Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.

Sep 2026 · Applied Radiation and Isotopes · Vol 238, pp. 112946 · 0 citations · 24 references
Medicine

TL;DR

The accuracy of TG-43 algorithm against Monte Carlo (MC) simulation is investigated, confirming its adequacy in homogeneous phantom, and the necessity of adopting Model-based calculation algorithms such as MC to improve the dosimetric precision and treatment accuracy in HDR brachytherapy is emphasized.

Abstract

The accuracy of dose calculation algorithms significantly influences brachytherapy treatment outcomes. TG-43 algorithm has been widely adopted due to its simplicity, which has standardized contemporary brachytherapy dose calculations worldwide. However, TG-43's homogeneity assumption potentially compromises clinical treatment accuracy. The present study aimed to investigate TG-43 accuracy against Monte Carlo (MC) simulation. Varian GammaMed Plus 192Ir High-dose-rate (HDR) brachytherapy source was modeled using TOPAS3.9 MC simulation-framework. Three well-defined test scenarios were examined: Homogeneous water phantom, TG-186 generic-applicator, and computed-tomography (CT) based intracavitary brachytherapy (ICBT) patient plan. Corresponding TG-43 based treatment plans were generated using Eclipse BrachyVision and imported into TOPAS for MC-simulation. Both TG-43 and MC dose distributions were evaluated using point-dose assessment, dose-volume metrics, dose-volume histogram, and γ-analysis with 2% Dose-difference and 1 mm Distance-to-agreement. In a homogeneous water phantom, TG-43 and MC exhibited close agreement, with dose-difference (Δ%) at the reference point and for High-risk clinical target volume (HRCTV) found to be below 1-3% and γ-passing rates ≥95%. However, in the presence of a TG-186 applicator, TG-43 substantially overestimated doses by 60-93% on shielded side and 4-13% on unshielded side. For CT-based ICBT patient plan, TG-43 overestimated HRCTV and organ-at-risk (OAR) doses relative to MC. The mean γ-passing rate in clinical ICBT plan was 90%, indicating moderate spatial agreement. Present study validated the accuracy of TG-43 algorithm against MC simulation confirming its adequacy in homogeneous phantom. However, significant deviations were observed in the presence of an applicator, and tissue heterogeneity revealed TG-43 limitations. TG-43 consistently overestimates dose owing to the homogeneity assumption. These findings emphasize the necessity of adopting Model-based calculation algorithms such as MC to improve the dosimetric precision and treatment accuracy in HDR brachytherapy.

View source

Similar papers

Open access Aug 2026

A Monte Carlo based dose verification tool for an HDR Ir‐192 brachytherapy source and its evaluation on extremity malignancies

Abstract Background TG‐43 formalism dose calculation for brachytherapy with the homogeneous water‐based assumption is widely used, which can lead to dose inaccuracy in treatment planning. Purpose Extremity malignancies are also treated with high‐dose‐rate (HDR) 192Ir interstitial brachytherapy, a treatment site that ha...

Peixiong Li, C. Kong, Rihui Zhang et al. · 0 citations
Sep 2026

Dosimetric comparison of HDR Co-60 and Ir-192 brachytherapy sources by model-based dose calculation algorithms and its clinical implication in different brachytherapy sites and techniques.

INTRODUCTION Numerous dosimetric studies have reported clinically insignificant differences between iridium-192 and cobalt-60. However, such comparison studies were conducted using TG-43 algorithms where heterogeneity is not accounted. Model based dose calculation algorithms (MBDCA) significantly improve dosimetric acc...

Raghavendra Hajare, Rohit Avinash Vadgaonkar, Taushiful Hoque et al. · 0 citations
Sep 2026

Physics in Medicine & Biology

This work establishes a robust, cross-validated computational baseline for brachytherapy dosimetry using general-purpose MC codes under realistic photon emission spectra under realistic, unsimplified photon emission spectra.

Lucas F Ara ´ ujo, Asif Amin, Adriano Sabino et al. · 0 citations
Open access Aug 2026

Comparison of Gonadal Dose Estimation Using Treatment Planning Systems and Monte Carlo Methods in Pelvic Radiotherapy.

INTRODUCTION Unintended irradiation of organs outside the target volume remains a concern in pelvic radiotherapy, particularly for radiosensitive structures such as the gonads. Analytical dose calculation algorithms used in treatment planning systems (TPS) may underestimate out-of-field doses. Monte Carlo (MC) methods...

D. Ribeiro, G. Campos, E. Carolino et al. · 0 citations
Sep 2026

Effects of grid size and statistical uncertainty parameters on dose calculations using the Monte Carlo algorithm in the Monaco treatment planning system for liver stereotactic body radiotherapy.

Stereotactic body radiotherapy (SBRT) for liver metastases requires accurate dose calculation. The Monte Carlo (MC) algorithm in the Monaco treatment planning system (TPS) provides high dose-calculation accuracy; however, user-defined grid size (GS) and statistical uncertainty (SU) settings influence computational effi...

Yunus Bayram, M. Saynak, Yavuz Aksoy et al. · 0 citations
Open access Sep 2026

Dosimetric Impact of Deformable Image Registration versus Simple Summation in Multi-Fraction HDR Brachytherapy for Cervical Cancer under Standard Protocol

DIR-based dose accumulation resulted in significantly lower HR-CTV dose estimates compared with SS, while dose differences for the bladder and rectum remained minimal, and future research should correlate DIR-based dosimetry with clinical outcomes to support anatomically adaptive brachytherapy.

Chitchaya Suwanraksa, Bheampos Rojkitjanuruk, Tanapat Suwanpan et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.