Jul 2026· American Journal of Cardiology· Vol 275, pp. 27-30· 0 citations· 10 references
Medicine
TL;DR
The Vectorial Angle is a novel, independent pre-procedural predictor of PPM after balloon-expandable TAVI, highlighting that the spatial orientation of the valve's maximal radial force relative to the conduction tissue is a potential anatomical determinant associated with conduction-system vulnerability, complementing conventional anatomical assessments such as MSL.
Abstract
Background
Conduction disturbances requiring permanent pacemaker (PPM) implantation remain a frequent complication of transcatheter aortic valve implantation (TAVI). This study aimed to evaluate the "Vectorial Angle"-a novel computed tomography (CT)-derived biomechanical parameter-alongside traditional anatomical and electrocardiographic predictors of PPM following balloon-expandable TAVI
Methods
This retrospective cohort study included 67 patients with severe symptomatic aortic stenosis undergoing TAVI with the balloon-expandable Myval system. The Vectorial Angle was defined on pre-procedural CT at the elliptical left ventricular outflow tract (e-LVOT) level as the angular deviation between the minor axis (representing the axis of maximal radial expansion force) and the midpoint of the right coronary to non-coronary (RCC-NCC) commissure (approximating the anatomical course of the conduction system).
Results
PPM implantation was required in 12 patients (17.9%). In multivariate logistic regression analysis, a narrower Vectorial Angle (OR 0.935 per 1° increment, 95% CI 0.889-0.982; p=0.007) and a shorter membranous septal length (MSL) (OR 0.447 per 1 mm increment, 95% CI 0.241-0.831; p=0.011) emerged as the only independent predictors of PPM requirement. Valve oversizing index was not associated with PPM implantation (p=0.954). Implantation depth was numerically greater in the PPM group but did not reach independent predictor status (p=0.079). Pre-existing right bundle branch block (RBBB) lost independent predictive value after adjustment for these structural parameters. Receiver operating characteristic (ROC) curve analysis for the Vectorial Angle yielded an area under the curve of 0.806 (p=0.001), with an optimal cutoff of ≤52°.
Conclusions
The Vectorial Angle is a novel, independent pre-procedural predictor of PPM after balloon-expandable TAVI. This finding highlights that the spatial orientation of the valve's maximal radial force relative to the conduction tissue is a potential anatomical determinant associated with conduction-system vulnerability, complementing conventional anatomical assessments such as MSL.
BACKGROUND
The Evolut and the Navitor transcatheter heart valves (THVs) represent the most widely used self-expanding implanted valves. Despite a similar implant concept, they markedly differ in design and hemodynamic performance. In both THVs, conduction disturbances following transcatheter aortic valve replacement (T...
Enrico Poletti, O. Oliva, Ahmad Hassan et al.· Catheterization and cardiova...· 0 citations
Introduction Transcatheter heart valve implantation into a failed surgical heart valve (THV-in-SHV) is an established reintervention strategy for degenerated aortic bioprostheses; however, early procedural outcomes, discharge hemodynamics, and standardized safety remain clinically relevant in routine practice. Material...
R. Sobczyński, J. Trębacz, Paweł Kleczyński et al.· Cardiology Journal· 0 citations
Among patients treated with newer-generation transcatheter valves, SEV demonstrated more favorable hemodynamic performance compared with BEV, however, short- and mid-term mortality rates remained similar between the two groups.
M. Baudo, S. Sicouri, Dimitrios E. Magouliotis et al.· Journal of Cardiovascular Me...· 1 citation
BACKGROUND
Data on early outcomes of the novel balloon-expandable (BE) Octapro transcatheter heart valve (THV) in real-world practice are limited. We sought to assess procedural safety and early outcomes of the Octapro in patients with severe native aortic stenosis (AS) undergoing transcatheter aortic valve implantatio...
A. Ielasi, E. Moscarella, Gregor Verček et al.· International Journal of Car...· 0 citations
Background: Conduction disturbances and the subsequent need for permanent pacemaker implantation (PPMI) remain a major limitation of transcatheter aortic valve replacement (TAVR) with self-expanding transcatheter heart valves (SE-THVs). The cusp-overlap technique (COT) has been proposed to optimize implantation depth a...
A. Ispas, D. Brunet, R. Saady et al.· Journal of Clinical Medicine· 0 citations
Introduction and objectives Left ventricular reverse remodeling (LVRR) after transcatheter aortic valve replacement (TAVR) is associated with outcomes, but the relevance of aortic-root geometry remains uncertain. We examined whether the annular eccentricity index (EI) and sinus of Valsalva height on computed tomography...
Xiao-Ting Li, Dai-Song Jiang, Jian Liu et al.· Frontiers in Cardiovascular...· 0 citations