n1.5 T = 22, n3T = 13) and a System Standard 130 phantom (n1.5 T = 19, n3T = 12) in 38...
H. Leskinen, Nea Pulkkinen, Antti Aarnio et al.· MAGMA· 0 citations
The wireless metasurface coil was more convenient and provided image quality and diagnostic performance comparable to the 12-channel commercial knee coil at 1.5 T.
Jie Gao, Xin-Xin Li, Zhong-Hai Chi et al.· Journal of Magnetic Resonanc...· 1 citation
Evaluated retrospective correction methods for spectral linewidth, signal-to-noise ratio, similarity, and metabolite quantification indicated that each provides practical benefits in improving the spectral linewidth, SNR, and similarity.
Chu-Yu Lee, Jia Xu, Bao-Lian Yang et al.· bioRxiv· 0 citations
The growing prevalence of neurodegenerative diseases has increased the number of patients treated with deep brain stimulation (DBS), many of whom will require magnetic resonance imaging (MRI) during their care. However, radiofrequency (RF)-induced heating of conductive DBS leads during MRI remains a primary safety conc...
A. Louka, Diego F. Martinez, Kat Zaraska et al.· NeuroImage· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.