Skip to content
Open access

Reproducibility of perfusion blood volume and ventilation imaging for lung function assessment

Sep 2026 · Physics in Medicine and Biology · Vol 71, pp. 185001 · 0 citations · 52 references
Physics Medicine

TL;DR

Despite systematic perfusion increases early in treatment, its spatial stability remained stronger, supporting DECT perfusion as a more robust biomarker for longitudinal lung function assessment and for informing adaptive, functionally-guided RT in NSCLC.

Abstract

Objective. To assess the reproducibility of dual-energy CT (DECT) derived perfusion blood volume (PBV) and four-dimensional CT (4DCT) derived lung ventilation images. Approach. We enrolled patients on an IRB-approved clinical trial of functional lung imaging. All had locally advanced non-small cell lung cancer (NSCLC) treated with definitive, conventionally fractionated concurrent chemoradiation radiotherapy (RT) with or without consolidation immunotherapy. Patients completed imaging at three timepoints: (1) RT simulation, (2) second-week of RT, and (3) six-months post-RT. Free-breathing 4DCT and contrast-enhanced DECT were acquired at each timepoint. To evaluate reproducibility of PBV and ventilation maps, voxel intensities within the lung parenchyma from exclusively the first two timepoints were binned according to relative image intensity. Binary masks were generated at 20%, 40%, 60%, and 80% thresholds. Dice-similarity-coefficients (DSC) and structural-similarity-index-measure (SSIM) were calculated between the corresponding masks to assess the robustness and reproducibility. Scatter plots were generated to visualize voxel-wise intensity agreement. Main Results. Reproducibility of ventilation and perfusion images was evaluated in 28 patients between baseline and 2 week follow-up. For perfusion, the 20%–100% threshold showed the highest DSC (Scanner A = 0.91 ± 0.06; Scanner B = 0.81 ± 0.08), with progressive decreases at higher thresholds (Scanner A: 40%–100%(0.81 ± 0.08), 60%–100%(0.65 ± 0.07), and 80%–100%(0.48 ± 0.07); Scanner B: 40%–100%(0.73 ± 0.13), 60%–100%(0.58 ± 0.08), and 80%–100%(0.39 ± 0.07), respectively). Ventilation images showed a similar trend, with DSC of 20%–100%(0.83 ± 0.05), 40%–100%(0.71 ± 0.07), 60%–100%(0.59 ± 0.10), and 80%–100%(0.46 ± 0.11). For perfusion, mean SSIM for Scanner A decreased from 20%–100%(0.72 ± 0.10) to 80%–100%(0.66 ± 0.14); Scanner B from 20%–100%(0.55 ± 0.10) to 80%–100%(0.54 ± 0.08). For ventilation, the SSIM had a similar trend and declined from 20%–100%(0.73 ± 0.11) to 80%–100%(0.68 ± 0.15). Significance. DECT-derived perfusion showed superior reproducibility to 4DCT ventilation between timepoints 1 and 2, with consistently higher DSC and SSIM across thresholds. Despite systematic perfusion increases early in treatment, its spatial stability remained stronger, supporting DECT perfusion as a more robust biomarker for longitudinal lung function assessment and for informing adaptive, functionally-guided RT in NSCLC.

Read PDF

Similar papers

Open access Aug 2026

Pulmonary ventilation images from planning CT images using deep learning on locally‐advanced NSCLC patients

Abstract Background Pulmonary ventilation imaging has become an increasingly important component in thoracic radiotherapy, as it is applied to functional avoidance radiotherapy, where radiation doses are strategically minimized to preserve functional lung regions. Purpose We developed a framework that generates compute...

Y. Katsuta, T. Hoshino, Takaya Yamamoto et al. · 0 citations
Open access Oct 2026

Independent Component Analysis for Non-Contrast-Enhanced Free-Breathing Pulmonary Ventilation and Perfusion MRI.

PURPOSE To introduce a robust approach for non-contrast-enhanced free-breathing functional lung MRI based on independent component analysis (ICA), addressing key limitations of standard methods such as sensitivity to irregular breathing or irregular heartbeat. METHODS Repeated dynamic single-slice MRI of ten healthy...

Laura Rozo-Pardo, R. Klaar, E. Schulz et al. · 0 citations
Open access Oct 2026

Evaluating Perfusion Quantification for Phase‐Resolved Functional Lung MRI Using Gaussian Filtering

Phase‐resolved functional lung (PREFUL) MRI enables contrast‐free lung perfusion quantification (QQuant). Current quantification uses a single median value per slice of maximal blood exchange fraction (ΔS), which limits regional assessment, whereas regional ΔS causes artifacts. This retrospective study aims to optimize...

Kirsten J. Schmidt, A. Voskrebenzev, J. Glandorf et al. · 0 citations
Open access Oct 2026

Feasibility of free-breathing 0.55-T lung MRI for structural and functional characterization in preterm infants with bronchopulmonary dysplasia.

BACKGROUND Neonatal lung diseases such as bronchopulmonary dysplasia (BPD) are currently characterized by clinical observations alone, owing to the lack of safe, non-invasive diagnostic tools for lung structure and function assessment. OBJECTIVE This study aims to investigate the feasibility of simultaneous structura...

Xin Miao, HaiThuy N. Nguyen, N. Lee et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.