Combining Clinical LAFOV PET/CT with a Digital Twin Providing Motion-Free Ground Truth Reveals Quantitative Trade-offs in Respiratory Motion Correction

Aug 2026 · medRxiv · 0 citations
Medicine

TL;DR

Clinical PET imaging with a realistic digital twin and corresponding motion-free ground truth enabled objective validation of respiratory MoCo beyond conventional clinical evaluation, highlighting the value of ground-truth-based validation for developing and benchmarking respiratory motion correction and quantitative PET on LAFOV PET/CT systems.

Abstract

Purpose: Respiratory motion remains a major source of quantitative bias in PET and becomes increasingly relevant for high-sensitivity long axial field-of-view (LAFOV) PET/CT. Although numerous respiratory motion correction (MoCo) methods have been proposed, their quantitative accuracy cannot be established clinically because a patient-specific motion-free reference is fundamentally unavailable in vivo. This study combined clinical PET imaging with a digital twin, a realistic representation of both the PET/CT system and the patient, to objectively validate respiratory MoCo against a corresponding motion-free reference. Methods: Twenty patients (10 [18F]FDG with predominantly pulmonary lesions and 10 [18F]SiFAlin-TATE with predominantly hepatic lesions; total 135 lesions) were analyzed. The digital twin combined a validated LAFOV PET/CT simulation model with an anatomically realistic phantom containing 14 lung and liver lesions, two patient-derived respiratory patterns, and respiratory motion amplitudes of 2 and 3 cm, generating patient-like datasets with corresponding motion-free references. Data-driven and image-based MoCo were evaluated using lesion morphology, SUVmean, SUVmax, and metabolic tumor volume (MTV). Results: In patients, data-driven MoCo produced larger SUVmean increases than image-based MoCo for liver (48.1{+/-}18.9% vs. 17.0 {+/-} 12.0%; p<0.01), lower-lung (32.5{+/-}21.2% vs. 16.3{+/-}15.6%, p=0.06), and upper-lung lesions (28.4{+/-}32.0% vs. 10.4 {+/-} 17.2%; p<0.01), with similar findings for SUVmax and larger MTV reductions. Simulation revealed marked motion-induced SUVmean underestimation before correction, particularly in liver (-31.2{+/-}6.8%) and lower lung (-15.5{+/-}13.9%). Relative to the motion-free reference, data-driven MoCo most accurately recovered hepatic uptake (4.3{+/-}11.7% vs. -10.0 {+/-} 9.2%; p=0.01) but overestimated pulmonary uptake (lower lung: 19.8{+/-}16.3% vs. -1.6 {+/-} 10.2%; p=0.02). SUVmax showed the same regional behavior, whereas image-based MoCo yielded MTV estimates closer to the reference. Quantitative recovery was largely independent of respiratory pattern, while larger motion amplitudes mainly affected image-based MoCo. Conclusion: Combining clinical PET with a realistic digital twin and corresponding motion-free ground truth enabled objective validation of respiratory MoCo beyond conventional clinical evaluation. Larger correction-induced quantitative changes should not be equated with greater quantitative accuracy. Instead, MoCo performance was region- and metric-dependent, highlighting the value of ground-truth-based validation for developing and benchmarking respiratory motion correction and quantitative PET on LAFOV PET/CT systems.

Read PDF

Similar papers

Open access Sep 2026

Quantitative accuracy and clinical utility of ultra-low dose CT protocols for attenuation correction in PET/CT

In PET/CT imaging, CT frequently accounts for more than 50% of the total radiation exposure, even when used solely for attenuation correction and anatomical localisation. This poses a major limitation for dose-sensitive populations, including pregnant women and paediatric patients. This study systematically evalu...

Georgios I. Angelis, Y. Gholami, K. Willowson et al. · 0 citations
Open access Aug 2026

SO(3)-based and structure-guided deformable registration for respiratory motion correction in thoracic PET

TLCE-morph, a Tri-Path Lie Convolution Encoder-based learning framework for deformable respiratory motion correction in thoracic PET, achieves the most favorable overall quantitative performance and shows more consistent local structural recovery in representative motion-sensitive regions.

Hui Zhou, Long-Xi He, Si-Yu Wang et al. · 0 citations
Open access Sep 2026

Quantification of patient motion in FDG-PET/CT brain imaging

Objectives FDG-PET is an important imaging tool in evaluation of neurological disorders, including drug-resistant epilepsy. Head motion can degrade PET image quality; however, the extent in routine examinations remains insufficiently characterized. This retrospective study aimed to quantify the magnitude and pattern of...

K. Mackova, K. Dudasova, P. Ježdík et al. · 0 citations
#software testing Open access Sep 2026

Differences in the quantitative parameter of lung lesions in dynamic chest FDG PET/CT imaging: framing protocol, reconstruction algorithm, software processing

The primary quantitative endpoint, the Ki value, remains highly robust across all evaluated framing protocols, software platforms, and reconstruction algorithms and facilitates cross-center data harmonization and clinical standardization for multi-center dynamic 18F-FDG PET/CT.

Ya-Rong Zhang, X. Wumener, Mao-Qun Zhang et al. · 0 citations
Sep 2026

EARL-based quantitative evaluation of administered activity reduction for [68Ga]Ga-DOTATOC PET/CT on a 32-cm axial FOV digital BGO system: Phantom analysis and retrospective list-mode clinical assessment.

PURPOSE The increased sensitivity of long-axial field-of-view (FOV) digital PET/CT systems enables potential reductions in administered activity and/or acquisition time with respect to standard-field-of-view systems and established clinical protocols, while preserving quantitative performance. This study aimed to evalu...

A. Piai, B. Bordigoni, L. Antunovic et al. · 0 citations
Open access Sep 2026

Effects of count rate performance on image quality in an LAFOV PET/CT system

This study investigated the relationship between body size, administered activity, and image quality (IQ) using two patient-like phantoms representing normal and obese body profiles, and demonstrated the count rate capabilities up to and beyond the saturation limit.

Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.