Abstract Background TG‐43 formalism dose calculation for brachytherapy with the homogeneous water‐based assumption is widely used, which can lead to dose inaccuracy in treatment planning. Purpose Extremity malignancies are also treated with high‐dose‐rate (HDR) 192Ir interstitial brachytherapy, a treatment site that has not been specifically investigated in prior comparative dosimetric studies, and the proximity of source dwell positions to bone and air introduces significant heterogeneity. BrachyPlanCheck, an in‐house brachytherapy second dose verification program for Monte Carlo (MC) simulations, incorporates these heterogeneities and is used for a comparative dosimetric study on extremity malignancies. Methods BrachyPlanCheck was commissioned and validated in both a homogeneous water phantom and a heterogeneous breast phantom. The clinical evaluation included 8 patient‐specific lower limb brachytherapy plans, comparing TG‐43 and MC calculation using BrachyPlanCheck. Dosimetric differences were evaluated using dose comparisons, gamma index analysis, and dose‐volume histogram (DVH) metrics for clinical target volume (CTV) and organ‐at‐risk (OAR), including bone and skin. Results In commissioning Level 1, MC simulations using BrachyPlanCheck demonstrated point dose agreement within ± 0.9%. Gamma passing rates exceeded 99.9% with criteria 1 mm/1% in commissioning Level 2. In clinical evaluation, TG‐43 calculation overestimated CTV V150 by up to 10.50% and CTV D90 by up to 7.25%. OAR doses were generally higher with TG‐43, with mean differences of 1.09% for bone D1cc (p = 0.109), 1.38% for bone D0.1cc (p = 0.078), 7.30% for skin D2cc (p = 0.008), and 5.68% for skin D0.1cc (p = 0.008). Conclusion Validated by phantom and patient study, BrachyPlanCheck demonstrated robust performance for independent secondary dose verification. TG‐43 calculation overestimated dose to CTV and OAR in extremity malignancy cases compared with BrachyPlanCheck.
The accuracy of TG-43 algorithm against Monte Carlo (MC) simulation is investigated, confirming its adequacy in homogeneous phantom, and the necessity of adopting Model-based calculation algorithms such as MC to improve the dosimetric precision and treatment accuracy in HDR brachytherapy is emphasized.
R. Shende, S. J. Dhoble, Yassir El Ghazi· Applied Radiation and Isotop...· 0 citations
INTRODUCTION
Numerous dosimetric studies have reported clinically insignificant differences between iridium-192 and cobalt-60. However, such comparison studies were conducted using TG-43 algorithms where heterogeneity is not accounted. Model based dose calculation algorithms (MBDCA) significantly improve dosimetric acc...
DIR-based dose accumulation resulted in significantly lower HR-CTV dose estimates compared with SS, while dose differences for the bladder and rectum remained minimal, and future research should correlate DIR-based dosimetry with clinical outcomes to support anatomically adaptive brachytherapy.
Chitchaya Suwanraksa, Bheampos Rojkitjanuruk, Tanapat Suwanpan et al.· Journal of Health Science an...· 0 citations
Abstract Background Frameless thermoplastic masks are standard for intracranial stereotactic radiosurgery (SRS), with the Encompass (CQ Medical) widely used on the Varian HyperArc platform. A candidate mask must minimize beam attenuation, permit accurate dose calculation, and enable reproducible setup, yet these are ra...
Adi Robinson, Matthew R. Caldwell· Journal of Applied Clinical...· 0 citations
BACKGROUND
High-dose-rate brachytherapy (HDRBT) is an established treatment for localized prostate cancer. Despite its efficacy, HDRBT can exhibit dosimetric discrepancies between planned and delivered doses. In vivo dosimetry (IVD) using metal-oxide-semiconductor field-effect transistors (MOSFETs) enables real-time mo...
S. Ruiz-Arrebola, D. Guirado, Rosa Fabregat-Borrás et al.· Physica medica (Testo stampa...· 0 citations
Background: This study aimed to evaluate the feasibility and dosimetric characteristics of 90Sr/90Y and 188Re sources for treating superficial oral rhabdomyosarcoma (RMS) using a surface mold brachytherapy approach.
Materials and Methods: In this Monte Carlo simulation study, TG-149 dosimetric parameters were calculat...
Shahriyar Seraj, Shiva Rahbar Yazdi, Soroush Najafi Sarvestani et al.· Iranian Journal of Pediatric...· 0 citations