Respiratory-Resolved Isotropic 3D T1 Mapping of the Carotid Vessel Wall With B1 + Correction (CARISMAT1C).

Sep 2026 · Magnetic Resonance in Medicine · 0 citations · 35 references
Medicine

TL;DR

C had high T1 accuracy in phantoms and high T1 precision in vivo in different sections of the carotid artery tree in different sections of the carotid artery tree.

Abstract

Purpose

To develop, optimize, and characterize a respiratory motion-resolved free-running isotropic 3D carotid vessel wall T1 mapping technique named CARISMAT1C.

Methods

An inversion-recovery gradient-echo free-running pulse sequence with interleaved double flip-angle (2FA) was implemented, and an extended-phase-graph dictionary was used to map the T1 relaxation time. In a phantom, the T1 accuracy was compared to that of a single-flip-angle (1FA) variant, the inversion-recovery spin-echo reference, and clinical routine MOLLI, while the T1 precision of several phyllotaxis-based radial trajectories was compared. The most precise sequence was used in 11 healthy volunteers (25 ± 3Y, 3F). A synthetic 3D gray-blood image was reconstructed from the source images and used to delineate the internal, external, and common carotid arteries. T1 values across artery sections were compared to those of MOLLI.

Results

In the phantom, the 2FA variant had higher T1 agreement with the IR-SE reference than 1FA or MOLLI. The top-to-bottom phyllotaxis trajectory with golden-step shuffling was the most precise and was retained for in vivo scans. End-expiratory carotid vessel wall T1 values (1185 ± 42 ms) were similar to those obtained with MOLLI (1162 ± 84 ms, p = 0.37) and showed less intersubject variability and high interscan repeatability. No significant differences were found across arterial segments in the vessel wall T1 values.

Conclusion

CARISMAT1C had high T1 accuracy in phantoms and high T1 precision in vivo in different sections of the carotid artery tree.

Read PDF

Similar papers

Open access Oct 2026

Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.

PURPOSE Hepatic MRI T1 relaxation time is a putative measure of fibrosis and inflammation but may be confounded by hepatic fat. This study evaluated a chemical shift-encoded Modified Look-Locker Inversion Recovery (CSE-MOLLI) technique for simultaneous water-specific T1 (water-T1), proton density fat fraction (PDFF), a...

Amol Pednekar, Joshua S. Greer, Pradipta Debnath et al. · 0 citations
Open access Sep 2026

Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T.

PURPOSE Investigating the feasibility of variable flip angle (VFA) T1 mapping for quantitative liver imaging at 0.55 T and evaluating the impact of RF spoiling correction and deep learning-based (DL) reconstruction on accuracy. METHODS A VFA protocol was developed and combined with DL reconstruction to improve quanti...

Majd Helo, M. Nickel, Thomas Kuestner · 0 citations
Open access Sep 2026

Native myocardial T1 mapping using rapid 1.5T T1FLASH and T1 MOLLI during breath-hold and free-breathing

Quantitative myocardial T1 mapping is an established technique in cardiovascular magnetic resonance (CMR). However, conventional T1 mapping requires relatively long breath-holds (BH), which may lead to breathing artifacts and reduced image quality in patients with limited BH capacity. The T1 fast low angle shot (T1FLAS...

L. Röwer, Yara Hussen-Demir, P. E. Verde et al. · 0 citations
Open access Sep 2026

Mapping R1 of fat, R1 of water, joint R2⁎, and proton density fat fraction at 3 T using the multi-echo MP2RAGE sequence.

PURPOSE To develop a technique for joint measurement of fat and water-specific longitudinal relaxation rates (R1f and R1w), effective transverse relaxation rate (R2∗), and proton density fat fraction (PDFF) combining the Multi-Echo Magnetization Prepared Two Rapid Acquisition of Gradient Echoes (ME-MP2RAGE) sequence an...

Jorge Campos Pazmiño, Marc-Antoine Fortin, Véronique Fortier et al. · 0 citations
Open access Aug 2026

Pre-Scan Breathing Pattern and Image Quality in Respiratory-Triggered T2-Weighted Liver MRI: Radial Versus Cartesian Fast Spin Echo.

BACKGROUND Respiratory-triggered T2-weighted imaging (T2WI) may use rectilinear sampling (respiratory-triggered fast recovery fast spin echo [RT-FRFSE]) or radial sampling (periodically rotated overlapping parallel lines with enhanced reconstruction [PROPELLER]). Their relative performance across breathing patterns is...

H. Park, Jimi Huh, K. Lee et al. · 1 citation
#graph neural networks Open access Oct 2026

Simultaneous T1 and T2 Mapping in the Brain With QuantoRAGE: An MP2RAGE Variant Using T2-Prepared Inversion.

PURPOSE To develop and validate a repeatable and reproducible approach, QuantoRAGE, for simultaneous whole-brain T1 and T2 mapping using adiabatic magnetization preparation. METHODS QuantoRAGE is a 3D FLASH-based sequence using an adiabatic T2-prepared inversion followed by two readout blocks. Sequence repetitions wi...

Natalia Pato Montemayor, Tâm Johan Nguyên, J. Marques et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.