Oct 2026· Journal of Laboratory Medicine· 0 citations· 22 references
TL;DR
The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage.
Abstract
Abstract Objectives We validated the analytical performance and clinical diagnostic efficacy of the eCL8000i point-of-care testing (POCT) system for high-sensitivity cardiac troponin T (hs-cTnT), following the standardized verification protocol for hs-cTnT POCT assays. Methods A two-phase evaluation was conducted: Phase I (564 samples) assessed precision, lower limit of quantitation (LoQ), method comparison, calibrator interchangeability, reference intervals and turnaround time (TAT) per CLSI guidelines; Phase II (721 emergency chest pain patients) used coronary angiography as the gold standard to evaluate AMI diagnostic performance vs. Roche Cobas e801 via receiver operating characteristic (ROC) analysis. Results Phase I showed total CVs of 5.22 % (low) and 1.69 % (high), qualified LoQ, strong correlation with Cobas e801 (r=0.998), acceptable calibrator interchangeability, sex-specific 99th percentile reference values, and TAT shortened to 17 min. Phase II revealed eCL8000i AUC 0.860 (95 % CI 0.833–0.885) vs. Cobas 0.891 (95 % CI 0.865–0.912); sensitivities/specificities were 74.8/85.1% and 82.0/84.3 %, with high negative predictive values (94.9 vs. 96.3 %). Conclusions The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage. Although the sample size for establishing sex-specific 99th percentile URLs (280 males, 284 females) did not fully meet the IFCC-recommended target of ≥400 per sex, our preliminary reference intervals provide a useful starting point for the Chinese population and require further validation in a larger cohort.
The POC assay showed high clinical concordance with CL-based ESC 0/1-h triage classification, with all discordances confined to adjacent categories, providing pre-implementation evidence supporting further prospective evaluation of POC hs-cTnI testing under routine bedside conditions.
N. Kongruttanachok, Naphat Chaijirawiwat, M. T. Angkana Chirapanuruk et al.· Clinica chimica acta; intern...· 0 citations
The study objectives were: 1) establish 99th percentile upper reference limits (URL) for the VITROS™ Immunodiagnostic Products hs Troponin I (VITROS hsTnI) Assay, 2) confirm VITROS hsTnI Assay meets analytical performance criteria for a high-sensitivity cardiac troponin assay established by the IFCC Task Force on...
A. Ott-Vasconi, M. Colvin, S. Jackson et al.· Clinical Chemistry· 0 citations
Objective
To perform a partial validation of a point-of-care assay (Vcheck cardiac troponin I [VC-cTnI; cTnI]) for measurement of cTnI in canine serum and to perform a method comparison between the VC-cTnI and an established chemiluminescent assay (ADVIA Centaur XP High-Sensitivity Troponin I [AC-cTnI-HS]).
Methods
F...
Bobbie Ditzler, A. Saunders, S. Wesselowski et al.· American Journal of Veterina...· 0 citations
BACKGROUND
This study evaluated the analytical performance and practicability of a novel point-of-care testing (POCT) device for measuring HbA1c and C-reactive protein (CRP) compared with central laboratory methods.
METHODS
Imprecision, bias, and analytical total error (TEa) were evaluated using quality control mater...
Ali Khatib, Álvaro Martínez Ortega, Noor Lachhab et al.· Annals of Clinical Biochemis...· 0 citations
The i-STAT hs-cTnI assay demonstrated high agreement with the Abbott Alinity platform for troponin-based risk classification and correctly identified all myocardial infarction cases, suggesting that the i-STAT assay may be safely used for early risk stratification of ED patients, particularly for identifying low-risk p...
Merih T. Tesfazghi, Elif Yucebay, Isabel Espinosa et al.· Clinical Chemistry· 0 citations
High-sensitivity cardiac troponin (hs-cTn) is a well-established diagnostic and prognostic biomarker in various clinical settings. Despite high myocardial specificity, concentrations of hs-cTn are influenced by a variety of factors like age, sex, prevalent cardiovascular disease and renal function, complicating p...
J. Lehmacher, A. Haack, N. Soerensen et al.· European Heart Journal· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.