Accuracy of maxillary positioning in semi-guided orthognathic surgery using 3D-printed cutting guides and bending plates: A retrospective study.

Oct 2026 · Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · Vol 54 12, pp. 109948 · 0 citations · 18 references
Medicine

TL;DR

Preliminary findings suggest that semi-guided orthognathic surgery can transfer the VSP with deviations within the range observed for the conventional splint-based workflow, and may represent a clinically promising alternative to conventional splint-based workflows and fully patient-specific implant-based approaches in orthognathic surgery.

Abstract

Additively manufactured interocclusal splints and patient-specific implants (PSI) are well-established methods for transferring virtual planning in orthognathic surgery. However, both methods entail specific limitations in terms of time and cost efficiency. This retrospective study evaluated the accuracy of maxillary positioning using a novel semi-guided workflow based on patient-specific cutting guides and pre-bent fixation plates. Nineteen patients who underwent bimaxillary orthognathic surgery with Le Fort I osteotomy at the University Hospital Basel between 2023 and 2025 were included. Pre- and postoperative CT or CBCT imaging in combination with virtual surgical planning (VSP) data were analysed to assess maxillary and mandibular positional deviations, osteotomy accuracy, and operative time. The newly described semi-guided workflow demonstrated a high osteotomy precision, with deviations ranging from 0.09 to 1.34 mm and a mean deviation of 0.56 mm. Overall, maxillary positioning showed no statistically significant intergroup differences compared to the conventional splint-based workflow. A significant difference of limited clinical relevance was found only for absolute vertical maxillary deviation, which was greater in the semi-guided group. In addition, mean operative time was 28.8 min shorter (p = 0.173) in the semi-guided group. These preliminary findings suggest that semi-guided orthognathic surgery can transfer the VSP with deviations within the range observed for the conventional splint-based workflow. The described hybrid workflow may therefore represent a clinically promising alternative to conventional splint-based workflows and fully patient-specific implant-based approaches in orthognathic surgery.

Read PDF

Similar papers

Open access Oct 2026

ACCURACY OF HARD TISSUE PREDICTIONS IN MONO- AND BIMAXILLARY ORTHOGNATHIC SURGERY USING 3D VIRTUAL SURGICAL PLANNING.

Virtual surgery planning demonstrated clinically acceptable accuracy for maxillary positioning in midsagittal plane, however, mandibular vertical repositioning did not meet predefined equivalence criteria in the selected angular parameters.

Sacha André Decho, K. Tolksdorf, Christoph Hennig et al. · 0 citations
Open access Sep 2026

Accuracy of Combined Osteotomy Guides and Patient-Specific Implants in Le Fort I Segmental Osteotomy Requiring Alveolar Bone Reduction

Background: Segmental Le Fort I osteotomy combined with interdental alveolar bone reduction is a technically challenging procedure, and the accuracy of patient-specific implants (PSIs) in this setting has not been well established. This study aimed to assess the three-dimensional (3D) repositioning accuracy and safety...

Xin-Yu Xu, Ling Wu, Zi-Li Li et al. · 0 citations
Sep 2026

Surgical cutting and repositioning guides for minimally invasive orthognathic surgery: a retrospective comparative cohort study of bilateral sagittal split osteotomy.

Customized surgical guides may improve the accuracy of minimally invasive orthognathic surgery, particularly in facilitating the transfer of three-dimensional virtual surgical planning (3D-VSP) to the operative setting. This retrospective comparative cohort study included patients undergoing bimaxillary surgery with si...

G. De Fazio, A. Diez-Fraile, T. Faro et al. · 0 citations
Open access Sep 2026

Novel Reverse Zygomatic Implant Approach: Expanded Cadaveric Validation, Surgical Accuracy, and Comparison of Orbital Access Routes—Part 2

Guided reverse zygomatic implant placement was technically feasible and reproducible regarding implant stability in this cadaveric model, however, cortical fracture risk and incomplete prosthetic correspondence indicate that further optimization and prospective clinical validation are required before routine clinical a...

A. Ferrer-Fuertes, F. Cuesta-González, R. Sieira-Gil et al. · 0 citations
Open access Sep 2026

Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.

Accuracy varied by guidance level and greater plan-to-implant accuracy was not accompanied by improved inter-implant parallelism, indicating that these outcomes should be evaluated separately.

Eric C. MacFadden, Chris M. Sobczak, A. Guentsch · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.