Skip to content
Review

The Role of Preprocedural CT in Redo Transcatheter Aortic Valve Replacement: A Radiologist's Perspective.

Aug 2026 · Radiology: Cardiothoracic Imaging · Vol 8 4, pp. e250200 · 0 citations · 28 references
Medicine

TL;DR

Differences are primarily related to the need to evaluate the characteristics of the index THV and the degree of commissural and coronary alignment achieved during the index TAVR procedure, which affect the selection of the type and size of the new THV and determine the risks of coronary obstruction and impeded future coronary access, as well as the potential for leaflet modification.

Abstract

As transcatheter aortic valve replacement (TAVR) has become a viable option across all surgical risk levels, younger patients are anticipated to outlive the durability of their index transcatheter heart valve (THV). When a THV fails, cardiac CT is an important tool to assess the mechanism of THV failure and aids in the differentiation between structural and nonstructural THV dysfunction mechanisms. After the identification of the index THV dysfunction mechanism, CT guides the sizing and selection of the new THV type. While the principles of preprocedural redo TAVR (often called TAV-in-TAV) CT assessment overlap with those for patients undergoing TAVR of surgical bioprosthetic valves, the assessment is different in several ways. These differences are primarily related to the need to evaluate the characteristics of the index THV and the degree of commissural and coronary alignment achieved during the index TAVR procedure, which affect the selection of the type and size of the new THV and determine the risks of coronary obstruction and impeded future coronary access, as well as the potential for leaflet modification. Keywords: Applications-CT, CT Angiography, Transcatheter Aortic Valve Implantation/Replacement, TAVI/TAVR Supplemental material is available for this article. ©RSNA, 2026.

View source

Similar papers

Open access Aug 2026

Redo-Transcatheter Aortic Valve Replacement With Self-Expanding Valves: A U.S. Registry Analysis.

In a national registry, redo-TAVR with the Evolut platform was associated with low procedural complication rates and clinical outcomes comparable with those observed with native TAVR and support redo-TAVR with Evolut as a feasible and effective treatment strategy in appropriately selected patients.

Toby Rogers, V. Bapat, S. Chetcuti et al. · 1 citation
Review Open access Aug 2026

Transcatheter atrioventricular valve-in-valve re-replacement with balloon-expandable prostheses: contemporary clinical relevance and place among redo valve interventions

At present, transcatheter valve-in-valve re-replacement in atrioventricular positions should be regarded not merely as an alternative for patients at prohibitive risk, but as a clinically justified component of the contemporary treatment strategy capable of substantially reducing procedural burden and accelerating reha...

A. Gorbatykh · 0 citations
Review Open access Aug 2026

Advanced 3D echocardiographic assessment of the aortic valve and aortic root complex: a repair-oriented perspective

Three-dimensional echocardiography, particularly transoesophageal 3DE, has become an essential imaging modality for integrating anatomical and functional information and plays a pivotal role in promoting a repair-oriented approach to the management of aortic valve disease.

T. Viva, C. Fiore, R. Dulgheru et al. · 0 citations
Editorial Open access Aug 2026

When redo-transcatheter aortic valve replacement (TAVR) is not enough: self-expanding TAVR explant and the next frontier of lifetime valve management

The central question is therefore no longer whether failed TAVR will be encountered, but whether the initial procedure and every subsequent valve or coronary intervention preserve a safe exit strategy when failure occurs.

Ahmad Shrouf, Ahmad Jabri, M. Abdalaziz et al. · 0 citations