Influence of Scan Body Geometry, Implant Angulation, and Interimplant Distance on the Accuracy of Maxillary Full-Arch Digital Implant Scans: An In Vitro Study

Jul 2026 · Medical Science Monitor · Vol 32 · 0 citations · 57 references
Medicine

TL;DR

Increased implant angulation and interimplant distance are associated with reduced accuracy and higher deviation values in digital full-arch scans, and the influence of scan body geometry varies depending on implant position.

Abstract

Background This study compared the accuracy of different scan body geometries under different implant angulation and interimplant distance conditions. Evidence regarding full-arch digital implant scanning under these conditions remains limited. Material/Methods In this in vitro study, 5 implants were placed in maxillary edentulous plaster models representing 3 implant systems (Bioinfinity, Meisinger, and Mode). The reference implant was positioned perpendicular to the ridge crest at the midline. Implant angulations (5° and 10°) and interimplant distances (5 mm and 10 mm) were standardized. Scan bodies were attached to the master model, and each model was scanned 10 times using an intraoral scanner (TRIOS 5). Three-dimensional deviations were calculated. Data were analyzed using 1-way ANOVA and Tukey HSD tests (α = 0.05). Results No significant differences were observed among the groups at 5° angulation (P > 0.05), whereas significant differences were detected at 10° angulation for implants 1 and 5 (P < 0.01). Bioinfinity showed higher deviation values at implant 1, while Meisinger demonstrated higher deviations at implant 5. At a 10-mm interimplant distance (implants 1 and 2), significant differences were found among the groups (P < 0.001), with Bioinfinity exhibiting lower deviation values. At a 5-mm distance, no significant difference was observed at implant 4 (P > 0.05), whereas a significant difference was found at implant 5 (P < 0.001), with the highest deviation in the Meisinger group. Conclusions Increased implant angulation and interimplant distance are associated with reduced accuracy and higher deviation values in digital full-arch scans. The influence of scan body geometry varies depending on implant position.

Read PDF

Similar papers

Open access Oct 2026

Influence of 3D Printing, Implant Macro-Design and Alveolar Ridge Morphology on Accuracy and Primary Stability in Static Computer-Assisted Implant Surgery: An In Vitro Study.

OBJECTIVES To assess the positional accuracy of static computer-assisted implant surgery (sCAIS) using guides printed by digital light processing (DLP) or material extrusion (ME), and to compare insertion torque (IT) and implant stability quotient (ISQ) of two tissue-level implants (Straumann TLC, TLX) in healed premol...

Malte Bagratuni, S. Abou-Ayash, Justus Pook et al. · 0 citations
Sep 2026

Influence of implant scan body material and intraoral scanner system on the accuracy of three-unit implant scans.

STATEMENT OF PROBLEM The accuracy of implant scans may be influenced by the intraoral scanner (IOS) system and the intraoral scan body (ISB) materials. However, evidence regarding their impact remains limited. PURPOSE The purpose of this in vitro study was to evaluate the trueness and precision of 3-unit implant scan...

Bahar Gurpinar, Burcu Kanat-Ertürk, Mehmet Karakuzu et al. · 0 citations
#software testing Open access Oct 2026

Clinical Accuracy of Complete Arch Implant Scans Recorded by Using a Customized Metallic Ring-Based Calibrated Framework: An In Vivo Study.

AIM To evaluate and compare the positional accuracy of complete arch intraoral digital implant scans obtained using a conventional scan body protocol and a customized metallic ring-based calibrated framework. MATERIALS AND METHODS A completely edentulous arch restored with six implants was scanned using two complete...

Miguel Gómez-Polo, Solange Vasquez-Ramos, Marta Revilla-León et al. · 0 citations
Open access Aug 2026

Effect of different splinting configurations on the accuracy of complete-arch digital implant impressions: An in vitro study

Background This in vitro study aimed to investigate the effect of different splinting configurations on the accuracy of digital impressions in fully edentulous arches. Materials and Methods Five configurations were tested on an edentulous maxillary phantom model containing six parallel implant analogs: G1: Non-splinted...

Simge Öner-Gedik, H. Yolcu, Pelin Türkoğlu et al. · 0 citations
Open access Sep 2026

Model-free digital-correction workflow for multi-implant restorations: A clinical proof-of-concept study.

PURPOSE This clinical proof-of-concept study evaluates the feasibility and accuracy of a digital workflow for fabricating and correcting implant-supported restorations of multiple dental implants in an edentulous maxilla using intraoral scanning and reverse-scan bodies. METHODS An intraoral scan (IOS) was performed u...

A. Kernen-Gintaute, M. Akulauskas, F. Kernen et al. · 0 citations
Open access Sep 2026

Accuracy and clinical passive fit of reverse and standard scan body protocols for complete-arch implant-supported digital impressions: An in vitro study.

Within the limitations of this in vitro study, the RSB protocol demonstrated accuracy comparable to the SCB protocol, with more stable implant position transfer in four-implant configurations and a high clinical passive fit rate, supporting its use as a reliable alternative in complete-arch digital implant workflows.

Rabia Kilicaslan Dasdelen, İkbal Leblebicioğlu Kurtuluş, K. Kılıç · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.