Impact of acquisition parameters in photon-counting computed tomography on the stability of virtual monoenergetic attenuation values and effective atomic number: a phantom study.
Aug 2026· European Journal of Radiology· Vol 204, pp.
113145
· 0 citations· 41 references
Medicine
TL;DR
P Pitch exerts the greatest influence on the stability of VMI CT numbers, whereas tube voltage and IQ level have relatively minor effects, supporting their reliability in quantitative PCCT applications.
Abstract
Objective
Photon-counting CT (PCCT) can generate virtual monoenergetic images (VMI) and effective atomic number (Zeff), demonstrating considerable potential for tissue quantification. However, the influence of key acquisition parameters on the stability of these quantitative metrics has not been fully elucidated. This study aimed to evaluate the reliability of VMI CT numbers and Zeff in abdominal organs under varying tube voltage, image quality (IQ) level, and pitch.
Methods
An adult male anthropomorphic abdominal phantom (Kyoto Kagaku CTU-41) was scanned using PCCT. The reference protocol was 120 kVp, image quality level (IQ level) 145, and pitch 1. Eight additional protocols were acquired by independently altering tube voltage (140 kVp, 90 kVp, Sn100 kVp), IQ level (115, 85, 55, 25), and pitch (3). VMI were reconstructed at 40, 60, 70, and 100 keV, and CT numbers and Zeff were measured in multiple organs. Differences from the reference protocol were calculated to assess the reliability of these quantitative indices under modified scan conditions.
Results
Across all protocols, VMI attenuation showed excellent correlation with the reference protocol (r ≥ 0.996). For IQ levels of 115, 85, and 55, all absolute CT number deviations were below 5 HU; at IQ level 25, only two measurements fell within the 5-10 HU range. At 140 kVp, all absolute deviations were below 5 HU. For the 90 kVp protocol, the prostate deviation at 40 keV was -11.33 HU, and five other measurements ranged between 5 and 10 HU. With the Sn100 kVp protocol, absolute deviations of the L1 vertebra and prostate exceeded 10 HU at 60 keV (L1 reached 19.33 HU), and three additional measurements were between 5 and 10 HU. Pitch alteration caused the greatest variability: only 75% of measurements had absolute deviations within 10 HU and 67.5% within 5 HU; deviations were prominent on 40 keV images, with the L1 vertebra reaching -41 HU. The intraclass correlation coefficient for Zeff was ≥0.999 under all conditions. Zeff deviations were within 0.2 for 140 kVp and IQ levels of 115, 85, and 55; the pancreas deviation was 0.34 at IQ level 25; at pitch 3, the maximum deviation (L1) was 0.37, and several organs showed deviations close to or exceeding 0.2.
Conclusions
Pitch exerts the greatest influence on the stability of VMI CT numbers, whereas tube voltage and IQ level have relatively minor effects. VMI at medium-to-high energy levels (≥70 keV) exhibit relative stability across different parameters and are recommended as the first choice for quantitative analysis. Zeff measurements remain highly stable under various acquisition conditions, supporting their reliability in quantitative PCCT applications.
This study investigated the kV dependence of VMI CT values in PCD-CT across a wide range of tissue types, keV levels, and kV settings for various phantom sizes, and to determine which kV settings yield the optimal contrast-to-noise ratio (CNR) for a given VMI energy and phantom size.
C. Jeukens, F. M. E. Pinckaers, B. Martens et al.· Investigative Radiology· 0 citations
This study investigated the relationship between body size, administered activity, and image quality (IQ) using two patient-like phantoms representing normal and obese body profiles, and demonstrated the count rate capabilities up to and beyond the saturation limit.
Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al.· Physics in Medicine and Biol...· 0 citations
The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.
E. Karić, M. Kukuljan, L. Lezaic et al.· Radiography· 0 citations
The findings support implementing standardized protocols with automatic dual-range reconstructions to enable efficient, individualized carotid plaque assessment in clinical practice and support implementing standardized protocols with automatic dual-range reconstructions for keV selection based on plaque composition.
Mirela Kostadinova, Lea B. Uebelacker, Tommaso D'Angelo et al.· Diagnostics· 0 citations
For spinal implants, PCD-CT offers effective approaches for metal artifact and 120 keV is preferred and the combination of iMAR and 120 keV introduce new artifacts and requires careful consideration.
Xin-Ying Huang, Zai-Zhu Zhang, Yong An et al.· European Journal of Radiolog...· 0 citations
Although DE-CBCT with VMI and frequency splitting improves HU agreement with CT, slightly improves soft tissue visualization, and reduces artifacts in head and neck phantom imaging, it did not reveal any additional targets compared with routine CBCT.
D. Jayakody, V. Riekki, M. Bode et al.· Biomedical engineering and p...· 0 citations