Oct 2026· Clinica chimica acta; international journal of clinical chemistry· pp.
122902
· 0 citations· 23 references
Medicine
TL;DR
The POC assay showed high clinical concordance with CL-based ESC 0/1-h triage classification, with all discordances confined to adjacent categories, providing pre-implementation evidence supporting further prospective evaluation of POC hs-cTnI testing under routine bedside conditions.
Abstract
Objective
Although point-of-care (POC) high-sensitivity cardiac troponin I (hs-cTnI) assays have demonstrated good analytical agreement with central laboratory (CL) testing, it remains uncertain whether they yield the same clinical triage decisions. We evaluated the clinical concordance of a POC hs-cTnI assay with a CL assay using the European Society of Cardiology (ESC) 0/1-h algorithm.
Methods
This prospective diagnostic agreement study included 127 patients presenting to the emergency department with suspected acute coronary syndrome. The primary outcome was concordance of patient classification according to the 0/1-h algorithm, assessed using the quadratic weighted kappa (κ) statistic. Secondary outcomes included quantitative agreement assessed by the intraclass correlation coefficient (ICC) and Bland-Altman analysis, and analytical comparability between whole blood and plasma on the POC platform.
Results
Overall, 114/127 patients (89.8%) were assigned to the same triage category (κ = 0.921; 95% CI, 0.878-0.964). All 13 discordant classifications (10.2%) occurred between adjacent categories, with no CL rule-in patient classified as POC rule-out. Quantitative agreement between assays was good (ICC = 0.882; 95%CI 0.835-0.917), with a mean bias of -5.27 ng/L. Whole blood and plasma showed close agreement (ICC = 0.999; 95%CI 0.998-0.999), with a mean bias of 0.50 ng/L.
Conclusions
The POC assay showed high clinical concordance with CL-based ESC 0/1-h triage classification, with all discordances confined to adjacent categories. These findings provide pre-implementation evidence supporting further prospective evaluation of POC hs-cTnI testing under routine bedside conditions.
The European Society of Cardiology (ESC) 0/1-hour algorithm is recommended for the rapid triage of patients with suspected non-ST-elevation myocardial infarction (NSTEMI). While this strategy is well-established for the reference high-sensitivity cardiac troponin T (hs-cTnT) assay, the algorithm is assay-dependen...
A. Nguyen, D. Dao, T. Phạm et al.· European Heart Journal· 0 citations
The i-STAT hs-cTnI assay demonstrated high agreement with the Abbott Alinity platform for troponin-based risk classification and correctly identified all myocardial infarction cases, suggesting that the i-STAT assay may be safely used for early risk stratification of ED patients, particularly for identifying low-risk p...
Merih T. Tesfazghi, Elif Yucebay, Isabel Espinosa et al.· Clinical Chemistry· 0 citations
High-sensitivity cardiac troponin (hs-cTn) assays have significantly enhanced the diagnosis in patients suspected of Acute Coronary Syndrome (ACS). New hs-cTn point-of-care (POC) devices, with a turnaround time <20 minutes, may reduce time to diagnosis compared to central laboratory testing. However, it remains u...
B. Backus, B. Schueler, L. E. van Daalen· European Heart Journal· 0 citations
The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage.
Fei Ding, Yu-Yao Li, Jia Li et al.· Journal of Laboratory Medici...· 0 citations
High-sensitivity cardiac troponin (hs-cTn) assays have significantly enhanced the diagnosis in patients suspected of acute coronary syndrome (ACS). New hs-cTn point-of-care (POC) devices (turnaround time <20 min) may significantly reduce time to diagnosis. We evaluated the diagnostic accuracy of four currently commerci...
L. E. van Daalen, B. Schueler, P. Collinson et al.· European journal of emergenc...· 0 citations
Advances in high-sensitivity cardiac troponin T (hs-cTnT) assay technology have improved the accuracy of detecting and quantifying cardiomyocyte injury in patients with suspected acute coronary syndrome (ACS). However, hs-cTnT concentrations may still be influenced by confounding factors unrelated to myocardial i...
A. Pujante-Martinez, C. Piqueras-Sánchez, J. Moreno-Fernández et al.· European Heart Journal· 0 citations