Contact Index for PulseSelect Catheter Apposition: Validation by Transesophageal Echocardiography.

Jul 2026 · Pacing and clinical electrophysiology : PACE · 0 citations · 3 references
Medicine

Abstract

Background

Reliable catheter-tissue apposition is important for effective pulmonary vein isolation using pulsed field ablation. The contact index, derived from real-time impedance rise on the EnSite X mapping system, may provide an objective surrogate for catheter contact; however, its utility with the PulseSelect catheter remains insufficiently defined.

Methods

In this single-center pilot study, patients with paroxysmal atrial fibrillation undergoing pulmonary vein isolation with the PulseSelect catheter under general anesthesia were evaluated. High-resolution transesophageal echocardiography was used as the reference standard to classify electrode-level catheter apposition as contact or non-contact. Contact index values from electrodes 3-7 were paired with transesophageal echocardiography findings, and receiver operating characteristic curve analysis was performed.

Results

A total of 316 paired imaging-electrical data points were analyzed. Interobserver agreement for transesophageal echocardiography-based contact assessment was excellent (Cohen's κ = 0.86). Contact index values were higher at contact than non-contact segments. A contact index cutoff of 10% yielded an area under the curve of 0.864, with 76.1% sensitivity and 82.9% specificity. In the central electrodes 4-6, the area under the curve was 0.871, with 80.1% sensitivity and 81.2% specificity.

Conclusions

Contact index was associated with transesophageal echocardiography-confirmed catheter-tissue apposition during pulsed field ablation using the PulseSelect catheter. A 10% threshold may be a useful candidate reference for real-time apposition assessment, requiring validation in larger studies.

View source

Similar papers

Open access Sep 2026

Impedance Changes During Intracardiac Echocardiography-Confirmed Catheter Contact in Pulsed Field Ablation.

PURPOSE Pulsed field ablation (PFA) has become an established treatment modality for atrial fibrillation (AF). Adequate catheter-to-tissue contact is essential for effective energy delivery. However, impedance characteristics associated with directly visualized catheter contact during PFA have not been well characteriz...

Kohki Kimura, Hirotoshi Nagae, Y. Yamamoto et al. · 1 citation
Open access Aug 2026

Construct Validity of a Novel Fiberoptic TAVR Guidewire for Invasive Pressure Measurements.

The fiberoptic guidewire effectively classified clinically relevant transvalvular gradients with limited bias post-TAVR, although it produced higher LVEDPs, possibly reflecting wire-induced leaflet interaction.

Geert A. A. Versteeg, M. Westerhof, N. Stens et al. · 0 citations
Open access Sep 2026

Clinical and experimental evaluation of averaged impedance drop during radiofrequency catheter ablation.

The dynamics of averaged impedance drop percentage (AID%), a novel ablation parameter, and its association with conduction gaps and lesion characteristics suggest that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric.

Nobuhiro Nakamura, S. Oka, A. Wakamiya et al. · 0 citations
Oct 2026

Electrogram Determinants of Acute Electrogram Attenuation During Pulmonary Vein Pulsed Field Ablation.

BACKGROUND Pulsed field ablation (PFA) is a nonthermal modality for pulmonary vein isolation in atrial fibrillation. Whether baseline intracardiac electrogram characteristics are associated with acute electrogram attenuation during PFA remains unclear. OBJECTIVES To evaluate the individual and combined associations o...

R. Usui, Akihiko Nogami, Y. Oda et al. · 0 citations
Oct 2026

Feasibility of tissue proximity indication during atrial fibrillation rhythm as an indicator of acute pulmonary vein reconnection using variable-loop pulsed field ablation catheter.

TPI assessment during AF rhythm is useful, particularly with version 8.1 and combined with bipolar voltage, despite a slight site-dependency, but the utility of these metrics was slightly site-dependent, with reduced accuracy observed at the roof/floor lines and the anterior region of the left inferior PV and the carin...

Tetsuma Kawaji, Saki Yamano, Ryuta Ishida et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.