Analytical Agreement and Diagnostic Performance of a Commercial Point-of-Care Lipid Testing System for Lipid Profile and Atherogenic Ratios

Sep 2026 · Diagnostics · Vol 16, pp. 3116 · 0 citations · 42 references

TL;DR

The Mission® Cholesterol 3-in-1 system shows good analytical agreement with laboratory testing, particularly for triglycerides and the HDL-C, TG/HDL-C, and TG/LDL-C ratios, which supports its use as a rapid point-of-care screening tool, especially in resource-limited settings.

Abstract

Background: Point-of-care testing (POCT) devices offer a rapid alternative for lipid assessment, but comprehensive analytical validation against reference laboratory methods remains limited. Objective: This study aimed to evaluate the agreement, systematic bias, and diagnostic performance of the Mission® Cholesterol 3-in-1 system for lipid profile measurements and atherogenic ratios. Methods: A prospective cross-sectional agreement study was conducted among 120 participants. Lipid profiles were measured using eleven Mission® Cholesterol 3-in-1 analyzers and compared with those measured using the reference laboratory method. Agreement was assessed using the intraclass correlation coefficient (ICC), concordance correlation coefficient (CCC), coefficient of determination (R2), root mean square error (RMSE), mean absolute error (MAE), percentage bias, Bland–Altman plots, Passing–Bablok regression, and diagnostic accuracy. Results: Triglycerides demonstrated excellent agreement (ICC = CCC = 0.972; bias = 5.19%), whereas total cholesterol, HDL-C, non-HDL-C, and LDL-C showed good agreement (ICC range: 0.817–0.854). Among the atherogenic ratios, TG/HDL-C (ICC = 0.932) and TG/LDL-C (ICC = 0.931) exhibited the highest concordance. Passing–Bablok regression revealed constant bias for all analytes, except for LDL-C estimated using the Sampson equation and the TG/LDL-C ratio, while proportional bias was observed for all analytes, except for triglycerides and the TG/LDL-C ratio. Despite these systematic differences, diagnostic performance remained excellent, with sensitivities of 94.29–98.33% and specificities of 68.33–91.80%. Conclusion: The Mission® Cholesterol 3-in-1 system shows good analytical agreement with laboratory testing, particularly for triglycerides and the HDL-C, TG/HDL-C, and TG/LDL-C ratios. Although systematic bias limits complete interchangeability with reference methods, its excellent diagnostic performance supports its use as a rapid point-of-care screening tool, especially in resource-limited settings.

Read PDF

Similar papers

Open access Aug 2026

Assessing the reliability and validity for metabolic syndrome diagnosis using point-of-care analyzers in a mobile setup - a pilot study

The metabolic syndrome (MetS) is characterized by having three or more of these risk factors: increased blood glucose (Glc), serum triglycerides (TG), blood pressure (BP), waist circumference (WC), and reduced high-density lipoprotein (HDL). Mobile diagnostics using point-of-care (POC) blood analyzers could help identi...

S. Houtenbos, A. Schraplau, P. Wippert et al. · 0 citations
Open access Aug 2026

Validation of Precision and Accuracy of Point-of-Care Testing Glucose Analyzers with a Standard Laboratory Method (Glucose Oxidase Method)

POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.

E. Nwobodo, N. T. Emmanuel · 0 citations
Open access Oct 2026

Validation of the eCL8000i point-of-care hs-cTnT assay: analytical performance and clinical utility in Chinese patients

Abstract Objectives We validated the analytical performance and clinical diagnostic efficacy of the eCL8000i point-of-care testing (POCT) system for high-sensitivity cardiac troponin T (hs-cTnT), following the standardized verification protocol for hs-cTnT POCT assays. Methods A two-phase evaluation was conducted: Phas...

Fei Ding, Yu-Yao Li, Jia Li et al. · 0 citations
Oct 2026

A-404 Determinants of Analytical Performance in Centrifugation–Based Hematocrit: Effects of Calibration and Internal Quality Control Frequency Across Hospital and Primary Point-of-Care Sites

Manual centrifugation–based hematocrit (Hct) remains widely used as point-of-care testing (POCT) in hospital laboratories and subdistrict health promotion hospitals (SDHHs), particularly in decentralized settings. However, evidence is limited on whether quality-related factors influence satisfactory performance w...

Wanvisa Treebuphachatsakul · 0 citations
Open access Sep 2026

A comparative analysis of different methods for LDL cholesterol calculation

Abstract Objectives Accurate estimation of low-density lipoprotein cholesterol (LDL-C) is essential for cardiovascular risk management. This study evaluated the performance of the Friedewald, Martin–Hopkins (MH), and National Institutes of Health Equation 2 (NIH2) methods compared with direct enzymatic measurement acro...

M. Ağıllı, Alpaslan Coşar, E. Karaismailoğlu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.