Musculoskeletal MRI Relaxometry Phantom for T1, T2, and T1ρ Measurements.

Oct 2026 · Journal of Magnetic Resonance Imaging · 0 citations · 44 references
Medicine

TL;DR

The phantom demonstrated excellent stability and reproducibility with temperature correction, supporting its use for quality assurance in longitudinal and multisite qMRI studies.

Abstract

Background

Quantitative MRI (qMRI) requires quality assurance methods that reflect in vivo conditions to ensure consistent measurements with low bias and high reproducibility over time, across multiple sites, and across different MRI scanner manufacturers and models.

Purpose

To develop a musculoskeletal (MSK) relaxometry phantom that spans the T1 and T2 (and T1ρ as an exploratory aim) ranges of articular cartilage. STUDY TYPE Longitudinal phantom study. PHANTOM A cylindrical phantom designed to fit within multiple knee coil configurations using dedicated support stands. The phantom contains 12 gel-filled vials spanning T1 values of 300-1200 ms, T2 values of 10-100 ms, and T1ρ values of 20-120 ms, and an MR-visible thermometer. FIELD STRENGTH/SEQUENCE 3T; single-echo spin-echo T2; inversion-recovery T1; spin-lock prepared T1ρ for NMR (spin-lock frequencies 300-5000 Hz) and MAPSS T1ρ for MRI (spin-lock frequency 500 Hz). ASSESSMENT NMR and two MRI systems from different vendors were used. A prototype phantom (n = 1), production phantoms (n = 12), and matched NMR samples were evaluated. NMR data acquired at multiple temperatures served as reference and enabled temperature correction of MRI measurements. The prototype phantom was scanned longitudinally (15 measurements over 725 days) to assess stability, while production phantoms were scanned at a single timepoint to evaluate batch consistency. STATISTICAL ANALYSIS Measurement variability and agreement were assessed using coefficients of variation (CVs), concordance correlation coefficients (CCCs), and Bland-Altman analysis for NMR-MRI comparison, longitudinal stability, batch consistency, and inter-vendor reproducibility.

Results

After temperature correction, longitudinal CVs were 1.1% (T1), 1.8% (T2), and 1.6% (T1ρ). Production phantoms showed inter-phantom CVs of 0.8% (T1), 1.2% (T2), and 0.9% (T1ρ). NMR and MRI measurements demonstrated strong agreement (CCCs > 0.99 for T1, T2, and T1ρ) with biases < 6%. Inter-vendor MRI comparisons yielded CCCs > 0.99 for T1 and T2. DATA

Conclusion

The phantom demonstrated excellent stability and reproducibility with temperature correction, supporting its use for quality assurance in longitudinal and multisite qMRI studies. EVIDENCE LEVEL 1. TECHNICAL EFFICACY Stage 1.

View source

Similar papers

Open access Sep 2026

Multidimensional T1−T2 Relaxation Imaging In vivo on a Portable 0.064 T MRI Scanner

Whole-brain multidimensional T1-T2 relaxation imaging is clinically feasible on a portable 0.064 T MRI scanner and may support further developments of quantitative multidimensional imaging at ultra-low field.

Ella Wilczynski, K. Magdoom, A. Avram et al. · 0 citations
Open access Sep 2026

Native myocardial T1 mapping using rapid 1.5T T1FLASH and T1 MOLLI during breath-hold and free-breathing

Quantitative myocardial T1 mapping is an established technique in cardiovascular magnetic resonance (CMR). However, conventional T1 mapping requires relatively long breath-holds (BH), which may lead to breathing artifacts and reduced image quality in patients with limited BH capacity. The T1 fast low angle shot (T1FLAS...

L. Röwer, Yara Hussen-Demir, P. E. Verde et al. · 0 citations
Open access Aug 2026

Robust T1ρ, T2, and T2* mapping via spin-locked MOLED with synthetic data-driven deep learning reconstruction

SL-MOLED provides accurate, repeatable, and artifact-robust quantitative mapping within a substantially reduced acquisition time, offering a promising framework for reliable multi-parametric MRI with potential for clinical translation.

Wei-Kun Chen, Qing Lin, Tai-Shan Kang et al. · 0 citations
Open access Sep 2026

Design of an In Vivo-Like Phantom for Quantitative Body CEST MRI.

PURPOSE To design and validate a reproducible in vivo-like fat-containing CEST phantom that enables robust quantitative CEST MRI for body applications by providing physiologically realistic relaxation properties, relevant endogenous CEST contrasts (amide, guanidino, rNOE, and semi-solid MT), and a controlled fat compon...

P. Menshchikov, Petr Bulanov, Neele Kempa et al. · 0 citations
Open access Oct 2026

Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.

PURPOSE Hepatic MRI T1 relaxation time is a putative measure of fibrosis and inflammation but may be confounded by hepatic fat. This study evaluated a chemical shift-encoded Modified Look-Locker Inversion Recovery (CSE-MOLLI) technique for simultaneous water-specific T1 (water-T1), proton density fat fraction (PDFF), a...

Amol Pednekar, Joshua S. Greer, Pradipta Debnath et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.