Skip to content

Clinical triage concordance of a point-of-care high-sensitivity cardiac troponin I assay using the ESC 0/1-hour algorithm.

Oct 2026 · Clinica chimica acta; international journal of clinical chemistry · pp. 122902 · 0 citations · 23 references
Medicine

TL;DR

The POC assay showed high clinical concordance with CL-based ESC 0/1-h triage classification, with all discordances confined to adjacent categories, providing pre-implementation evidence supporting further prospective evaluation of POC hs-cTnI testing under routine bedside conditions.

Abstract

Objective

Although point-of-care (POC) high-sensitivity cardiac troponin I (hs-cTnI) assays have demonstrated good analytical agreement with central laboratory (CL) testing, it remains uncertain whether they yield the same clinical triage decisions. We evaluated the clinical concordance of a POC hs-cTnI assay with a CL assay using the European Society of Cardiology (ESC) 0/1-h algorithm.

Methods

This prospective diagnostic agreement study included 127 patients presenting to the emergency department with suspected acute coronary syndrome. The primary outcome was concordance of patient classification according to the 0/1-h algorithm, assessed using the quadratic weighted kappa (κ) statistic. Secondary outcomes included quantitative agreement assessed by the intraclass correlation coefficient (ICC) and Bland-Altman analysis, and analytical comparability between whole blood and plasma on the POC platform.

Results

Overall, 114/127 patients (89.8%) were assigned to the same triage category (κ = 0.921; 95% CI, 0.878-0.964). All 13 discordant classifications (10.2%) occurred between adjacent categories, with no CL rule-in patient classified as POC rule-out. Quantitative agreement between assays was good (ICC = 0.882; 95%CI 0.835-0.917), with a mean bias of -5.27 ng/L. Whole blood and plasma showed close agreement (ICC = 0.999; 95%CI 0.998-0.999), with a mean bias of 0.50 ng/L.

Conclusions

The POC assay showed high clinical concordance with CL-based ESC 0/1-h triage classification, with all discordances confined to adjacent categories. These findings provide pre-implementation evidence supporting further prospective evaluation of POC hs-cTnI testing under routine bedside conditions.

View source

Similar papers

Open access Oct 2026

Diagnostic performance of a novel high-sensitivity cardiac troponin T assay in the ESC 0/1-hour algorithm for the early diagnosis of NSTEMI

The European Society of Cardiology (ESC) 0/1-hour algorithm is recommended for the rapid triage of patients with suspected non-ST-elevation myocardial infarction (NSTEMI). While this strategy is well-established for the reference high-sensitivity cardiac troponin T (hs-cTnT) assay, the algorithm is assay-dependen...

A. Nguyen, D. Dao, T. Phạm et al. · 0 citations
Oct 2026

A-041 Clinical Performance and Risk Classification Agreement of Point-of-Care i-STAT High-Sensitivity Troponin I Compared with Abbott Alinity hs-cTnI in ED Patients with Suspected Acute Coronary Syndrome

The i-STAT hs-cTnI assay demonstrated high agreement with the Abbott Alinity platform for troponin-based risk classification and correctly identified all myocardial infarction cases, suggesting that the i-STAT assay may be safely used for early risk stratification of ED patients, particularly for identifying low-risk p...

Merih T. Tesfazghi, Elif Yucebay, Isabel Espinosa et al. · 0 citations
Review Open access Oct 2026

What is the diagnostic accuracy of different high sensitive point-of-care troponin assays for the assessments of patients presenting with chest pain at the emergency department?

High-sensitivity cardiac troponin (hs-cTn) assays have significantly enhanced the diagnosis in patients suspected of Acute Coronary Syndrome (ACS). New hs-cTn point-of-care (POC) devices, with a turnaround time <20 minutes, may reduce time to diagnosis compared to central laboratory testing. However, it remains u...

B. Backus, B. Schueler, L. E. van Daalen · 0 citations
Open access Oct 2026

Validation of the eCL8000i point-of-care hs-cTnT assay: analytical performance and clinical utility in Chinese patients

The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage.

Fei Ding, Yu-Yao Li, Jia Li et al. · 0 citations
Review Oct 2026

Diagnostic accuracy of high-sensitivity point-of-care troponin assays in emergency department for patients presenting with chest pain: a systematic review.

High-sensitivity cardiac troponin (hs-cTn) assays have significantly enhanced the diagnosis in patients suspected of acute coronary syndrome (ACS). New hs-cTn point-of-care (POC) devices (turnaround time <20 min) may significantly reduce time to diagnosis. We evaluated the diagnostic accuracy of four currently commerci...

L. E. van Daalen, B. Schueler, P. Collinson et al. · 0 citations
Open access Oct 2026

Clinical performance of a sixth-generation high-sensitivity cardiac troponin T assay compared with standard high-sensitivity troponin T in a real-world cohort

Advances in high-sensitivity cardiac troponin T (hs-cTnT) assay technology have improved the accuracy of detecting and quantifying cardiomyocyte injury in patients with suspected acute coronary syndrome (ACS). However, hs-cTnT concentrations may still be influenced by confounding factors unrelated to myocardial i...

A. Pujante-Martinez, C. Piqueras-Sánchez, J. Moreno-Fernández et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.