Aug 2026· Frontiers in Oncology· Vol 16· 0 citations· 35 references
Medicine
TL;DR
It is suggested that HyperSight provides HU accuracy closer to that of simulation CT, improved image quality, and more accurate CBCT-based dose recalculation than Halcyon 3.1, thereby supporting the identification of clinically meaningful anatomical changes.
Abstract
Introduction Anatomical changes such as seroma shrinkage, soft-tissue deformation, and tumor-bed displacement frequently occur during breast radiotherapy and may compromise target coverage or increase unnecessary dose to surrounding tissues when the initial plan is maintained throughout treatment. Offline adaptive radiotherapy (ART) requires CBCT systems with CT-equivalent HU accuracy and robust dose-recalculation accuracy to reliably quantify these changes. This study evaluated whether the newly integrated HyperSight CBCT on the Halcyon platform provides high image quality, HU-to-electron density accuracy, and dosimetric reliability to support offline ART in breast cancer. Methods Twelve patients who underwent treatment with both Halcyon 3.1 and Halcyon-HyperSight systems were analyzed. Image quality was assessed using structural similarity metrics (SSIM, RMSE, NRMSE), uniformity, signal and noise characteristics, contrast-to-noise ratio (CNR), and edge-based indices. HU-to-ED curves were generated using a tissue-equivalent electron-density phantom and applied for CBCT-based dose recalculation. Results Compared with Halcyon 3.1, HyperSight demonstrated significantly improved structural similarity to simulation CT, with higher SSIM and lower RMSE/NRMSE, as well as significantly improved uniformity and edge sharpness. HU-to-ED linearity for HyperSight closely matched simulation CT across soft-tissue and bone-equivalent materials, whereas Halcyon 3.1 exhibited HU bias and large variability. Dosimetrically, HyperSight accurately reproduced CT-based CI_Paddick and HI values, showing markedly reduced deviations relative to Halcyon 3.1. Clinical evaluation further showed that HyperSight identified volumetric and positional changes in both the whole breast and tumor bed, identifying regions susceptible to high-dose spill or under-coverage when no plan adaptation was performed. Discussion These findings suggest that HyperSight provides HU accuracy closer to that of simulation CT, improved image quality, and more accurate CBCT-based dose recalculation than Halcyon 3.1, thereby supporting the identification of clinically meaningful anatomical changes. HyperSight may therefore serve as a practical imaging option for offline ART in breast radiotherapy and may help reduce reliance on repeat simulation CT in selected cases while supporting adaptive treatment assessment.
HyperSight has the potential to be an alternative image used for dose calculation directly on HS-CBCT in the ART workflow, particularly in prostate cancer, and demonstrates good image quality and CT number accuracy.
PURPOSE
To evaluate the dosimetric impact of anatomical changes during whole-breast irradiation (WBI) and to assess the utility of offline adaptive radiation therapy (ART) in patients treated with the field-in-field (FIF) technique.
METHODS
A dose of 42.4 Gy in 16 fractions was prescribed to the whole-breast planning...
Sangwook Lim, Kyung Ran Park, Dong Hyun Lee et al.· Physica medica (Testo stampa...· 0 citations
Abstract Background Cone‐beam computed tomography (CBCT) is increasingly used in adaptive radiotherapy (ART), yet the dosimetric accuracy and image quality of advanced CBCT imaging platforms—such as Halcyon HyperSight 4.0 (HH), which combines an upgraded detector panel with advanced iterative reconstruction, scatter‐co...
B. Park, Taeyoon Kim, Kwan-Woo Lee· Journal of Applied Clinical...· 0 citations
iCBCT with a dedicated HU-density calibration curve provides dose calculation accuracy comparable to planning CT, supporting reliable offline ART in prostate radiotherapy.
María Aurora Vicedo González, M. T. García Hernández, M. L. Alcaraz Lozano et al.· Physica medica (Testo stampa...· 0 citations
Daily CBCT image-guided couch corrections maintain dosimetric accuracy during hypofractionated prostate radiotherapy and support safe implementation of tight PTV margins in resource-constrained public-sector healthcare settings.
Leanore Legoale-Sebopa, Sheynaz Bassa, H. V. Maselesele· South African Journal of Onc...· 0 citations
Abstract Background Cone‐beam computed tomography (CBCT) is integral to image‐guided radiation therapy and increasingly used for adaptive radiotherapy and CBCT‐based dose calculation. Conventional kV‐CBCT systems on C‐arm TrueBeam linacs with Feldkamp–Davis–Kress (FDK) reconstruction have historically been limited by r...
Niamh Clarke, J. Hughes, D. Wanigaratne et al.· Journal of Applied Clinical...· 0 citations