Sep 2026· European Journal of Radiology· Vol 205, pp.
113245
· 0 citations· 27 references
Medicine
TL;DR
A single 140-kV protocol with 68-70-keV VMIs yielded low errors, supporting potential opportunistic liver fat assessment across body sizes, including large body habitus.
Abstract
Objective
To identify common photon-counting CT (PCCT) virtual monochromatic imaging (VMI) settings for accurate hepatic fat fraction (FF) quantification across body sizes, including large body habitus.
Methods
Six non-iodinated fat lesions with known FFs (5%-40%) were embedded in anthropomorphic liver phantoms representing medium (25 × 32.5 cm2) and large (31 × 39 cm2) abdomens. Phantoms were repeatedly scanned on a PCCT system (NAEOTOM Alpha) at 120 and 140 kV. CT numbers were measured on VMIs from 40-190 keV in 1-keV increments. Linear regression between ground-truth FF and Hounsfield units (HU) was used to estimate FF, and the settings minimizing relative root-mean-square error (rRMSE) in both phantoms were identified. Prediction accuracy was further assessed using leave-one-out cross-validation on the four lesions with FF of 10%-30%, with each prediction based on a regression model fitted from the remaining five lesions.
Results
At 140 kV, 69-keV VMI demonstrated the best overall performance, with FF (%) = - 0.690HU + 36.84 (R2 > 0.996) and rRMSE ≤ 3.6% for both phantoms. Robust performance (rRMSE ≤ 5%) was maintained across 68-70 keV. Individual-lesion relative errors ranged from - 3.9% to 5.8%, while cross-validated prediction relative errors for FF 10%-30% ranged from - 5.59% to 5.98%. Performance improved over previously reported single- and dual-energy CT results, while using 40% less dose.
Conclusion
PCCT provided consistent hepatic FF quantification across phantom sizes. A single 140-kV protocol with 68-70-keV VMIs yielded low errors, supporting potential opportunistic liver fat assessment across body sizes, including large body habitus.
RATIONALE AND OBJECTIVES
To evaluate the robustness and grading performance of Photon-counting CT (PCCT)-derived fat fraction (CTFF) in liver fat quantification across different protocols and cohorts with varying liver conditions, using proton density fat fraction (PDFF) as the reference.
MATERIALS AND METHODS
The pr...
Xin-Xin Xu, Jie Yuan, Xin-Xin Cai et al.· Academic Radiology· 0 citations
RATIONALE AND OBJECTIVES
To evaluate the feasibility and accuracy of photon-counting CT (PCCT)-derived parameters for quantifying vertebral bone marrow adipose tissue (BMAT), using MRI proton density fat fraction (PDFF) as the reference standard.
MATERIALS AND METHODS
Adult participants with low back pain were prospe...
D. Shu, Jian-Ping Wang, Ding Liang et al.· European Journal of Radiolog...· 0 citations
The findings support implementing standardized protocols with automatic dual-range reconstructions to enable efficient, individualized carotid plaque assessment in clinical practice and support implementing standardized protocols with automatic dual-range reconstructions for keV selection based on plaque composition.
Mirela Kostadinova, Lea B. Uebelacker, Tommaso D'Angelo et al.· Diagnostics· 0 citations
Abstract Background In abdominal spectral Computed Tomography (CT), organ dose assessment is important for radiation safety but remains challenging due to the complexity of Monte Carlo (MC) simulations. Purpose This study aimed to compare the correlations of three size‐specific dose estimate (SSDE) metrics (including t...
Tian Qin, Qing-Ting Sun, Yi-Han Fan et al.· Journal of Applied Clinical...· 0 citations
Accurate patient-specific mean glandular dose (MGD) estimation in breast computed tomography (BCT) requires anatomically realistic models, but high-resolution patient-derived phantoms impose high computational demands on Monte Carlo (MC) dosimetry. We previously developed and validated an EGSnrc-based framework for syn...
A. Entezam, A. Pakzad, C. Hall et al.· 0 citations
PURPOSE
This study aimed to (i) compare TotalSegmentator Fast and Normal inference modes on low-dose PET/CT, (ii) evaluate whether CT denoising improves segmentation longitudinal consistency, and (iii) examine whether the selected pipeline can capture long-term within-participant volumetric trends over approximately 13...
Rika Kobayashi, Tetsuro Sekine, Shogo Imai et al.· Japanese Journal of Radiolog...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.