Aug 2026· Scientific Reports· Vol 16· 0 citations· 11 references
Medicine
TL;DR
The study concludes that the Wondfo POCT device demonstrates acceptable agreement with standard laboratory testing within therapeutic ranges, thereby offering a viable alternative for rapid, decentralized INR monitoring, particularly in outpatient care, and can potentially improve patient management and safety.
Abstract
Reliable INR monitoring is crucial for patients on warfarin to prevent thrombotic or bleeding events. This study evaluated the accuracy of the Wondfo POCT device compared to standard laboratory methods (Stago STA-R Max3 and ACL TOP 300) in a primary healthcare setting. The goal was to assess whether this point-of-care testing can provide timely and reliable results for better patient management in community-based clinics. In a prospective, cross-sectional study, samples from 74 patients on warfarin and 20 healthy controls were analyzed. Agreement was assessed using Bland-Altman plots, Deming regression, and CLSI POCT14-A criteria. Deming regression demonstrated strong linear correlations across all comparisons (r > 0.92, R² > 0.84). The mean bias for Wondfo (capillary) remained within CLSI POCT14-A agreement limits for all INR ranges when compared with both Stago and ACL TOP 300 platforms (all SE < AE). A very strong correlation was also observed between Wondfo venous and capillary results (r = 0.954, R² = 0.908). The study concludes that the Wondfo POCT device demonstrates acceptable agreement with standard laboratory testing within therapeutic ranges, thereby offering a viable alternative. It is a reliable and practical solution for rapid, decentralized INR monitoring, particularly in outpatient care, and can potentially improve patient management and safety.
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels, particularly at higher glucose levels.
Ufuoma Ohwo, K. Digban, C. Okafor· Apollo Medicine· 0 citations
Background: Point-of-care (POC) blood glucose meters shorten turnaround time for bedside glycemic management, but critical illness introduces physiologic and analytical conditions that can degrade agreement with central laboratory measurements. Objective: To synthesize evidence on the analytical and clinical accuracy o...
C. Amorado, Leah Quinto, Mark Kerwin F. Sayas et al.· International Journal of Hea...· 1 citation
ABL90 demonstrated the best overall concordance with the reference analyzer in terms of both creatinine measurement and eGFR categorization, followed by i-STAT and pHOx Ultra, which may be attributable to their manufacturer-stated traceability to the NIST reference standards.
S. Yun, Hee-Jun Yun, K. Kim et al.· Clinical Chemistry· 0 citations
Abstract Objectives We validated the analytical performance and clinical diagnostic efficacy of the eCL8000i point-of-care testing (POCT) system for high-sensitivity cardiac troponin T (hs-cTnT), following the standardized verification protocol for hs-cTnT POCT assays. Methods A two-phase evaluation was conducted: Phas...
Fei Ding, Yu-Yao Li, Jia Li et al.· Journal of Laboratory Medici...· 0 citations
Point of care (POC) prothrombin time (PT)/international normalized ratio (INR) testing can significantly enhance coagulation monitoring by increasing test access, improving therapeutic management and adherence, decreasing patient travel, increasing clinic throughput, and patient satisfaction. However, implementin...
Marie Good, Stephanie Kinney, Erin E. Schuler· Clinical Chemistry· 0 citations
POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.
E. Nwobodo, N. T. Emmanuel· IOASD Journal of Medical and...· 0 citations