Transcatheter tricuspid valve replacement—a comparison of clinical and procedural outcomes of two different prostheses

Sep 2026 · Frontiers in Cardiovascular Medicine · Vol 13 · 0 citations · 21 references
Medicine

TL;DR

TTVR with both, the Cardiovalve and EVOQUE prothesis is an effective treatment option, leading to profound TR reduction and meaningful clinical improvement at 30-day follow-up.

Abstract

Background Transcatheter Tricuspid Valve Replacement (TTVR) has emerged as a treatment option for patients with severe tricuspid regurgitation (TR) who are at high surgical risk and unsuitable for tricuspid transcatheter edge-to-edge repair (T-TEER). Objectives This study aims to compare procedural, clinical and echocardiographic outcomes of two different TTVR devices: the Cardiovalve (Venus MedTech) and the EVOQUE (Edwards Lifesciences) prosthesis. Methods We performed a retrospective, single-center cohort analysis including all consecutive patients who were treated with TTVR. Primary endpoints included TR reduction, NYHA class and safety events according to TVARC criteria. Results Overall, 25 consecutive patients were included in the analysis. The Cardiovalve prosthesis was implanted in 12 patients and the EVOQUE prosthesis in 13 patients. Intraprocedural success was 100%. There was a significant reduction in TR grade (4.20 vs. 0.48; p < 0.001) and NYHA class (2.72 vs. 1.88; p = 0.004) for the overall cohort at 30-day follow-up, but no differences between the two devices (TR grade 0.50 vs. 0.46; p = 0.408 and NYHA class 1.75 vs. 2.00; p = 0.813; Cardiovalve and EVOQUE respectively). Mortality rate at 30 days was 0% in both groups. One patient in each group (8.0%) required new pacemaker implantation. Conclusion TTVR with both, the Cardiovalve and EVOQUE prothesis is an effective treatment option, leading to profound TR reduction and meaningful clinical improvement at 30-day follow-up. Patient selection and long term follow up need to be the focus of future studies.

Read PDF

Similar papers

Open access Aug 2026

Transapical transcatheter mitral valve replacement with apical tethering system. Single-center mid-term results.

OBJECTIVES In high-risk patients, transcatheter mitral valve replacement (TMVR) is an alternative to surgery and transcatheter edge-to-edge repair (TEER). This study aims to evaluate the mid-term results of TMVR with the Tendyne System in a real-world population at a single center. METHODS Between November 2018 and D...

L. Besola, Cristina Di Franco, F. Giorgi et al. · 0 citations
Open access Sep 2026

TricValve system implantation in patients with cardiac implantable electronic devices: A multicentre observational study.

OBJECTIVES Tricuspid regurgitation (TR) is associated with high morbidity and mortality. The TricValve bicaval valve system has emerged as a transcatheter option for patients with severe symptomatic TR who are not candidates for surgery or orthotopic interventions. However, its safety in patients with pre-existing card...

K. Osipenko, Iuliana Coti, Paul Werner et al. · 0 citations
Review Sep 2026

Tricuspid Transcatheter Edge-to-Edge Repair in Patients with Transvalvular Cardiac Implantable Electronic Device Leads: A Systematic Review and Meta-Analysis.

Severe tricuspid regurgitation (TR) is associated with adverse outcomes, particularly in patients with cardiac implantable electronic devices (CIEDs), in whom transvalvular leads may contribute to valve dysfunction. Evidence supporting tricuspid transcatheter edge-to-edge repair (T-TEER) in this population remains limi...

Lorenzo Moscato, Francesco Stabile, G. Vadalá et al. · 0 citations
Open access Sep 2026

Temporal Trends in Transcatheter Tricuspid Valve Interventions After FDA Approval in the United States

Background Transcatheter tricuspid valve interventions (TTVIs) have recently received commercial approval; however, contemporary adoption trends in early postapproval phase are not well described. Objectives The objectives of the study were to characterize trends in utilization of transcatheter tricuspid edge-to-edge r...

Bharat Rawlley, Maximilian Reisinger, Hannah R. Murphy et al. · 0 citations
Aug 2026

Hemodynamic Changes After Transcatheter Valve Repair and Replacement for Tricuspid Regurgitation.

TEER was associated with a greater reduction in right atrial pressure and more favorable RV remodeling at 30 days, as opposed to TTVR, which achieved more complete and sustained elimination of tricuspid regurgitation with higher forward stroke volume.

Mohamad S. Alabdaljabar, Mohamed Elhadi, J. Thaden et al. · 0 citations
Case report Open access Sep 2026

Transcatheter Tricuspid Valve Replacement in High-Risk Elderly Patients: Procedural Feasibility, Complexity, and Delayed Conduction Disturbances.

The feasibility of TTVR in elderly patients with complex anatomy is demonstrated and effective in selected elderly high-risk patients with severe TR and challenging anatomy and delayed conduction disturbances as a clinically important complication requiring extended rhythm surveillance are highlighted.

L. Trabucchi, Sergio Torralbas Fitz, Rodolfo Arce et al. · 0 citations