Toward an optimized PET reconstruction protocol for spinal cord: superiority of OSEM with TOF and PSF

Aug 2026 · EJNMMI Physics · Vol 13 · 0 citations · 51 references
Medicine

TL;DR

The OSEM + PSF + TOF reconstruction method demonstrated significant advantages in both image quality and quantitative assessment of FDG uptake, making it the recommended choice for spinal cord PET image reconstruction.

Abstract

Although multiple PET reconstruction approaches have been successfully applied in brain and oncologic imaging, their performance in the context of spinal cord imaging remains unverified. This study aims to identify the optimal reconstruction method for spinal cord PET image analysis based on assessments of image quality and quantitative uptake. We retrospectively analyzed 18F-fluorodeoxyglucose (FDG) PET/MR data from 22 patients with cervical spinal cord tumors. PET images were reconstructed using Ordered Subset Expectation Maximization (OSEM), OSEM + Point Spread Function (PSF), OSEM + Time-of-Flight (TOF), and OSEM + TOF + PSF algorithms. Image quality was assessed by noise, signal-to-noise ratio (SNR), and signal-to-background ratio (SBR), both overall and within equally sized small (n = 11) and large (n = 11) tumor subgroups. Tumor uptake metrics, including SUVmean, SUVmax, SUVRmean, and SUVRmax of the lesions, were quantified for each method. The differences among reconstruction methods were assessed using the Friedman test. There were no significant differences in image noise among reconstruction methods (P > 0.05). Compared to OSEM reconstruction, the OSEM + PSF + TOF method significantly improved SNR (Z = 3.62, P < 0.01) and SBR (Z = 2.80, P = 0.03); the SBR of OSEM + PSF + TOF was also significantly higher than that of OSEM + PSF (Z = 3.04, P = 0.01). In large tumors, while image noise remained unaffected by the reconstruction method (P > 0.05), both SNR and SBR were significantly impacted (both P < 0.05). Specifically, OSEM + TOF + PSF achieved significantly higher SNR compared to OSEM (Z = 3.30, P < 0.01), and higher SBR compared to OSEM + PSF (Z = 3.14, P = 0.01). Quantitative lesion uptake metrics (SUVmax, SUVmean, SUVRmax, and SUVRmean) were significantly higher for OSEM + TOF + PSF compared to both OSEM and OSEM + PSF reconstructions (all P < 0.05). Additionally, SUVmean (Z = 2.92, P = 0.02) and SUVRmean (Z = 2.83, P = 0.03) values from OSEM + TOF + PSF reconstruction were significantly higher than those from OSEM + TOF. The OSEM + PSF + TOF reconstruction method demonstrated significant advantages in both image quality and quantitative assessment of FDG uptake, making it the recommended choice for spinal cord PET image reconstruction.

Read PDF

Similar papers

Open access Oct 2026

Optimization of PET acquisition time for thoracic lesions in 18F-FDG PET/MRI: a retrospective analysis

This study aimed to define the minimum positron emission tomography (PET) acquisition time for chest PET/magnetic resonance imaging (MRI), providing image quality and lesion detectability comparable to a 20-min reference. In this retrospective study, list-mode data from 101 patients undergoing chest 18F-fluorodeoxygluc...

Hai-Feng Gu, Yue Jiang, Jun Cai et al. · 0 citations
Sep 2026

MRI-guided versus OSEM Amyloid Brain PET Reconstruction for Reader Agreement, Visual Consistency, and Diagnostic Accuracy.

Background Ordered-subset expectation-maximization (OSEM) PET reconstruction is limited by spatial resolution and partial-volume effects, contributing to interreader variability and false positives and highlighting the need for more comprehensive alternative approaches. Purpose To evaluate whether MRI-guided brain PET...

Yu Cai, Chen Lin, Manoj K. Jain et al. · 1 citation
Open access Sep 2026

Evaluation of Intracranial Lesions Using Deep Learning-Based Reconstruction in Canine Brain MRI: Comparison With Conventional Reconstruction on T2-Weighted and FLAIR Sequences.

Obtaining high-resolution brain MRI in small animals is challenging due to the inherent trade-off between image quality and acquisition time. Deep learning-based reconstruction (DLR) has emerged as a solution to improve image quality without prolonging scan time; however, its clinical utility and potential risks, such...

Wooseok Jin, Hye-Ran Na, Sang-Kwon Lee et al. · 0 citations
Open access Oct 2026

CT protocol optimization for oncologic surveillance imaging: phantom-based comparison of multi-scan and single-scan acquisition strategies

Optimize CT acquisition protocols reduced radiation dose while preserving the evaluated objective image quality metrics in the cranial, thoracic, and upper abdominal regions, and was identified as a candidate protocol for subsequent clinical and reader performance validation.

D. Kostova-Lefterova, F. Simeonov, D. Hitova-Topkarova et al. · 0 citations
Open access Sep 2026

Evaluation of accelerated whole-body diffusion weighted imaging with deep learning reconstruction in patients with metastatic prostate cancer: assessment of image quality and ADC estimates.

DL-based reconstruction enables faster WB-DWI acquisition without compromising image quality or ADC quantification, supporting clinical implementation in metastatic prostate cancer staging and therapy assessment.

E. Kousi, M. Rata, Doh-Mu Koh et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.